Pulmonary nodule volumetric measurement variability as a function of CT slice thickness and nodule morphology.

Myria Petrou1, Leslie E Quint, Bin Nan, Laurence H Baker.   

Abstract

OBJECTIVE: The purpose of our study was to assess differences in volumetric measurements of pulmonary nodules obtained using different CT slice thicknesses; correlate these differences with nodule size, shape, and margination; and compare measurements generated by two different software packages.
MATERIALS AND METHODS: Seventy-five individual nodules identified on 29 lowdose, unenhanced, MDCT chest examinations were selected for volumetric analysis. Each image data set was reconstructed in three ways (slice thickness/reconstruction interval): 1.25 mm/0.625 mm, 2.5 mm/2 mm, and 5 mm/2.5 mm. Volumetric measurements were made on all 75 nodules at 1.25- and 2.5-mm slice thicknesses and on 57 of 75 nodules at the 5-mm thickness using Volume Analysis software. For 69 of 75 nodules, measurements were obtained on 1.25- and 2.5-mm-thick sections using a different commercially available software system, LN500 R2 software. Volume variability between different slice thicknesses was correlated with nodule diameter, shape, and margination using multiple linear regression. Percent differences between measurements obtained with the two software systems were calculated. Significance of relative volume differences between slice thicknesses and software packages was assessed using a one-sample Student's t-test.
RESULTS: Although statistically significant differences in volumes between different section thicknesses were seen only for the tiny nodule size group, many individual nodules showed substantial volume variation. Significant differences were seen in nodule volume variability for smaller nodules (3-10 mm) compared with larger nodules (> or = 11 mm) (p < 0.0001), as well as spiculated compared with smooth nodules, within a single size group (p < 0.05). No effect of nodule shape (round vs elongated) was noted. Statistically significant differences in measurements obtained with the two software systems were seen only with 2.5-mm-thick sections (p = 0.001).
CONCLUSION: CT slice thickness variation resulted in significant differences in volume measurements for tiny nodules. A spiculated margin was shown to have a significant effect on nodule volume variability within a single size group. Use of different software packages resulted in significant volume measurement differences at the 2.5-mm CT slice thickness.

Mesh:

Year:  2007        PMID: 17242235     DOI: 10.2214/AJR.05.1063

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  48 in total

1.  Zone of transition: a potential source of error in tumor volume estimation.

Authors:  Lijuan Zhang; David F Yankelevitz; Claudia I Henschke; Artit C Jirapatnakul; Anthony P Reeves; Darryl Carter
Journal:  Radiology       Date:  2010-08       Impact factor: 11.105

2.  Adrenal gland volume measurement in septic shock and control patients: a pilot study.

Authors:  Stephanie Nougaret; B Jung; S Aufort; G Chanques; S Jaber; B Gallix
Journal:  Eur Radiol       Date:  2010-06-04       Impact factor: 5.315

3.  CT tumor volume measurement in advanced non-small-cell lung cancer: Performance characteristics of an emerging clinical tool.

Authors:  Mizuki Nishino; Mengye Guo; David M Jackman; Pamela J DiPiro; Jeffrey T Yap; Tak K Ho; Hiroto Hatabu; Pasi A Jänne; Annick D Van den Abbeele; Bruce E Johnson
Journal:  Acad Radiol       Date:  2010-10-30       Impact factor: 3.173

Review 4.  Noncalcified lung nodules: volumetric assessment with thoracic CT.

Authors:  Marios A Gavrielides; Lisa M Kinnard; Kyle J Myers; Nicholas Petrick
Journal:  Radiology       Date:  2009-04       Impact factor: 11.105

5.  Imprecision in automated volume measurements of pulmonary nodules and its effect on the level of uncertainty in volume doubling time estimation.

Authors:  Paul J Nietert; James G Ravenel; William M Leue; James V Miller; Katherine K Taylor; Elizabeth S Garrett-Mayer; Gerard A Silvestri
Journal:  Chest       Date:  2009-01-13       Impact factor: 9.410

6.  Three-dimensional analysis of pulmonary nodules: variability of semiautomated volume measurements between different versions of the same software.

Authors:  M F Rinaldi; T Bartalena; L Braccaioli; N Sverzellati; S Mattioli; E Rimondi; G Rossi; M Zompatori; G Battista; R Canini
Journal:  Radiol Med       Date:  2010-01-15       Impact factor: 3.469

7.  Pulmonary nodule volume: effects of reconstruction parameters on automated measurements--a phantom study.

Authors:  James G Ravenel; William M Leue; Paul J Nietert; James V Miller; Katherine K Taylor; Gerard A Silvestri
Journal:  Radiology       Date:  2008-05       Impact factor: 11.105

8.  Exploring Variability in CT Characterization of Tumors: A Preliminary Phantom Study.

Authors:  Binsheng Zhao; Yongqiang Tan; Wei Yann Tsai; Lawrence H Schwartz; Lin Lu
Journal:  Transl Oncol       Date:  2014-02-01       Impact factor: 4.243

9.  Computed tomography assessment of response to therapy: tumor volume change measurement, truth data, and error.

Authors:  Michael F McNitt-Gray; Luc M Bidaut; Samuel G Armato; Charles R Meyer; Marios A Gavrielides; Charles Fenimore; Geoffrey McLennan; Nicholas Petrick; Binsheng Zhao; Anthony P Reeves; Reinhard Beichel; Hyun-Jung Grace Kim; Lisa Kinnard
Journal:  Transl Oncol       Date:  2009-12       Impact factor: 4.243

10.  Optimal image reconstruction for detection and characterization of small pulmonary nodules during low-dose CT.

Authors:  SayedMasoud Hashemi; Hatem Mehrez; Richard S C Cobbold; Narinder S Paul
Journal:  Eur Radiol       Date:  2014-03-22       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.