Development and prospective evaluation of an automated software system for quality control of quantitative 99mTc-MAG3 renal studies.

Russell D Folks1, Ernest V Garcia, Andrew T Taylor.   

Abstract

UNLABELLED: Quantitative nuclear renography has numerous potential sources of error. We previously reported the initial development of a computer software module for comprehensively addressing the issue of quality control (QC) in the analysis of radionuclide renal images. The objective of this study was to prospectively test the QC software.
METHODS: The QC software works in conjunction with standard quantitative renal image analysis using a renal quantification program. The software saves a text file that summarizes QC findings as possible errors in user-entered values, calculated values that may be unreliable because of the patient's clinical condition, and problems relating to acquisition or processing. To test the QC software, a technologist not involved in software development processed 83 consecutive nontransplant clinical studies. The QC findings of the software were then tabulated. QC events were defined as technical (study descriptors that were out of range or were entered and then changed, unusually sized or positioned regions of interest, or missing frames in the dynamic image set) or clinical (calculated functional values judged to be erroneous or unreliable).
RESULTS: Technical QC events were identified in 36 (43%) of 83 studies. Clinical QC events were identified in 37 (45%) of 83 studies. Specific QC events included starting the camera after the bolus had reached the kidney, dose infiltration, oversubtraction of background activity, and missing frames in the dynamic image set.
CONCLUSION: QC software has been developed to automatically verify user input, monitor calculation of renal functional parameters, summarize QC findings, and flag potentially unreliable values for the nuclear medicine physician. Incorporation of automated QC features into commercial or local renal software can reduce errors and improve technologist performance and should improve the efficiency and accuracy of image interpretation.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17337654      PMCID: PMC3205090     

Source DB:  PubMed          Journal:  J Nucl Med Technol        ISSN: 0091-4916


  24 in total

1.  Quality of dynamic radionuclide renal imaging: multicentre evaluation using a functional renal phantom.

Authors:  J O Heikkinen; J T Kuikka; A K Ahonen; P J Rautio
Journal:  Nucl Med Commun       Date:  2001-09       Impact factor: 1.690

2.  UK audit and analysis of quantitative parameters obtained from gamma camera renography.

Authors:  A S Houston; D R Whalley; J V Skrypniuk; P H Jarritt; J S Fleming; P S Cosgriff
Journal:  Nucl Med Commun       Date:  2001-05       Impact factor: 1.690

3.  [Recommendations for carrying out nuclear medicine kidney function tests with a gamma camera and computer].

Authors:  H Bergmann; A Mostbeck; K Kletter; R Nicoletti; M Oberladstätter
Journal:  Acta Med Austriaca Suppl       Date:  1991

4.  Quality assurance in nuclear medicine software and 'COST'.

Authors:  K E Britton; E Busemann Sokole
Journal:  Nucl Med Commun       Date:  1990-04       Impact factor: 1.690

5.  Estimation of residual urine and urine flow rates without urethral catheterization.

Authors:  B S Strauss; M D Blaufox
Journal:  J Nucl Med       Date:  1970-02       Impact factor: 10.057

6.  Diuretic renography with the addition of quantitative gravity-assisted drainage in infants and children.

Authors:  D C Wong; M A Rossleigh; R H Farnsworth
Journal:  J Nucl Med       Date:  2000-06       Impact factor: 10.057

7.  Renal clearance determination with one blood sample: improved accuracy and universal applicability by a new calculation principle.

Authors:  B Bubeck
Journal:  Semin Nucl Med       Date:  1993-01       Impact factor: 4.446

8.  Special considerations in the pediatric use of radionuclides for kidney studies.

Authors:  J M Ash; V F Antico; D L Gilday; S Houle
Journal:  Semin Nucl Med       Date:  1982-10       Impact factor: 4.446

9.  Pelviureteric junction obstruction. The value of a post-micturition view in 99mTc DTPA diuretic renography.

Authors:  I Gordon; R M Mialdea-Fernandez; A M Peters
Journal:  Br J Urol       Date:  1988-05

10.  Increasing the dose of furosemide in patients with azotemia and suspected obstruction.

Authors:  Angela Hunsche; Hejung Press; Andrew Taylor
Journal:  Clin Nucl Med       Date:  2004-03       Impact factor: 7.794

View more
  8 in total

1.  iRENEX: a clinically informed decision support system for the interpretation of ⁹⁹mTc-MAG3 scans to detect renal obstruction.

Authors:  Ernest V Garcia; Andrew Taylor; Russell Folks; Daya Manatunga; Raghuveer Halkar; Bital Savir-Baruch; Eva Dubovsky
Journal:  Eur J Nucl Med Mol Imaging       Date:  2012-05-30       Impact factor: 9.236

2.  SNMMI Procedure Standard/EANM Practice Guideline for Diuretic Renal Scintigraphy in Adults With Suspected Upper Urinary Tract Obstruction 1.0.

Authors:  Andrew T Taylor; David C Brandon; Diego de Palma; M Donald Blaufox; Emmanuel Durand; Belkis Erbas; Sandra F Grant; Andrew J W Hilson; Anni Morsing
Journal:  Semin Nucl Med       Date:  2018-03-16       Impact factor: 4.446

3.  Automated patient motion detection and correction in dynamic renal scintigraphy.

Authors:  Russell D Folks; Daya Manatunga; Ernest V Garcia; Andrew T Taylor
Journal:  J Nucl Med Technol       Date:  2011-05-12

Review 4.  Computer-assisted diagnosis in renal nuclear medicine: rationale, methodology, and interpretative criteria for diuretic renography.

Authors:  Andrew T Taylor; Ernest V Garcia
Journal:  Semin Nucl Med       Date:  2014-03       Impact factor: 4.446

5.  Development of a relational database to capture and merge clinical history with the quantitative results of radionuclide renography.

Authors:  Russell D Folks; Bital Savir-Baruch; Ernest V Garcia; Liudmila Verdes; Andrew T Taylor
Journal:  J Nucl Med Technol       Date:  2012-09-25

6.  Decision support systems in diuresis renography.

Authors:  Andrew Taylor; Amita Manatunga; Ernest V Garcia
Journal:  Semin Nucl Med       Date:  2008-01       Impact factor: 4.446

7.  Diagnostic performance of an expert system for interpretation of 99mTc MAG3 scans in suspected renal obstruction.

Authors:  Andrew Taylor; Ernest V Garcia; Jose Nilo G Binongo; Amita Manatunga; Raghuveer Halkar; Russell D Folks; Eva Dubovsky
Journal:  J Nucl Med       Date:  2008-01-16       Impact factor: 10.057

8.  Estimation of Renal Function Using Unenhanced Computed Tomography in Upper Urinary Tract Stones Patients.

Authors:  Jiali Li; Yang Xun; Cong Li; Yunfeng Han; Yaqi Shen; Xuemei Hu; Daoyu Hu; Zheng Liu; Shaogang Wang; Zhen Li
Journal:  Front Med (Lausanne)       Date:  2020-07-03
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.