Literature DB >> 24371233

Microbial biofilms on needleless connectors for central venous catheters: comparison of standard and silver-coated devices collected from patients in an acute care hospital.

Elizabeth Perez1, Margaret Williams, Jesse T Jacob, Mary Dent Reyes, Sheri Chernetsky Tejedor, James P Steinberg, Lori Rowe, Satishkumar Ranganathan Ganakammal, Shankar Changayil, M Ryan Weil, Rodney M Donlan.   

Abstract

Microorganisms may colonize needleless connectors (NCs) on intravascular catheters, forming biofilms and predisposing patients to catheter-associated infection (CAI). Standard and silver-coated NCs were collected from catheterized intensive care unit patients to characterize biofilm formation using culture-dependent and culture-independent methods and to investigate the associations between NC usage and biofilm characteristics. Viable microorganisms were detected by plate counts from 46% of standard NCs and 59% of silver-coated NCs (P=0.11). There were no significant associations (P>0.05, chi-square test) between catheter type, side of catheter placement, number of catheter lumens, site of catheter placement, or NC placement duration and positive NC findings. There was an association (P=0.04, chi-square test) between infusion type and positive findings for standard NCs. Viable microorganisms exhibiting intracellular esterase activity were detected on >90% of both NC types (P=0.751), suggesting that a large percentage of organisms were not culturable using the conditions provided in this study. Amplification of the 16S rRNA gene from selected NCs provided a substantially larger number of operational taxonomic units per NC than did plate counts (26 to 43 versus 1 to 4 operational taxonomic units/NC, respectively), suggesting that culture-dependent methods may substantially underestimate microbial diversity on NCs. NC bacterial communities were clustered by patient and venous access type and may reflect the composition of the patient's local microbiome but also may contain organisms from the health care environment. NCs provide a portal of entry for a wide diversity of opportunistic pathogens to colonize the catheter lumen, forming a biofilm and increasing the potential for CAI, highlighting the importance of catheter maintenance practices to reduce microbial contamination.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24371233      PMCID: PMC3957745          DOI: 10.1128/JCM.02220-13

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  29 in total

1.  Validation of the Scan RDI for routine microbiological analysis of process water.

Authors:  Stephen P Costanzo; Roya N Borazjani; Patrick J McCormick
Journal:  PDA J Pharm Sci Technol       Date:  2002 Jul-Aug

2.  Prevalence of the use of central venous access devices within and outside of the intensive care unit: results of a survey among hospitals in the prevention epicenter program of the Centers for Disease Control and Prevention.

Authors:  Michael Climo; Dan Diekema; David K Warren; Loreen A Herwaldt; Trish M Perl; Lance Peterson; Theresa Plaskett; Connie Price; Kent Sepkowitz; Steve Solomon; Jerry Tokars; Victoria J Fraser; Edward Wong
Journal:  Infect Control Hosp Epidemiol       Date:  2003-12       Impact factor: 3.254

3.  Health care-associated bloodstream infections associated with negative- or positive-pressure or displacement mechanical valve needleless connectors.

Authors:  William R Jarvis; Cathryn Murphy; Keri K Hall; Pamela J Fogle; Tobi B Karchmer; Glenys Harrington; Cassandra Salgado; Eve T Giannetta; Carol Cameron; Robert J Sherertz
Journal:  Clin Infect Dis       Date:  2009-12-15       Impact factor: 9.079

Review 4.  Biofilm elimination on intravascular catheters: important considerations for the infectious disease practitioner.

Authors:  Rodney M Donlan
Journal:  Clin Infect Dis       Date:  2011-04-15       Impact factor: 9.079

5.  Biofilm formation by gram-negative bacteria on central venous catheter connectors: effect of conditioning films in a laboratory model.

Authors:  R Murga; J M Miller; R M Donlan
Journal:  J Clin Microbiol       Date:  2001-06       Impact factor: 5.948

Review 6.  Microbial ecology of the skin.

Authors:  R R Roth; W D James
Journal:  Annu Rev Microbiol       Date:  1988       Impact factor: 15.500

7.  Increased rate of catheter-related bloodstream infection associated with use of a needleless mechanical valve device at a long-term acute care hospital.

