Ocular response analyzer versus Goldmann applanation tonometry for intraocular pressure measurements.

Jose M Martinez-de-la-Casa1, Julian Garcia-Feijoo, Ana Fernandez-Vidal, Carmen Mendez-Hernandez, Julian Garcia-Sanchez.   

Abstract

PURPOSE: To establish correlations between intraocular pressure (IOP) measurements obtained with the ocular response analyzer (ORA) and the Goldmann applanation tonometer (GAT). The effects of central corneal thickness on the measures obtained were also examined.
METHODS: This was a cross-sectional study. IOP was determined in 48 eyes of 48 patients with glaucoma In all patients, central corneal thickness (CCT) was measured by ultrasound pachymetry.
RESULTS: ORA readings were consistently higher than GAT measurements (Goldmann-correlated IOP - IOP GAT mean difference, 7.2 +/- 3.5 mm Hg; corneal-compensated IOP - IOP GAT mean difference, 8.3 +/- 4.0 mm Hg) However, differences were not constant and increased with increasing IOP GAT readings, both with respect to Goldmann-correlated IOP (slope = 0.623, P < 0.0001) and corneal-compensated IOP (slope = 0.538, P < 0.0001). Both pressure measurements provided by the ORA showed significant correlation with CCT (CCT versus Goldmann-correlated IOP: r = 0.460, P = 0.001; CCT versus corneal-compensated IOP: r = 0.442, P = 0.001). No significant effects of corneal curvature or refraction on any of the pressures were observed.
CONCLUSIONS: The ORA significantly overestimates IOP compared with the GAT. Differences between both sets of measures increase as the GAT-determined IOP increases. ORA readings seem to be affected by central corneal thickness.

Entities:  

Mesh:

Year:  2006        PMID: 17003433     DOI: 10.1167/iovs.06-0158

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  26 in total

1.  Repeatability of intra-ocular pressure and central corneal thickness measurements provided by a non-contact method of tonometry and pachymetry.

Authors:  Domenico Schiano Lomoriello; Marco Lombardo; Laura Tranchina; Francesco Oddone; Sebastiano Serrao; Pietro Ducoli
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-10-28       Impact factor: 3.117

Review 2.  Tonometers-which one should I use?

Authors:  Kanza Aziz; David S Friedman
Journal:  Eye (Lond)       Date:  2018-02-19       Impact factor: 3.775

3.  Correlations between Corneal Biomechanics and Glaucoma Severity in Patients with Primary Open Angle Glaucoma.

Authors:  Dana Dascalescu; Catalina Corbu; Mihaela Constantin; Miruna Cristea; Catalina Ionescu; Miruna Cioboata; Liliana Voinea
Journal:  Maedica (Bucur)       Date:  2015-09

4.  Comparison of ocular response analyzer, dynamic contour tonometer and Goldmann applanation tonometer.

Authors:  Charlotte Renier; Thierry Zeyen; Steffen Fieuws; Sofie Vandenbroeck; Ingeborg Stalmans
Journal:  Int Ophthalmol       Date:  2010-05-25       Impact factor: 2.031

5.  Reproducibility of ocular response analyzer measurements and their correlation with central corneal thickness.

Authors:  Joanna Wasielica-Poslednik; Fatmire Berisha; Shakhsanam Aliyeva; Norbert Pfeiffer; Esther M Hoffmann
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-08-10       Impact factor: 3.117

6.  Corneal biomechanical properties measured by the ocular response analyzer in acromegalic patients.

Authors:  Emine Sen; Yasemin Tutuncu; Melike Balikoglu-Yilmaz; Ufuk Elgin; Dilek Berker; Faruk Ozturk; Serdar Guler
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-05-13       Impact factor: 3.117

7.  Effect of corneal astigmatism on intraocular pressure measurement using ocular response analyzer and Goldmann applanation tonometer.

Authors:  Mana Hagishima; Kazutaka Kamiya; Fusako Fujimura; Tetsuya Morita; Nobuyuki Shoji; Kimiya Shimizu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-09-29       Impact factor: 3.117

8.  Effect of biometric characteristics on biomechanical properties of the cornea in cataract patient.

Authors:  Xue-Fei Song; Achim Langenbucher; Zisis Gatzioufas; Berthold Seitz; Moatasem El-Husseiny
Journal:  Int J Ophthalmol       Date:  2016-06-18       Impact factor: 1.779

9.  [Measurement of viscoelastic corneal parameters (corneal hysteresis) in patients with primary open angle glaucoma].

Authors:  B Schroeder; A Hager; A Kutschan; W Wiegand
Journal:  Ophthalmologe       Date:  2008-10       Impact factor: 1.059

10.  Repeatability of the Novel Intraocular Pressure Measurement From Corvis ST.

Authors:  Masato Matsuura; Hiroshi Murata; Yuri Fujino; Mieko Yanagisawa; Yoshitaka Nakao; Shunsuke Nakakura; Yoshiaki Kiuchi; Ryo Asaoka
Journal:  Transl Vis Sci Technol       Date:  2019-06-24       Impact factor: 3.283

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.