Literature DB >> 17328995

Biomechanical evaluation of the modified double-plating fixation for the distal radius fracture.

Hsin-Yi Kathy Cheng1, Chun-Li Lin, Yu-Hao Lin, Alvin Chao-Yu Chen.   

Abstract

BACKGROUND: Distal radius fracture is among the most common type of skeletal injuries. To conquer the surgical and biomechanical complications of the most-frequent used double-plating operation for this fracture, modified double-plating technique was proposed in this study. The aim of this study was to investigate the biomechanical interactions of double-plating, modified double-plating and traditional single plating fixations coupled with various load conditions using nonlinear finite element analysis.
METHODS: A three-dimensional finite element distal radius fracture model with three fixation methods (double-plating, modified double-plating and single) was generated based on computer tomography data. After model verification and validation, frictional (contact) elements were used to simulate the interface condition between the fixation plates and the bony surface. The rigidity, stress values and displacements at the radius end were observed under axial, bending and torsion load conditions.
FINDINGS: The simulated results showed that the modified double-plating model demonstrated the highest rigidity and the least displacement among the three techniques in bending, but not in axial compression (similar results across the three) and torsion (modified double-plating technique possessed lowest rigidity). The maximum von Mises stress for bone was lower in modified double-plating model as well. These results indicated that modified double-plating technique demonstrated a better structural strength against bending with the least potential of fracture fragments and screw loosening.
INTERPRETATION: Although a lower torsional rigidity, modified double-plating technique was a better choice in distal radius fracture fixation since the bending force, which has the potential to separate the fracture ends, is more detrimental in hindering fracture healing.

Entities:  

Mesh:

Year:  2007        PMID: 17328995     DOI: 10.1016/j.clinbiomech.2006.12.010

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  12 in total

1.  Biomechanical study of different plate configurations for distal humerus osteosynthesis.

Authors:  M Bogataj; F Kosel; R Norris; M Krkovic; M Brojan
Journal:  Med Biol Eng Comput       Date:  2015-02-08       Impact factor: 2.602

2.  Interfragmentary lag screw and locking plate combination in simple distal femoral fractures: A finite element analysis.

Authors:  Jun Zhang; Yan Wei; Guoding Li; Jian Wang; Youjia Xu
Journal:  Acta Orthop Traumatol Turc       Date:  2021-01       Impact factor: 1.511

3.  Biomechanical analysis of the wrist arthroplasty in rheumatoid arthritis: a finite element analysis.

Authors:  M N Bajuri; Mohammed Rafiq Abdul Kadir; Malliga Raman Murali; T Kamarul
Journal:  Med Biol Eng Comput       Date:  2012-11-03       Impact factor: 2.602

4.  Finite element analysis of the wrist in stroke patients: the effects of hand grip.

Authors:  Muhammad Hanif Ramlee; Gan Kok Beng; Nazri Bajuri; Mohammed Rafiq Abdul Kadir
Journal:  Med Biol Eng Comput       Date:  2017-12-05       Impact factor: 2.602

5.  Validation of radiocarpal joint contact models based on images from a clinical MRI scanner.

Authors:  Joshua E Johnson; Terence E McIff; Phil Lee; E Bruce Toby; Kenneth J Fischer
Journal:  Comput Methods Biomech Biomed Engin       Date:  2012-05-25       Impact factor: 1.763

6.  Validation of a new multiscale finite element analysis approach at the distal radius.

Authors:  Joshua E Johnson; Karen L Troy
Journal:  Med Eng Phys       Date:  2017-03-31       Impact factor: 2.242

7.  Biomechanical comparison of different volar fracture fixation plates for distal radius fractures.

Authors:  Kareem Sobky; Todd Baldini; Kenneth Thomas; Joel Bach; Allison Williams; Jennifer Moriatis Wolf
Journal:  Hand (N Y)       Date:  2007-09-07

8.  Finite element analysis of osteosynthesis screw fixation in the bone stock: an appropriate method for automatic screw modelling.

Authors:  Jan Wieding; Robert Souffrant; Andreas Fritsche; Wolfram Mittelmeier; Rainer Bader
Journal:  PLoS One       Date:  2012-03-28       Impact factor: 3.240

9.  Finite element analysis of different locking plate fixation methods for the treatment of ulnar head fracture.

Authors:  Yue Zhang; Qin Shao; Chensong Yang; Changqing Ai; Di Zhou; Yang Yu; Guixin Sun
Journal:  J Orthop Surg Res       Date:  2021-03-15       Impact factor: 2.359

10.  Biomechanical comparison of osteoporotic distal radius fractures fixed by distal locking screws with different length.

Authors:  Xiong Liu; Wei-dong Wu; Ya-feng Fang; Mei-chao Zhang; Wen-hua Huang
Journal:  PLoS One       Date:  2014-07-31       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.