Authors:  Cassandra D Salgado; Libby Chinnes; Tammy H Paczesny; J Robert Cantey
Journal:  Infect Control Hosp Epidemiol       Date:  2007-05-14       Impact factor: 3.254

8.  Culture-independent microbiological analysis of foley urinary catheter biofilms.

Authors:  Daniel N Frank; Shandra S Wilson; Allison L St Amand; Norman R Pace
Journal:  PLoS One       Date:  2009-11-12       Impact factor: 3.240

9.  A large genome center's improvements to the Illumina sequencing system.

Authors:  Michael A Quail; Iwanka Kozarewa; Frances Smith; Aylwyn Scally; Philip J Stephens; Richard Durbin; Harold Swerdlow; Daniel J Turner
Journal:  Nat Methods       Date:  2008-12       Impact factor: 28.547

10.  Use of cultivation-dependent and -independent techniques to assess contamination of central venous catheters: a pilot study.

Authors:  Mette K S Larsen; Trine R Thomsen; Claus Moser; Niels Høiby; Per H Nielsen
Journal:  BMC Clin Pathol       Date:  2008-10-28
View more
  9 in total

1.  Microbial diversity on intravascular catheters from paediatric patients.

Authors:  L Zhang; N Marsh; D Long; M Wei; M Morrison; C M Rickard
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2015-10-29       Impact factor: 3.267

2.  Virulence profile: Rodney Donlan.

Authors:  Rodney Donlan
Journal:  Virulence       Date:  2016-08-10       Impact factor: 5.882

3.  Microbial biofilms associated with intravascular catheter-related bloodstream infections in adult intensive care patients.

Authors:  L Zhang; J Gowardman; M Morrison; N Runnegar; C M Rickard
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2015-11-26       Impact factor: 3.267

Review 4.  Disinfection of Needleless Connector Hubs: Clinical Evidence Systematic Review.

Authors:  Nancy L Moureau; Julie Flynn
Journal:  Nurs Res Pract       Date:  2015-05-14

Review 5.  Microorganisms in Confined Habitats: Microbial Monitoring and Control of Intensive Care Units, Operating Rooms, Cleanrooms and the International Space Station.

Authors:  Maximilian Mora; Alexander Mahnert; Kaisa Koskinen; Manuela R Pausan; Lisa Oberauner-Wappis; Robert Krause; Alexandra K Perras; Gregor Gorkiewicz; Gabriele Berg; Christine Moissl-Eichinger
Journal:  Front Microbiol       Date:  2016-10-13       Impact factor: 5.640

6.  Antimicrobial activity of Ti-ZrN/Ag coatings for use in biomaterial applications.

Authors:  Anthony J Slate; David J Wickens; Mohamed El Mohtadi; Nina Dempsey-Hibbert; Glen West; Craig E Banks; Kathryn A Whitehead
Journal:  Sci Rep       Date:  2018-01-24       Impact factor: 4.379

7.  Comparative Analysis of Bacterial Community Composition and Structure in Clinically Symptomatic and Asymptomatic Central Venous Catheters.

Authors:  Franziska A Stressmann; Elodie Couve-Deacon; Delphine Chainier; Ashwini Chauhan; Aimee Wessel; Sylvaine Durand-Fontanier; Marie-Christine Escande; Irène Kriegel; Bruno Francois; Marie-Cécile Ploy; Christophe Beloin; Jean-Marc Ghigo
Journal:  mSphere       Date:  2017-09-27       Impact factor: 4.389

8.  Molecular Comparison of Bacterial Communities on Peripheral Intravenous Catheters and Matched Skin Swabs.

Authors:  Md Abu Choudhury; Nicole Marsh; Shahera Banu; David L Paterson; Claire M Rickard; David J McMillan
Journal:  PLoS One       Date:  2016-01-05       Impact factor: 3.240

9.  Contaminant-Activated Visible Light Photocatalysis.

Authors:  Vijay Krishna; Wei Bai; Zhao Han; Akihiro Yano; Abhinav Thakur; Angelina Georgieva; Kristy Tolley; Joseph Navarro; Ben Koopman; Brij Moudgil
Journal:  Sci Rep       Date:  2018-01-30       Impact factor: 4.379

  9 in total