Filtering techniques for the removal of ventilator artefact in oesophageal pulse oximetry.

K Shafqat1, D P Jones, R M Langford, P A Kyriacou.   

Abstract

The oesophagus has been shown to be a reliable site for monitoring blood oxygen saturation (SpO(2)). However, the photoplethysmographic (PPG) signals from the lower oesophagus are frequently contaminated by a ventilator artefact making the estimation of SpO(2) impossible. A 776th order finite impulse response (FIR) filter and a 695th order interpolated finite impulse response (IFIR) filter were implemented to suppress the artefact. Both filters attenuated the ventilator artefact satisfactorily without distorting the morphology of the PPG when processing recorded data from ten cardiopulmonary bypass patients. The IFIR filter was the better since it conformed more closely to the desired filter specifications and allowed real-time processing. The average improvements in signal-to-noise ratio (SNR) achieved by the FIR and IFIR filters for the fundamental component of the red PPG signals with respect to the fundamental component of the artefact were 57.96 and 60.60 dB, respectively. The corresponding average improvements achieved by the FIR and IFIR filters for the infrared PPG signals were 54.83 and 60.96 dB, respectively. Both filters were also compared with their equivalent tenth order Butterworth filters. The average SNR improvements for the FIR and IFIR filters were significantly higher than those for the Butterworth filters.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16937215     DOI: 10.1007/s11517-006-0089-2

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  7 in total

1.  Evaluation of oesophageal pulse oximetry in patients undergoing cardiothoracic surgery.

Authors:  P A Kyriacou; S L Powell; D P Jones; R M Langford
Journal:  Anaesthesia       Date:  2003-05       Impact factor: 6.955

Review 2.  Pulse oximetry.

Authors:  J F Kelleher
Journal:  J Clin Monit       Date:  1989-01

3.  Evaluation of oesophageal reflectance pulse oximetry in major burns patients.

Authors:  S K Pal; P A Kyriacou; S Kumaran; S Fadheel; R Emamdee; R M Langford; D P Jones
Journal:  Burns       Date:  2005-01-20       Impact factor: 2.744

4.  Pulse oximetry during low cardiac output and hypothermia states immediately after open heart surgery.

Authors:  H Pälve; A Vuori
Journal:  Crit Care Med       Date:  1989-01       Impact factor: 7.598

5.  A prospective study of intraoperative pulse oximetry failure.

Authors:  P R Freund; P T Overand; J Cooper; L Jacobson; S Bosse; B Walker; K L Posner; F W Cheney
Journal:  J Clin Monit       Date:  1991-07

6.  Investigation of oesophageal photoplethysmographic signals and blood oxygen saturation measurements in cardiothoracic surgery patients.

Authors:  P A Kyriacou; S Powell; R M Langford; D P Jones
Journal:  Physiol Meas       Date:  2002-08       Impact factor: 2.833

7.  Predictors of pulse oximetry data failure.

Authors:  D L Reich; A Timcenko; C A Bodian; J Kraidin; J Hofman; M DePerio; S N Konstadt; T Kurki; J B Eisenkraft
Journal:  Anesthesiology       Date:  1996-04       Impact factor: 7.892

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.