Comparison of optic nerve head topography in healthy adults using a Heidelberg retina tomograph and retinal thickness analyzer.

Andreja Rekic1, Marjanca Breznik, Barbara Cvenkel.   

Abstract

PURPOSE: To compare optic nerve head topographies measured using a Heidelberg retina tomograph I (HRT) and a retinal thickness analyzer (RTA) and thereby to determine clinical agreement between the two devices. Also, to investigate the reproducibility of optic nerve head measurements using the HRT and RTA.
METHODS: The study included 50 normal left eyes of 50 white adults. Ophthalmologic examination, keratometry, fundus photography, and examination with HRT and RTA were performed. To determine reproducibility, 10 volunteers had repeated examinations after a week. Mean (SD), median, and 5th and 95th percentiles were calculated for the HRT and RTA measurements and differences in measurements between HRT and RTA were tested for statistical significance. Clinical agreement was assessed with limits of agreement and reproducibility by a repeatability coefficient and intraclass correlation coefficient (ICC).
RESULTS: Statistically significant differences (P < 0.05) between measurements by HRT and RTA were observed for all topographic parameters, except for rim area (P = 0.051) and height variation contour (P = 0.054). The limits of agreement between HRT and RTA were wide. The repeatability coefficient for HRT was good (< or =0.10) for all parameters, except for retinal nerve fibre layer (RNFL) cross-sectional area (0.28). The repeatability coefficient for RTA was >0.10 for cup area (0.15), rim area (0.19), maximum cup depth (0.13), height variation contour (0.11), and RNFL cross-sectional area (0.14). The ICC was good (> or =90%) for all parameters, except for mean RNFL thickness (89%) for HRT, and height variation contour (84%) for RTA.
CONCLUSION: The observed differences within the limits of agreement were clinically important. Therefore, the two devices cannot be used interchangeably in clinical practice.

Entities:  

Mesh:

Year:  2007        PMID: 17426923     DOI: 10.1007/s10792-007-9051-z

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.029


  13 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  Comparison of optic disc topography measured by Retinal Thickness Analyzer with measurement by Heidelberg Retina Tomograph II.

Authors:  Noriko Itai; Masaki Tanito; Etsuo Chihara
Journal:  Jpn J Ophthalmol       Date:  2003 Mar-Apr       Impact factor: 2.447

3.  [Correlations between retinal thickness analyzer (RTA) and confocal scanning laser tomography (HRT) in optic disc analysis].

Authors:  J M Martínez de la Casa; J García Feijoó; A Castillo Gómez; J García Sánchez
Journal:  Arch Soc Esp Oftalmol       Date:  2004-01

4.  A new method for rapid mapping of the retinal thickness at the posterior pole.

Authors:  R Zeimer; M Shahidi; M Mori; S Zou; S Asrani
Journal:  Invest Ophthalmol Vis Sci       Date:  1996-09       Impact factor: 4.799

Review 5.  Optic nerve and retinal nerve fiber layer analyzers in glaucoma.

Authors:  David S Greenfield
Journal:  Curr Opin Ophthalmol       Date:  2002-04       Impact factor: 3.761

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Agreement among 3 optical imaging methods for the assessment of optic disc topography.

Authors:  Esther M Hoffmann; Christopher Bowd; Felipe A Medeiros; Catherine Boden; Franz H Grus; Rupert R Bourne; Linda M Zangwill; Robert N Weinreb
Journal:  Ophthalmology       Date:  2005-10-10       Impact factor: 12.079

8.  Optic nerve head analyser and Heidelberg retina tomograph: accuracy and reproducibility of topographic measurements in a model eye and in volunteers.

Authors:  P Janknecht; J Funk
Journal:  Br J Ophthalmol       Date:  1994-10       Impact factor: 4.638

9.  Clinically detectable nerve fiber atrophy precedes the onset of glaucomatous field loss.

Authors:  A Sommer; J Katz; H A Quigley; N R Miller; A L Robin; R C Richter; K A Witt
Journal:  Arch Ophthalmol       Date:  1991-01

10.  Progression of disc and field damage in early glaucoma.

Authors:  T G Zeyen; J Caprioli
Journal:  Arch Ophthalmol       Date:  1993-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.