Multi-detector row CT attenuation measurements: assessment of intra- and interscanner variability with an anthropomorphic body CT phantom.

Bernard A Birnbaum1, Nicole Hindman, Julie Lee, James S Babb.   

Abstract

PURPOSE: To determine the dependence of absolute computed tomographic (CT) attenuation values on multi-detector row CT scanner type, convolution kernel, and tube current by using an anthropomorphic phantom.
MATERIALS AND METHODS: A customized phantom was designed with tissue-equivalent materials to simulate contrast material-enhanced liver, spleen, pancreas, aorta, kidney, 0- and 50-HU cylindric renal cysts, muscle, and fat. The phantom was scanned with five multi-detector row CT scanners (LightSpeed QXi, GE Healthcare, Milwaukee, Wis; MX8000, Philips Medical Systems, Best, the Netherlands; and Volume Zoom, Sensation 16 and Sensation 64, Siemens Medical Solutions, Forchheim, Germany) on five separate occasions with 120 kVp, low and high tube current settings, 3.00-3.75-mm section thickness, 50% overlap, and standard and high-spatial-resolution kernels. Standardized regions of interest (ROIs) were used to obtain 3510 attenuation measurements. Attenuation dependence on scanner, kernel, and tube current was evaluated by using F tests derived with mixed-model regression. Within the mixed-model framework, the Tukey honestly significant difference procedure and a Bonferroni multiple comparison correction were used to assess differences among imaging regimens and tube current settings, respectively, in terms of tissue attenuation and ROI standard deviation.
RESULTS: Tube current had no significant effect (P>.4) on observed tissue attenuation. Significant (P<.0001) differences were observed between imaging regimens with respect to mean attenuation for each tissue type. Convolution kernel modification had an inconsistent effect on tissue attenuation, depending on the scanner. All multi-detector row CT scanners displayed intrascanner variability in tissue attenuation (minimum range: 8.4 HU for fat tissue with the Sensation 16; maximum range: 63.4 HU for liver tissue with the Sensation 64). The scanners behaved differently at the lower range of the CT number scale, where 0-HU cyst attenuation ranged from -15.7 to 23.9 HU and one vendor's equipment showed significantly lower mean attenuation values.
CONCLUSION: CT attenuation values vary significantly between different manufacturers' multi-detector row CT scanners, among different generations of multi-detector row CT scanning equipment, and with individual combinations of scanner and convolution kernel. SUPPLEMENTAL MATERIAL: http://radiology.rsnajnls.org/cgi/content/full/242/1/109/DC1 Copyright (c) RSNA, 2007.

Entities:  

Mesh:

Year:  2007        PMID: 17185663     DOI: 10.1148/radiol.2421052066

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  68 in total

1.  Precision of the measurement of CT numbers: comparison of dual-energy CT spectral imaging with fast kVp switching and conventional CT with phantoms.

Authors:  Izuru Matsuda; Masaaki Akahane; Jiro Sato; Masaki Katsura; Shigeru Kiryu; Naoki Yoshioka; Akira Kunimatsu; Kenji Ino; Kuni Ohtomo
Journal:  Jpn J Radiol       Date:  2011-12-02       Impact factor: 2.374

Review 2.  Quantitative Assessment of Liver Fat with Magnetic Resonance Imaging and Spectroscopy.

Authors:  Scott B Reeder; Irene Cruite; Gavin Hamilton; Claude B Sirlin
Journal:  J Magn Reson Imaging       Date:  2011-09-16       Impact factor: 4.813

3.  Deep learning-based image restoration algorithm for coronary CT angiography.

Authors:  Fuminari Tatsugami; Toru Higaki; Yuko Nakamura; Zhou Yu; Jian Zhou; Yujie Lu; Chikako Fujioka; Toshiro Kitagawa; Yasuki Kihara; Makoto Iida; Kazuo Awai
Journal:  Eur Radiol       Date:  2019-04-08       Impact factor: 5.315

4.  Diagnostic performance of dual-energy CT and subtraction CT for renal lesion detection and characterization.

Authors:  Ali Pourvaziri; Anushri Parakh; Amirkasra Mojtahed; Avinash Kambadakone; Dushyant Vasudeo Sahani
Journal:  Eur Radiol       Date:  2019-05-27       Impact factor: 5.315

Review 5.  Noncalcified lung nodules: volumetric assessment with thoracic CT.

Authors:  Marios A Gavrielides; Lisa M Kinnard; Kyle J Myers; Nicholas Petrick
Journal:  Radiology       Date:  2009-04       Impact factor: 11.105

Review 6.  Dual energy MDCT assessment of renal lesions: an overview.

Authors:  Achille Mileto; Daniele Marin; Rendon C Nelson; Giorgio Ascenti; Daniel T Boll
Journal:  Eur Radiol       Date:  2013-10-04       Impact factor: 5.315

Review 7.  [Qualitative and quantitative CT analysis of acute pulmonary failure].

Authors:  A W Reske; M Seiwerts
Journal:  Radiologe       Date:  2009-08       Impact factor: 0.635

8.  ACCURATUM: improved calcium volume scoring using a mesh-based algorithm--a phantom study.

Authors:  Stefan C Saur; Hatem Alkadhi; Lotus Desbiolles; Gábor Székely; Philippe C Cattin
Journal:  Eur Radiol       Date:  2008-09-26       Impact factor: 5.315

9.  Validation of the use of calibration factors between the iodine concentration and the computed tomography number measured outside the objects for estimation of iodine concentration inside the objects: phantom experiment.

Authors:  Kentaro Takanami; Shuichi Higano; Kei Takase; Tomohiro Kaneta; Takayuki Yamada; Hiroki Ishiya; Issei Mori; Shoki Takahashi
Journal:  Radiat Med       Date:  2008-05-29

10.  Rat lumbar vertebrae bone densitometry using multidetector CT.

Authors:  Yi-Xiang J Wang; James F Griffith; Hua Zhou; Kai Chow Choi; Vivian W Y Hung; David K W Yeung; Ling Qin; Anil T Ahuja
Journal:  Eur Radiol       Date:  2008-10-29       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.