Osteolysis caused by wear particles from polyethylene in the artificial hip joints is a serious issue. We have used photo-induced radical graft polymerization to graft 2-methacryloyloxyethyl phosphorylcholine (MPC) polymer onto the surface of cross-linked polyethylene (CLPE-g-MPC) in order to reduce friction and wear at the bearing surface of the joint. The physical and mechanical properties of CLPE and CLPE-g-MPC were not significantly different, expect that the friction coefficient of untreated CLPE cups was 0.0075, compared with 0.0009 for CLPE-g-MPC cup, an 88% reduction. After 3.0 x 10(6) cycles in the hip joint simulator test, we could not observe any wear of CLPE-g-MPC cups. We concluded that the advantage of photo-induced radical graft polymerization technique was that the grafted MPC polymer gave a high lubricity only on the surface and has no effect on the bulk properties of the CLPE substrate.Entities: