Increased rate of catheter-related bloodstream infection associated with use of a needleless mechanical valve device at a long-term acute care hospital.

Cassandra D Salgado1, Libby Chinnes, Tammy H Paczesny, J Robert Cantey.   

Abstract

OBJECTIVE: To determine whether introduction of a needleless mechanical valve device (NMVD) at a long-term acute care hospital was associated with an increased frequency of catheter-related bloodstream infection (BSI).
DESIGN: For patients with a central venous catheter in place, the catheter-related BSI rate during the 24-month period before introduction of the NMVD, a period in which a needleless split-septum device (NSSD) was being used (hereafter, the NSSD period), was compared with the catheter-related BSI rate during the 24-month period after introduction of the NMVD (hereafter, the NMVD period). The microbiological characteristics of catheter-related BSIs during each period were also compared. Comparisons and calculations of relative risks (RRs) with 95% confidence intervals (CIs) were performed using chi (2) analysis.
RESULTS: Eighty-six catheter-related BSIs (3.86 infections per 1,000 catheter-days) occurred during the study period. The rate of catheter-related BSI during the NMVD period was significantly higher than that during the NSSD period (5.95 vs 1.79 infections per 1,000 catheter-days; RR, 3.32 [95% CI, 2.88-3.83]; P<.001). A significantly greater percentage of catheter-related BSIs during the NMVD period were caused by gram-negative organisms, compared with the percentage recorded during the NSSD period (39.5% vs 8%; P=.007). Among catheter-related BSIs due to gram-positive organisms, the percentage caused by enterococci was significantly greater during the NMVD period, compared with the NSSD period (54.8% vs 13.6%; P=.004). The catheter-related BSI rate remained high during the NMVD period despite several educational sessions regarding proper use of the NMVD.
CONCLUSIONS: An increased catheter-related BSI rate was temporally associated with use of a NMVD at the study hospital, despite several educational sessions regarding proper NMVD use. The current design of the NMVD may be unsafe for use in certain patient populations.

Entities:  

Mesh:

Year:  2007        PMID: 17520541     DOI: 10.1086/516800

Source DB:  PubMed          Journal:  Infect Control Hosp Epidemiol        ISSN: 0899-823X            Impact factor:   3.254


  15 in total

1.  Summary of recommendations: Guidelines for the Prevention of Intravascular Catheter-related Infections.

Authors:  Naomi P O'Grady; Mary Alexander; Lillian A Burns; E Patchen Dellinger; Jeffrey Garland; Stephen O Heard; Pamela A Lipsett; Henry Masur; Leonard A Mermel; Michele L Pearson; Issam I Raad; Adrienne G Randolph; Mark E Rupp; Sanjay Saint
Journal:  Clin Infect Dis       Date:  2011-05       Impact factor: 9.079

2.  Guidelines for the prevention of intravascular catheter-related infections.

Authors:  Naomi P O'Grady; Mary Alexander; Lillian A Burns; E Patchen Dellinger; Jeffrey Garland; Stephen O Heard; Pamela A Lipsett; Henry Masur; Leonard A Mermel; Michele L Pearson; Issam I Raad; Adrienne G Randolph; Mark E Rupp; Sanjay Saint
Journal:  Clin Infect Dis       Date:  2011-04-01       Impact factor: 9.079

3.  Needleless closed system does not reduce central venous catheter-related bloodstream infection: a retrospective study.

Authors:  Mitsuru Ishizuka; Hitoshi Nagata; Kazutoshi Takagi; Keiichi Kubota
Journal:  Int Surg       Date:  2013 Jan-Mar

4.  epic3: national evidence-based guidelines for preventing healthcare-associated infections in NHS hospitals in England.

Authors:  H P Loveday; J A Wilson; R J Pratt; M Golsorkhi; A Tingle; A Bak; J Browne; J Prieto; M Wilcox
Journal:  J Hosp Infect       Date:  2014-01       Impact factor: 3.926

5.  Outbreak column 19: needleless connectors (NCs) tales from nine outbreaks.

Authors:  Evonne Curran
Journal:  J Infect Prev       Date:  2016-06-30

6.  Blood drawn through valved catheter hub connectors carries a significant risk of contamination.

Authors:  R J Sherertz; T B Karchmer; E Palavecino; W Bischoff
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2011-05-01       Impact factor: 3.267

7.  Surface micropattern reduces colonization and medical device-associated infections.

Authors:  Binjie Xu; Qiuhua Wei; M Ryan Mettetal; Jie Han; Lindsey Rau; Jinfeng Tie; Rhea M May; Eric T Pathe; Shravanthi T Reddy; Lauren Sullivan; Albert E Parker; Donald H Maul; Anthony B Brennan; Ethan E Mann
Journal:  J Med Microbiol       Date:  2017-10-06       Impact factor: 2.472

8.  Microbial biofilms on needleless connectors for central venous catheters: comparison of standard and silver-coated devices collected from patients in an acute care hospital.

Authors:  Elizabeth Perez; Margaret Williams; Jesse T Jacob; Mary Dent Reyes; Sheri Chernetsky Tejedor; James P Steinberg; Lori Rowe; Satishkumar Ranganathan Ganakammal; Shankar Changayil; M Ryan Weil; Rodney M Donlan
Journal:  J Clin Microbiol       Date:  2013-12-26       Impact factor: 5.948

9.  Reducing the degree of colonisation of venous access catheters by continuous passive disinfection.

Authors:  Fernando Gutiérrez Nicolás; Gloria Julia Nazco Casariego; María Micaela Viña Romero; Jonathan González García; Ruth Ramos Diaz; Jose Antonio Perez Perez
Journal:  Eur J Hosp Pharm       Date:  2015-10-12

10.  Even the Simplest Devices May Malfunction: Split Septum Design Revisited.

Authors:  Rotem Naftalovich; Steven Char; Andrew J Iskander; Daniel Naftalovich
Journal:  Biomed Instrum Technol       Date:  2021-01-01