Literature DB >> 17205566

Stress analysis of the anterior tibial post in posterior stabilized knee prostheses.

Chang-Hung Huang1, Jiann-Jong Liau, Chun-Hsiung Huang, Cheng-Kung Cheng.   

Abstract

Recent retrieval studies have indicated a high incidence of polyethylene wear on the anterior tibial post caused by impingement. This study investigated the influences of post-cam design features and component alignment on the stress distribution in the anterior tibial post when subjected to the impingement loading. Two three-dimensional finite element models of posterior stabilized knee prostheses were constructed, one with flat on flat (FF) and another with curve on curve (CC) contact surfaces between anterior tibial post and femoral cam. The polyethylene insert was modeled with elastoplastic properties. Nine cases, three hyperextension angles (0 degrees , 5 degrees , and 10 degrees ) combined with three axial tibial rotations (0 degrees , 2.5 degrees , and 5 degrees ) simulating different component alignments were analyzed. A vertical compressive load of 2,000 N and an extension moment of 45 Nm were applied simultaneously. The FF model had larger stress increases than the CC model in both hyperextension and tibial rotation compared with the neutral position. The maximum increase for the FF model was 68% in peak contact stress, 125% in von Mises stress, and 58% in tensile stress in the extreme case of 10 degrees of hyperextension combined with 5 degrees of axial rotation. Stress concentration was found at the anterior corner of the post in the FF model; this was not found in the CC model. The curve on curve design can reduce edge loading on the tibial post, especially during axial tibiofemoral rotation.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17205566     DOI: 10.1002/jor.20336

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  7 in total

1.  The effect of posterior tibial slope on knee flexion in posterior-stabilized total knee arthroplasty.

Authors:  Xiaojun Shi; Bin Shen; Pengde Kang; Jing Yang; Zongke Zhou; Fuxing Pei
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-06-02       Impact factor: 4.342

2.  Implant design influences tibial post wear damage in posterior-stabilized knees.

Authors:  Mark M Dolan; Natalie H Kelly; Joseph T Nguyen; Timothy M Wright; Steven B Haas
Journal:  Clin Orthop Relat Res       Date:  2011-01       Impact factor: 4.176

3.  Tibial post wear in posterior-stabilized knee replacements is design-dependent.

Authors:  Bridgette D Furman; Joseph Lipman; Mordechai Kligman; Timothy M Wright; Steven B Haas
Journal:  Clin Orthop Relat Res       Date:  2008-08-14       Impact factor: 4.176

4.  Stress distribution of the patellofemoral joint in the anatomic V-shape and curved dome-shape femoral component: a comparison of resurfaced and unresurfaced patellae.

Authors:  Chang-Hung Huang; Lin-I Hsu; Ting-Kuo Chang; Tai-Yuan Chuang; Shih-Liang Shih; Yung-Chang Lu; Chen-Sheng Chen; Chun-Hsiung Huang
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-12-25       Impact factor: 4.342

5.  Effect of Post-Cam Design for Normal Knee Joint Kinematic, Ligament, and Quadriceps Force in Patient-Specific Posterior-Stabilized Total Knee Arthroplasty by Using Finite Element Analysis.

Authors:  Yong-Gon Koh; Juhyun Son; Oh-Ryong Kwon; Sae Kwang Kwon; Kyoung-Tak Kang
Journal:  Biomed Res Int       Date:  2018-09-19       Impact factor: 3.411

6.  Mechanics of post-cam engagement during simulated dynamic activity.

Authors:  Clare K Fitzpatrick; Chadd W Clary; Adam J Cyr; Lorin P Maletsky; Paul J Rullkoetter
Journal:  J Orthop Res       Date:  2013-04-19       Impact factor: 3.494

Review 7.  Biomechanical considerations in the design of high-flexion total knee replacements.

Authors:  Cheng-Kung Cheng; Colin J McClean; Yu-Shu Lai; Wen-Chuan Chen; Chang-Hung Huang; Kun-Jhih Lin; Chia-Ming Chang
Journal:  ScientificWorldJournal       Date:  2014-05-06
  7 in total