Literature DB >> 17003709

Contour and paired-point registration in a model for image-guided surgery.

P Daniel Knott1, Pete S Batra, Robert S Butler, Martin J Citardi.   

Abstract

OBJECTIVES/HYPOTHESIS: This study assesses target registration error (TRE) of contour-based registration (CBR) and paired-point registration (PPR) for endoscopic sinus surgery. STUDY
DESIGN: The experimental registration model consisted of a replica of a human head (Sawbones #1345-27; Pacific Research Laboratories, Vashon, WA). Twelve surface fiducial markers were affixed to the simulated skin on the model, and titanium screws were placed in the regions of the anterior ethmoid (AE) and sphenoid face (SF). An axial computed tomography scan (1-mm slice thickness) was then obtained.
METHODS: Registration was then performed on the InstaTrak 3500 Plus (GE Surgical Navigation & Visualization, Lawrence, MA) with the standard PPR protocol with 12 points and CBR protocol with 500, 250, 125, 50, and 4 points. TRE was then calculated at the AE and SF targets.
RESULTS: Target registration error was significantly lower for paired-point registration compared with contour-based registration (AE, 0.5 mm vs. 1.5 mm, P < .0001; SF, 0.8 mm vs. 1.5 mm, P < .0001). Among contour-based registration protocols, target registration error at the sphenoid face was lowest with 50 points (1.5 mm, P < .02). At the anterior ethmoid, contour-based registration with 50 points produced a lower target registration error than contour-based registration with 125 points (1.5 vs. 1.8 mm, P < .01). Other target registration error values for CBR were similar at both regions. The target registration error interquartile range was lowest with CBR-125 at both regions.
CONCLUSIONS: Paired-point registration provided significantly lower target registration error. Contour-based registration with 125 points (and possibly as few as 50 points) produces clinically acceptable target registration error. This registration model elucidates important concepts about registration for surgical navigation for sinus surgery.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17003709     DOI: 10.1097/01.mlg.0000234939.23796.7a

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  6 in total

1.  Is there an equivalence of non-invasive to invasive referenciation in computer-aided surgery?

Authors:  Tanja D Grauvogel; Juergen Grauvogel; Susan Arndt; Ansgar Berlis; Wolfgang Maier
Journal:  Eur Arch Otorhinolaryngol       Date:  2012-05-05       Impact factor: 2.503

2.  Predicting anatomical landmarks and bone morphology of the femur using local region matching.

Authors:  Cong-Bo Phan; Seungbum Koo
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-02-12       Impact factor: 2.924

3.  Prospects and limitations of different registration modalities in electromagnetic ENT navigation.

Authors:  Eric Soteriou; Juergen Grauvogel; Roland Laszig; Tanja Daniela Grauvogel
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-05-05       Impact factor: 2.503

Review 4.  Progress in surgical management of chronic rhinosinusitis and nasal polyposis.

Authors:  Neil Bhattacharyya
Journal:  Curr Allergy Asthma Rep       Date:  2007-06       Impact factor: 4.806

5.  Navigation accuracy after automatic- and hybrid-surface registration in sinus and skull base surgery.

Authors:  Tanja Daniela Grauvogel; Paul Engelskirchen; Wiebke Semper-Hogg; Juergen Grauvogel; Roland Laszig
Journal:  PLoS One       Date:  2017-07-10       Impact factor: 3.240

6.  Computed tomography-guided navigation assisted drainage for inaccessible deep neck abscess: A case report.

Authors:  Sung Min Han; Hee Sung Chae; Ha Neul Lee; Hyun Jong Jeon; Jeong Pyo Bong; Ji Hoon Kim
Journal:  Medicine (Baltimore)       Date:  2019-03       Impact factor: 1.817

  6 in total