Psychophysical evaluation of calibration curve for diagnostic LCD monitor.

Masanobu Uemura1, Yoshiyuki Asai, Michihiro Yamaguchi, Hideki Fujita, Yuuko Shintani, Shigeru Sanada.   

Abstract

PURPOSE: In 1998, Digital Imaging Communications in Medicine (DICOM) proposed a calibration tool, the grayscale standard display function (GSDF), to obtain output consistency of radiographs. To our knowledge, there have been no previous reports of investigating the relation between perceptual linearity and detectability on a calibration curve.
MATERIALS AND METHODS: To determine a suitable calibration curve for diagnostic liquid crystal display (LCD) monitors, the GSDF and Commission Internationale de l'Eclairage (CIE) curves were compared using psychophysical gradient delta and receiver operating characteristic (ROC) analysis for clinical images.
RESULTS: We succeeded in expressing visually recognized contrast directly using delta instead of the just noticeable difference (JND) index of the DICOM standard. As a result, we found that the visually recognized contrast at low luminance areas on the LCD monitor calibrated by the CIE curve is higher than that calibrated by the GSDF curve. On the ROC analysis, there was no significant difference in tumor detectability between GSDF and CIE curves for clinical thoracic images. However, the area parameter Az of the CIE curve is superior to that of the GSDF curve. The detectability of tumor shadows in the thoracic region on clinical images using the CIE curve was superior to that using the GSDF curve owing to the high absolute value of delta in the low luminance range.
CONCLUSION: We conclude that the CIE curve is the most suitable tool for calibrating diagnostic LCD monitors, rather than the GSDF curve.

Entities:  

Mesh:

Year:  2006        PMID: 17186319     DOI: 10.1007/s11604-006-0085-3

Source DB:  PubMed          Journal:  Radiat Med        ISSN: 0288-2043


  10 in total

1.  Development of a digital image database for chest radiographs with and without a lung nodule: receiver operating characteristic analysis of radiologists' detection of pulmonary nodules.

Authors:  J Shiraishi; S Katsuragawa; J Ikezoe; T Matsumoto; T Kobayashi; K Komatsu; M Matsui; H Fujita; Y Kodera; K Doi
Journal:  AJR Am J Roentgenol       Date:  2000-01       Impact factor: 3.959

2.  Development and evaluation of a new gray-scale test pattern to adjust gradients of thoracic CT imaging.

Authors:  M Yamaguchi; H Fujita; M Uemura; Y Asai; H Wakae; M Ishifuro
Journal:  Eur Radiol       Date:  2004-04-02       Impact factor: 5.315

3.  Psychophysical analysis of monitor display functions affecting observer diagnostic performance of CT image on liquid crystal display monitors.

Authors:  M Yamaguchi; H Fujita; Y Asai; M Uemura; Y Ookura; M Matsumoto; T Johkoh
Journal:  Eur Radiol       Date:  2005-08-16       Impact factor: 5.315

4.  Evaluation of grey-scale standard display function as a calibration tool for diagnostic liquid crystal display monitors using psychophysical analysis.

Authors:  Y Asai; Y Shintani; M Yamaguchi; M Uemura; M Matsumoto; H Kanamori
Journal:  Med Biol Eng Comput       Date:  2005-05       Impact factor: 2.602

5.  Comparison of gradation characteristics between three typical medical imaging display devices using psychophysical analysis.

Authors:  Masanobu Uemura; Yoshiyuki Asai; Michihiro Yamaguchi; Yuko Shintani; Sigeru Sanada
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2005-12-20

Review 6.  ROC methodology in radiologic imaging.

Authors:  C E Metz
Journal:  Invest Radiol       Date:  1986-09       Impact factor: 6.016

7.  Some practical issues of experimental design and data analysis in radiological ROC studies.

Authors:  C E Metz
Journal:  Invest Radiol       Date:  1989-03       Impact factor: 6.016

8.  The meaning and use of the area under a receiver operating characteristic (ROC) curve.

Authors:  J A Hanley; B J McNeil
Journal:  Radiology       Date:  1982-04       Impact factor: 11.105

9.  Determination of optimum film density range for roentgenograms from visual effects.

Authors:  H Kanamori
Journal:  Acta Radiol Diagn (Stockh)       Date:  1966-07

10.  Spatial resolution by the human visual system. The effect of mean retinal illuminance.

Authors:  A S Patel
Journal:  J Opt Soc Am       Date:  1966-05
  10 in total
  2 in total

1.  Investigation of optimal viewing size for detecting nodular ground-glass opacity on high-resolution computed tomography with cine-mode display.

Authors:  Michihiro Yamaguchi; Yuichi Bessho; Tatsuro Inoue; Yoshiyuki Asai; Tomoshige Matsumoto; Kenya Murase
Journal:  Radiol Phys Technol       Date:  2010-08-19

2.  Comparison of the commercial color LCD and the medical monochrome LCD using randomized object test patterns.

Authors:  Jay Wu; Tung H Wu; Rou P Han; Shu J Chang; Cheng T Shih; Jing Y Sun; Shih M Hsu
Journal:  PLoS One       Date:  2012-05-31       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.