Incidence and clinical impact of coronary stent fracture after sirolimus-eluting stent implantation.

Jiro Aoki1, Gaku Nakazawa, Kengo Tanabe, Angela Hoye, Hirosada Yamamoto, Tomohiro Nakayama, Yoshinobu Onuma, Yasutomi Higashikuni, Syuji Otsuki, Atsuhiko Yagishita, Sen Yachi, Hiroyoshi Nakajima, Kazuhiro Hara.   

Abstract

BACKGROUND: Stent fracture is one of the possible causes of restenosis after sirolimus-eluting stents (SES) implantation. The aim of our study was to evaluate the prevalence and clinical impact of coronary stent fracture after SES implantation.
METHODS: From our prospective institutional database, 280 patients were treated solely with SES from August 2004 to June 2005. Among the 280 patients, 256 patients with a total of 307 lesions underwent follow-up angiography on an average of 240 days after the procedure.
RESULTS: Stent fractures were observed in eight (2.6%) lesions. Of the eight lesions with stent fracture, five were located in the right coronary artery (RCA), two in the saphenous vein (SV) graft, and one in the left anterior descending coronary artery. The stent fractures were all in the locations that served as hinges during vessel movement in the cardiac contraction cycle. Seven of the eight stent fractures were adjacent to the edge of previously implanted or overlapped stent. Significant multivariate predictors of stent fracture were SV graft location (Odds ratio 35.88; 95% confidence interval 2.73-471.6, P = 0.006), implanted stent length (Odds ratio 1.04; 95% confidence interval 1.01-1.07, P = 0.02), and RCA location (Odds ratio 10.00; 95% confidence interval 1.11-89.67, P = 0.04). In-stent binary restenosis rate was 37.5% and target lesion repeat revascularization rate was 50.0% in patients with stent fracture.
CONCLUSIONS: Stent fracture was likely to be affected by mechanical stress provoked by rigid structures and hinge points. Stent fracture might be associated with the high incidence of target lesion revascularization. (c) 2006 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17195199     DOI: 10.1002/ccd.20950

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  33 in total

1.  Long-term angiographic outcomes of post-sirolimus-eluting stent restenosis in Japanese patients.

Authors:  Nobuyuki Miyai; Noriyuki Kinoshita; Keisuke Oota; Takeshi Yamada; Reo Nakamura; Hidekazu Irie; Tetsuo Hashimoto; Shyunichi Tamaki; Hiroaki Matsubara
Journal:  Heart Vessels       Date:  2010-10-27       Impact factor: 2.037

2.  Prevalence and prognosis of coronary stent gap detected by multi-detector CT: a follow-up study.

Authors:  Xinghua Zhang; Li Yang; Haiyue Ju; Fan Zhang; Jian Wu; Bin He; Yundai Chen
Journal:  Eur Radiol       Date:  2012-04-27       Impact factor: 5.315

Review 3.  Are we aware of stent fracture?

Authors:  R D Acar; M Bulut; M Akcakoyun
Journal:  Herz       Date:  2013-10-25       Impact factor: 1.443

4.  Coronary in-stent restenosis: predisposing clinical and stent-related factors, diagnostic performance and analyses of inaccuracies in 320-row computed tomography angiography.

Authors:  Yung-Liang Wan; Pei-Kwei Tsay; Chun-Chi Chen; Yu-Hsiang Juan; Yu-Chieh Huang; Wen-Hui Chan; Ming-Shien Wen; I-Chang Hsieh
Journal:  Int J Cardiovasc Imaging       Date:  2016-04-22       Impact factor: 2.357

5.  Impact of sirolimus-eluting stent fractures without early cardiac events on long-term clinical outcomes: a multislice computed tomography study.

Authors:  Tsuyoshi Ito; Masashi Kimura; Mariko Ehara; Mitsuyasu Terashima; Kenya Nasu; Yoshihisa Kinoshita; Maoto Habara; Etsuo Tsuchikane; Takahiko Suzuki
Journal:  Eur Radiol       Date:  2014-02-25       Impact factor: 5.315

6.  Impact of the bifurcation angle on major cardiac events after cross-over single stent strategy in unprotected left main bifurcation lesions: 3-dimensional quantitative coronary angiographic analysis.

Authors:  Kisaki Amemiya; Takenori Domei; Masashi Iwabuchi; Shinichi Shirai; Kenji Ando; Masahiko Goya; Hiroyoshi Yokoi; Masakiyo Nobuyoshi
Journal:  Am J Cardiovasc Dis       Date:  2014-12-29

7.  Validating Fatigue Safety Factor Calculation Methods for Cardiovascular Stents.

Authors: 
Journal:  J Biomech Eng       Date:  2018-06-01       Impact factor: 2.097

8.  Complete fracture of sirolimus-eluting stent in a saphenous vein graft to left anterior descending artery.

Authors:  Sun Hong Yoo; Seung Won Jin; Sung Ho Her; Hee Jeoung Yoon; Hyoung Doo Kim; Yun Sun Im; Ki Bae Seung; Jae Hyung Kim
Journal:  Korean Circ J       Date:  2009-06-30       Impact factor: 3.243

9.  Late thrombosis of sirolimus-eluting stent: a multifactorial problem.

Authors:  Igor Kranjec; Andreja Cerne
Journal:  Case Rep Med       Date:  2009-11-04

10.  Therapeutic strategy for in-stent restenosis based on the restenosis pattern after drug-eluting stent implantation.

Authors:  Ki-Hun Kim; Doo-Il Kim; Il-Hwan Kim; Jong-Yoon Kim; Yang-Chun Han; Sang-Hoon Seol; Ung Kim; Tae-Hyun Yang; Dae-Kyeong Kim; Dong-Soo Kim; Sang-Hee Lee; Jong-Seon Park; Young-Jo Kim; Dong-Gu Shin; Yoon-Kyung Cho; Chang-Wook Nam; Seung-Ho Hur; Kwon-Bae Kim
Journal:  Korean Circ J       Date:  2009-10-28       Impact factor: 3.243

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.