Comparison of conventional versus computer-navigated acetabular component insertion.

Rolf G A Haaker1, Kai Tiedjen, Andreas Ottersbach, Frank Rubenthaler, Marthin Stockheim, James B Stiehl.   

Abstract

This retrospective study compared the efficacy of computer navigation and conventional freehand techniques to place acetabular component orientation in the target position of acetabular cup inclination of 45 degrees and anteversion of 20 degrees . We selected 69 patients who had undergone total hip arthroplasty with freehand cup insertion who had computed tomography (CT) to plan for acetabular cup placement of the contralateral side. This group was compared with 98 patients who underwent CT-based cup insertion, and all had postoperative CT. After CT-based cup placement, average cup position was 43 degrees inclination (95% confidence interval [CI], 0.97; range, 30 degrees -58 degrees ) and 22.2 degrees anteversion (95% CI, 1.72; range, 5 degrees -38 degrees ). For freehand, average cup position was 45.7 degrees inclination (95% CI, 2.63 degrees ; range, 26 degrees -64 degrees ) and 28.5 degrees anteversion (95% CI, 3.80 degrees ; range, 9 degrees -53 degrees ). F ratio was 5.56 for inclination and 3.67 for anteversion (P < .0001). This study demonstrated substantial statistical improvement in accuracy of cup placement using CT-based navigation compared with freehand methods.

Entities:  

Mesh:

Year:  2007        PMID: 17275626     DOI: 10.1016/j.arth.2005.10.018

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  27 in total

1.  Intra- and extra-articular planes of reference for use in total hip arthroplasty: a preliminary study.

Authors:  Jerome Hausselle; Pierre Etienne Moreau; Loic Wessely; Emmanuel de Thomasson; Ayman Assi; Sebastien Parratte; Jerome Essig; Wafa Skalli
Journal:  Int Orthop       Date:  2012-03-21       Impact factor: 3.075

2.  Intraoperative radiographs for placing acetabular components in hip resurfacing arthroplasty.

Authors:  Thomas P Gross; Fei Liu; Lee Webb
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

3.  Acetabular component positioning using anatomic landmarks of the acetabulum.

Authors:  Yong-Chan Ha; Jeong Joon Yoo; Young-Kyun Lee; Jin Young Kim; Kyung-Hoi Koo
Journal:  Clin Orthop Relat Res       Date:  2012-07-10       Impact factor: 4.176

4.  Anteroposterior Radiographs Are More Accurate than Cross-Table Lateral Radiographs for Acetabular Anteversion Assessment: a Retrospective Cohort Study.

Authors:  Peter K Sculco; Alexander S McLawhorn; Kaitlin M Carroll; Benjamin A McArthur; David J Mayman
Journal:  HSS J       Date:  2015-10-19

5.  Comparison of robotic-assisted and conventional acetabular cup placement in THA: a matched-pair controlled study.

Authors:  Benjamin G Domb; Youssef F El Bitar; Adam Y Sadik; Christine E Stake; Itamar B Botser
Journal:  Clin Orthop Relat Res       Date:  2013-08-29       Impact factor: 4.176

6.  Navigated non-image-based positioning of the acetabulum during total hip replacement.

Authors:  Jean-Yves Jenny; Cyril Boeri; Jean-Claude Dosch; Marius Uscatu; Eugen Ciobanu
Journal:  Int Orthop       Date:  2007-11-13       Impact factor: 3.075

7.  Six sigma analysis of minimally invasive acetabular arthroplasty: a preliminary investigation.

Authors:  David A Heck; James B Stiehl
Journal:  Clin Orthop Relat Res       Date:  2009-05-22       Impact factor: 4.176

8.  Intra- and intersurgeon variability in image-free navigation system for THA.

Authors:  Hirotsugu Ohashi; Masanori Matsuura; Yusaku Okamoto; Yoshiaki Okajima
Journal:  Clin Orthop Relat Res       Date:  2009-04-21       Impact factor: 4.176

9.  Component alignment in hip resurfacing using computer navigation.

Authors:  Chris Bailey; Rehan Gul; Mark Falworth; Steven Zadow; Roger Oakeshott
Journal:  Clin Orthop Relat Res       Date:  2008-10-30       Impact factor: 4.176

10.  Does computer-assisted surgery benefit leg length restoration in total hip replacement? Navigation versus conventional freehand.

Authors:  Alfonso Manzotti; Pietro Cerveri; Elena De Momi; Chris Pullen; Norberto Confalonieri
Journal:  Int Orthop       Date:  2009-11-11       Impact factor: 3.075

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.