Central venous pulse pressure analysis using an R-synchronized pressure measurement system.

Yoshihisa Fujita1, Daisuke Hayashi, Shinya Wada, Naoki Yoshioka, Takeshi Yasukawa, Gunther Pestel.   

Abstract

OBJECTIVE: The information derived from central venous catheters is underused. We developed an EKG-R synchronization and averaging system to obtained distinct CVP waveforms and analyzed components of these.
METHODS: Twenty-five paralyzed surgical patients undergoing CVP monitoring under mechanical ventilation were studied. CVP and EKG signals were analyzed employing our system, the mean CVP and CVP at end-diastole during expiration were compared, and CVP waveform components were measured using this system.
RESULTS: CVP waveforms were clearly visualized in all patients. They showed the a peak to be 1.8+/- 0.7 mmHg, which was the highest of three peaks, and the x trough to be lower than the y trough (-1.6+/- 0.7 mmHg and -0.9+/- 0.5 mmHg, respectively), with a mean pulse pressure of 3.4 mmHg. The difference between the mean CVP and CVP at end-diastole during expiration was 0.58+/- 0.81 mmHg.
CONCLUSIONS: The mean CVP can be used as an index of right ventricular preload in patients under mechanical ventilation with regular sinus rhythm. Our newly developed system is useful for clinical monitoring and for education in circulatory physiology.

Entities:  

Mesh:

Year:  2006        PMID: 17053869     DOI: 10.1007/s10877-006-9035-y

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  6 in total

1.  Prognostic importance of elevated jugular venous pressure and a third heart sound in patients with heart failure.

Authors:  M H Drazner; J E Rame; L W Stevenson; D L Dries
Journal:  N Engl J Med       Date:  2001-08-23       Impact factor: 91.245

Review 2.  Arterial and central venous pressure monitoring.

Authors:  James A L Pittman; John Sum Ping; Jonathan B Mark
Journal:  Int Anesthesiol Clin       Date:  2004

3.  Comparison of methods of measuring pulmonary artery pressure.

Authors:  J L Lundstedt
Journal:  Am J Crit Care       Date:  1997-07       Impact factor: 2.228

4.  Accurate, automated, continuously displayed pulmonary artery pressure measurement.

Authors:  M M Mitchell; E A Meathe; B R Jones; T E Donch; W G Ricks; J L Benumof; L J Saidman
Journal:  Anesthesiology       Date:  1987-09       Impact factor: 7.892

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Comparison of pulmonary artery and central venous pressure waveform measurements via digital and graphic measurement methods.

Authors:  T S Ahrens; L Schallom
Journal:  Heart Lung       Date:  2001 Jan-Feb       Impact factor: 2.210

  6 in total
  3 in total

1.  Do radial arterial pressure curves have diagnostic validity for identify severe aortic stenosis?

Authors:  Naoki Yoshioka; Yoshihisa Fujita; Takeshi Yasukawa; Itsuro Sano; Masako Kiso; Masayuki Nakayama; Yuka Yoshida
Journal:  J Anesth       Date:  2010-02       Impact factor: 2.078

Review 2.  Using what you get: dynamic physiologic signatures of critical illness.

Authors:  Andre L Holder; Gilles Clermont
Journal:  Crit Care Clin       Date:  2015-01       Impact factor: 3.598

3.  Detection of Junctional Ectopic Tachycardia by Central Venous Pressure.

Authors:  Xin Tan; Yanwan Dai; Ahmed Imtiaz Humayun; Haoze Chen; Genevera I Allen; Parag N Jain
Journal:  Artif Intell Med Conf Artif Intell Med (2005-)       Date:  2021-06-08
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.