Least detectable concentration and dynamic range of three immunoassay systems using the same antibody.

Thomas R Glass1, Naoya Ohmura, Hiroshi Saiki.   

Abstract

The least detectable concentration (LDC) and dynamic range (DR) of three immunoassay systems are compared using four distinct antibodies (all specific for estradiol but with different affinities) on each system. The systems evaluated include the industry standard, ELISA, and two biosensors, surface plasmon resonance and kinetic exclusion. In all cases, the measurements of inhibition curves (response vs estradiol concentration) were contracted to outside experts (the biosensor manufacturers themselves in the case of the biosensors), each of whom was supplied with the same blind samples. Each biosensor manufacturer also reported an estimate of the equilibrium dissociation constant (Kd) for each of the antibodies. The LDC and DR observed for the kinetic exclusion biosensor are consistent with an interpretation of Kd limited detection while that from the other biosensor and ELISA show limits of detection somewhat above those expected for Kd limited performance. The determined LDC and DR of each biosensor is self-consistent in the sense that none of the inhibition data contradicts theoretical limits associated with the Kd as measured on that system; however, some contradictions are apparent across platforms. The use of multiple antibodies of differing Kd improves confidence that the observed differences in performance are associated with the immunoassay system rather than the particular analyte.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17256970     DOI: 10.1021/ac061288z

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  4 in total

1.  Uncertainty estimates of purity measurements based on current information: toward a "live validation" of purity methods.

Authors:  Izydor Apostol; Drew Kelner; Xinzhao Grace Jiang; Gang Huang; Jette Wypych; Xin Zhang; Jessica Gastwirt; Kenneth Chen; Szilan Fodor; Suminda Hapuarachchi; Dave Meriage; Frank Ye; Leszek Poppe; Wojciech Szpankowski
Journal:  Pharm Res       Date:  2012-08-15       Impact factor: 4.200

2.  Theoretical limitations of quantification for noncompetitive sandwich immunoassays.

Authors:  Christine F Woolley; Mark A Hayes; Prasun Mahanti; S Douglass Gilman; Tom Taylor
Journal:  Anal Bioanal Chem       Date:  2015-09-05       Impact factor: 4.142

3.  Kinetic Exclusion Assay of Biomolecules by Aptamer Capture.

Authors:  Mark H Smith; Daniel Fologea
Journal:  Sensors (Basel)       Date:  2020-06-18       Impact factor: 3.576

4.  Influence of particle size on the SARS-CoV-2 spike protein detection using IgG-capped gold nanoparticles and dynamic light scattering.

Authors:  C B P Ligiero; T S Fernandes; D L D'Amato; F V Gaspar; P S Duarte; M A Strauch; J G Fonseca; L G R Meirelles; P Bento da Silva; R B Azevedo; G Aparecida de Souza Martins; B S Archanjo; C D Buarque; G Machado; A M Percebom; C M Ronconi
Journal:  Mater Today Chem       Date:  2022-04-22
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.