Prostate cancer detection with an improved resonance sensor system: parameter evaluation in a silicone model and on human prostate tissue in vitro.

Peter Lindberg1, Britt Andersson, Anders Bergh, Börje Ljungberg, Olof Lindahl.   

Abstract

The purpose of this study was to improve a resonance sensor system for prostate cancer detection and evaluate its performance on silicone with different hardness. Furthermore, to investigate if the instrument could distinguish between cancerous and normal prostate tissue in one in vitro prostate specimen. The system could measure the frequency shift, impression depth and the rise time of the force signal. The frequency shift, impression depth and the rise time described the relative hardness of silicone (n = 50, P < 0.05). The results from measurements on the prostate specimen indicated that there is a significant difference in the parameter data between cancerous and normal prostate tissue (n = 15, P < 0.05). The parameters' impression depth and force rise time adds important information for cancer detection. Further studies on prostate tissue with different tumour types must be performed in order to understand the full value of the new sensor system.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17093952     DOI: 10.1007/s11517-006-0129-y

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  8 in total

1.  A catheter tactile sensor for measuring hardness of soft tissue: measurement in a silicone model and in an in vitro human prostate model.

Authors:  A Eklund; A Bergh; O A Lindahl
Journal:  Med Biol Eng Comput       Date:  1999-09       Impact factor: 2.602

Review 2.  Cancer burden in the year 2000. The global picture.

Authors:  D M Parkin; F I Bray; S S Devesa
Journal:  Eur J Cancer       Date:  2001-10       Impact factor: 9.162

3.  An applanation resonator sensor for measuring intraocular pressure using combined continuous force and area measurement.

Authors:  Anders Eklund; Per Hallberg; Christina Lindén; Olof A Lindahl
Journal:  Invest Ophthalmol Vis Sci       Date:  2003-07       Impact factor: 4.799

4.  Modeling of micturition characteristics based on prostatic stiffness modulation induced using hormones and adrenergic antagonists.

Authors:  O Sadao; C E Constantinou
Journal:  IEEE Trans Biomed Eng       Date:  1995-08       Impact factor: 4.538

Review 5.  Transrectal ultrasound of the prostate: innovations and future applications.

Authors:  R G Aarnink; H P Beerlage; J J De La Rosette; F M Debruyne; H Wijkstra
Journal:  J Urol       Date:  1998-05       Impact factor: 7.450

Review 6.  Prostatitis and prostate cancer: implications for prostate cancer screening.

Authors:  Jun Kawakami; D Robert Siemens; J Curtis Nickel
Journal:  Urology       Date:  2004-12       Impact factor: 2.649

7.  Evaluation of the digital rectal examination as a screening test for prostate cancer. Rotterdam section of the European Randomized Study of Screening for Prostate Cancer.

Authors:  F H Schröder; P van der Maas; P Beemsterboer; A B Kruger; R Hoedemaeker; J Rietbergen; R Kranse
Journal:  J Natl Cancer Inst       Date:  1998-12-02       Impact factor: 13.506

8.  Impression technique for the assessment of oedema: comparison with a new tactile sensor that measures physical properties of tissue.

Authors:  O A Lindahl; S Omata
Journal:  Med Biol Eng Comput       Date:  1995-01       Impact factor: 2.602

  8 in total
  1 in total

Review 1.  Microfabricated tactile sensors for biomedical applications: a review.

Authors:  Paola Saccomandi; Emiliano Schena; Calogero Maria Oddo; Loredana Zollo; Sergio Silvestri; Eugenio Guglielmelli
Journal:  Biosensors (Basel)       Date:  2014-11-03
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.