Evaluation of performance across the dynamic range of the Abbott RealTime HIV-1 assay as compared to VERSANT HIV-1 RNA 3.0 and AMPLICOR HIV-1 MONITOR v1.5 using serial dilutions of 39 group M and O viruses.

Priscilla Swanson1, Shihai Huang, Klara Abravaya, Carmen de Mendoza, Vincent Soriano, Sushil G Devare, John Hackett.   

Abstract

Performance of the Abbott m2000 instrument system and the Abbott RealTime HIV-1 assay was evaluated using a panel of 37 group M (subtypes A-D, F, G, CRF01_AE, CRF02_AG and unique recombinant forms) and 2 group O virus isolates. Testing was performed on 273 sample dilutions and compared to VERSANT HIV-1 RNA 3.0 (bDNA) and AMPLICOR HIV-1 MONITOR v1.5 (Monitor v1.5) test results. RealTime HIV-1, bDNA, and Monitor v1.5 tests quantified 87%, 78%, and 81% of samples, respectively. RealTime HIV-1 detected an additional 31 samples at < 40 copies/mL. For group M, RealTime HIV-1 dilution profiles and viral loads were highly correlated with bDNA and Monitor v1.5 values; 87% and 89% of values were within 0.5 log(10) copies/mL. In contrast, the group O viruses were not detected by Monitor v1.5 and were substantially underquantified by approximately 2 log(10) copies/mL in bDNA relative to the RealTime HIV-1 assay. Sequence analysis revealed that RealTime HIV-1 primer/probe binding sites are highly conserved and exhibit fewer nucleotide mismatches relative to Monitor v1.5. The automated m2000 system and RealTime HIV-1 assay offer the advantages of efficient sample processing and throughput with reduced "hands-on" time while providing improved sensitivity, expanded dynamic range and reliable quantification of genetically diverse HIV-1 strains.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17184853     DOI: 10.1016/j.jviromet.2006.11.026

Source DB:  PubMed          Journal:  J Virol Methods        ISSN: 0166-0934            Impact factor:   2.014


  21 in total

1.  Performance of NucliSens HIV-1 EasyQ Version 2.0 compared with six commercially available quantitative nucleic acid assays for detection of HIV-1 in China.

Authors:  Sihong Xu; Aijing Song; Jianhui Nie; Xiuhua Li; Youchun Wang
Journal:  Mol Diagn Ther       Date:  2010-10-01       Impact factor: 4.074

Review 2.  Human immunodeficiency virus type 1 subtype distribution in the worldwide epidemic: pathogenetic and therapeutic implications.

Authors:  L Buonaguro; M L Tornesello; F M Buonaguro
Journal:  J Virol       Date:  2007-07-18       Impact factor: 5.103

3.  Evaluation of the Abbott investigational use only realtime HIV-1 assay and comparison to the Roche Amplicor HIV-1 monitor test, version 1.5.

Authors:  Michael T Pyne; Eric Q Konnick; Amit Phansalkar; David R Hillyard
Journal:  J Mol Diagn       Date:  2009-05-21       Impact factor: 5.568

4.  Evaluation of the Cavidi ExaVir Load assay (version 3) for plasma human immunodeficiency virus type 1 load monitoring.

Authors:  Vicki L Greengrass; Megan M Plate; Pauline M Steele; Justin T Denholm; Catherine L Cherry; Lisa M Morris; Anna Hearps; Suzanne M Crowe
Journal:  J Clin Microbiol       Date:  2009-07-15       Impact factor: 5.948

5.  Sequence conservation of the region targeted by the Abbott RealTime HBV viral load assay in clinical specimens.

Authors:  Gavin A Cloherty; James Rhoads; Thomas P Young; Neil T Parkin; Vera Holzmayer; Lilly Yuen; Carolyn Mullen
Journal:  J Clin Microbiol       Date:  2013-01-23       Impact factor: 5.948

6.  Modification of the Abbott RealTime assay for detection of HIV-1 plasma RNA viral loads less than one copy per milliliter.

Authors:  Steven A Yukl; Peilin Li; Katsuya Fujimoto; Harry Lampiris; Chuanyi M Lu; C Bradley Hare; Steven G Deeks; Teri Liegler; Mark Pandori; Diane V Havlir; Joseph K Wong
Journal:  J Virol Methods       Date:  2011-04-22       Impact factor: 2.014

7.  Comparison of the RealTime HIV-1, COBAS TaqMan 48 v1.0, Easy Q v1.2, and Versant v3.0 assays for determination of HIV-1 viral loads in a cohort of Canadian patients with diverse HIV subtype infections.

Authors:  Deirdre Church; Daniel Gregson; Tracie Lloyd; Marina Klein; Brenda Beckthold; Kevin Laupland; M John Gill
Journal:  J Clin Microbiol       Date:  2010-11-17       Impact factor: 5.948

8.  Single real-time reverse transcription-PCR assay for detection and quantification of genetically diverse HIV-1, SIVcpz, and SIVgor strains.

Authors:  Lucie Etienne; Sabrina Eymard-Duvernay; Avelin Aghokeng; Christelle Butel; Marjorie Monleau; Martine Peeters
Journal:  J Clin Microbiol       Date:  2012-12-19       Impact factor: 5.948

Review 9.  The development of antiretroviral therapy and its impact on the HIV-1/AIDS pandemic.

Authors:  Samuel Broder
Journal:  Antiviral Res       Date:  2009-12-16       Impact factor: 5.970

10.  Evaluation of the Abbott m2000 RealTime human immunodeficiency virus type 1 (HIV-1) assay for HIV load monitoring in South Africa compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 assays.

Authors:  Lesley E Scott; Lara D Noble; Jackie Moloi; Linda Erasmus; Willem D F Venter; Wendy Stevens
Journal:  J Clin Microbiol       Date:  2009-05-06       Impact factor: 5.948