[Radiation exposure with 3D rotational angiography of the skull].

D Gosch1, W Kurze, F Deckert, T Schulz, A Patz, T Kahn.   

Abstract

PURPOSE: Determination and comparison of radiation exposure for examinations of the skull with unsubtracted 3D Rotational Angiography (3D RA) and 2D Digital Subtraction Angiography (2D DSA).
MATERIALS AND METHODS: Measurements were carried out with a skull of an Alderson phantom for 3D RA and for 2D DSA in p. a. and lateral projections using an Innova 4100 angiography system with a digital flat panel detector from GE Healthcare. 45 thermoluminescent dosimeters TLD 100H from Harshaw were placed inside the phantom to measure organ doses. In addition the dose area product was recorded and the effective dose was calculated using the Monte Carlo program PCXMC.
RESULTS: For a biplanar DSA run (lateral and p. a. projection), the organ doses were 4 to 5 times higher and the effective dose was 4 times higher than for a 3D RA even though the number of images for the two DSA runs was only half of that for 3D RA.
CONCLUSION: The radiation exposure for unsubtracted 3D RA using a flat panel detector is significantly lower than for biplanar DSA. Using 3D RA in place of 2D DSA can reduce the radiation exposure of patients in neuroradiology procedures.

Entities:  

Mesh:

Year:  2006        PMID: 16953480     DOI: 10.1055/s-2006-926955

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  3 in total

1.  Artificial Intelligence-Based 3D Angiography for Visualization of Complex Cerebrovascular Pathologies.

Authors:  S Lang; P Hoelter; M Schmidt; C Strother; C Kaethner; M Kowarschik; A Doerfler
Journal:  AJNR Am J Neuroradiol       Date:  2021-09-09       Impact factor: 4.966

2.  Dose comparison of classical 2-plane DSA and 3D rotational angiography for the assessment of intracranial aneurysms.

Authors:  N Guberina; U Lechel; M Forsting; C Mönninghoff; U Dietrich; A Ringelstein
Journal:  Neuroradiology       Date:  2016-03-12       Impact factor: 2.804

3.  Effect of ROI filtering in 3D cone-beam rotational angiography on organ dose and effective dose in cerebral investigations.

Authors:  Fabian Göpfert; Ralph Schmidt; Jörg Wulff; Klemens Zink
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.