Literature DB >> 17081862

Internal anterior chamber diameter using optical coherence tomography compared with white-to-white distances using automated measurements.

Thomas Kohnen1, Martin Christian Thomala, Magdalena Cichocki, Anja Strenger.   

Abstract

PURPOSE: To compare internal horizontal anterior chamber (AC) diameter determined by optical coherence tomography (OCT) and horizontal corneal diameter (white-to-white [WTW]) using automated measurements.
SETTING: Department of Ophthalmology, Johann Wolfgang Goethe-University, Frankfurt am Main, Germany.
METHODS: Internal AC diameter and WTW distance was measured in 52 eyes of 26 patients using the Visante OCT (Carl Zeiss Meditec), IOLMaster (Carl Zeiss Meditec), and Orbscan IIz topography system (Bausch & Lomb). Statistical evaluation was performed using the Bland-Altman method and regression analysis for comparison of measurement techniques. The Kruskal-Wallis test was used to measure the repeatability of each device.
RESULTS: The mean internal AC diameter was 12.45 mm +/- 0.53 (SD) with OCT; the mean WTW distance was 12.17 +/- 0.45 mm with the IOLMaster and 11.84 +/- 0.41 mm with the Orbscan IIz. A positive regression was determined for AC diameter and both WTW measurements. Measurement values varied little between both WTW measurement systems (R(2) = 0.9384).
CONCLUSIONS: Anterior chamber measurement using optical coherence tomography (Visante OCT) was easy to handle and showed good repeatability. The internal horizontal diameter of the AC was larger than the horizontal corneal diameter determined by automated WTW measurements (IOLMaster, Orbscan IIz). Optical coherence tomography with the Visante OCT allows direct measurement of the AC width.

Entities:  

Mesh:

Year:  2006        PMID: 17081862     DOI: 10.1016/j.jcrs.2006.08.023

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  16 in total

1.  Evaluation of myopic corneal diameter with the Orbscan II Topography System.

Authors:  Yi Zha; Wangqiang Feng; Xiaohui Han; Jianqiu Cai
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-06-02       Impact factor: 3.117

2.  [Biometry of the anterior eye segment for implantation of phakic anterior chamber lenses. A comparison of current measurement devices].

Authors:  L Kiraly; G Duncker
Journal:  Ophthalmologe       Date:  2012-03       Impact factor: 1.059

3.  White-to-white corneal diameter: normal values in healthy Iranian population obtained with the Orbscan II.

Authors:  Hamid Gharaee; Mojtaba Abrishami; Masoud Shafiee; Asieh Ehsaei
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

Review 4.  Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.

Authors:  Alberto Domínguez-Vicent; Cari Pérez-Vives; Teresa Ferrer-Blasco; Santiago García-Lázaro; Robert Montés-Micó
Journal:  Int J Ophthalmol       Date:  2016-07-18       Impact factor: 1.779

5.  Repeatability assessment of anterior segment biometric measurements under accommodative and nonaccommodative conditions using an anterior segment OCT.

Authors:  Noelia Martínez-Albert; Jose J Esteve-Taboada; Robert Montés-Micó
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-11-02       Impact factor: 3.117

6.  Factors influencing corneal flap thickness in laser in situ keratomileusis with a femtosecond laser.

Authors:  Chan Young Kim; Ji Hye Song; Kyoung Sun Na; So-Hyang Chung; Choun-Ki Joo
Journal:  Korean J Ophthalmol       Date:  2011-01-17

7.  Comparison of Eyemetrics and Orbscan automated method to determine horizontal corneal diameter.

Authors:  Arvind Venkataraman; Sapna K Mardi; Sarita Pillai
Journal:  Indian J Ophthalmol       Date:  2010 May-Jun       Impact factor: 1.848

Review 8.  Clinical and research applications of anterior segment optical coherence tomography - a review.

Authors:  Jose Luiz Branco Ramos; Yan Li; David Huang
Journal:  Clin Exp Ophthalmol       Date:  2008-11-05       Impact factor: 4.207

9.  Study of distribution of white-to-white corneal diameter and anterior chamber depth in study population obtained with optical biometry using intraocular lens (IOL) master.

Authors:  Kritika Singh; Sandeep Gupta; P S Moulick; Neeraj Bhargava; Alok Sati; Gagandeep Kaur
Journal:  Med J Armed Forces India       Date:  2018-08-14

10.  Limitations in imaging common conjunctival and corneal pathologies with fourier-domain optical coherence tomography.

Authors:  Hakan Demirci; Daniel W Steen
Journal:  Middle East Afr J Ophthalmol       Date:  2014 Jul-Sep