Literature DB >> 16919375

Biomechanical comparison of a unique locking plate versus a standard plate for internal fixation of proximal humerus fractures in a cadaveric model.

Stephen Walsh1, Rudy Reindl, Edward Harvey, Gregory Berry, Lorne Beckman, Thomas Steffen.   

Abstract

BACKGROUND: Newer internal fixation devices with a locking mechanism between the plate and the screw have recently been released. The efficacy of these plates in the proximal humerus has yet to be fully described. There is a need to compare the biomechanical properties of efficacy of plate fixation with or without locking screws for surgery of two-part proximal humerus fractures. Multiple-plane locking plate and cloverleaf plate designs were tested to determine their ability to maintain fixation on the humeral head.
METHODS: Eight matched pairs of cadaveric shoulders with 7-millimeter osteotomy defects at the surgical neck simulating two-part fractures of the proximal humerus were loaded to failure in thirty degrees of glenohumeral abduction. One side was repaired with a proximal humerus locking plate and the other with a cloverleaf plate. The rotator cuff musculature was then loaded via a servo-hydraulic testing machine under displacement control to simulate the deforming forces present in vivo.
FINDINGS: The average maximum load to failure was greater in proximal humerus locking plates than in cloverleaf plates (876 versus 712; P=0.04).
INTERPRETATION: In the cadaveric, two-part proximal humerus fracture model that was created, the locking plate displayed significantly greater holding power of the humeral head. Clinical relevance is unproven but may be manifested in vivo as improved early range of motion exercises and functional outcome.

Entities:  

Mesh:

Year:  2006        PMID: 16919375     DOI: 10.1016/j.clinbiomech.2006.06.005

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  25 in total

Review 1.  Surgical treatment options in patients with impaired bone quality.

Authors:  Norman A Johanson; Jody Litrenta; Jay M Zampini; Frederic Kleinbart; Haviva M Goldman
Journal:  Clin Orthop Relat Res       Date:  2011-08       Impact factor: 4.176

2.  Locking buttons increase fatigue life of locking plates in a segmental bone defect model.

Authors:  Marc Tompkins; David J Paller; Douglas C Moore; Joseph J Crisco; Richard M Terek
Journal:  Clin Orthop Relat Res       Date:  2012-10-27       Impact factor: 4.176

3.  The calcar screw in angular stable plate fixation of proximal humeral fractures--a case study.

Authors:  Georg Osterhoff; Christian Ossendorf; Guido A Wanner; Hans-Peter Simmen; Clément M Werner
Journal:  J Orthop Surg Res       Date:  2011-09-24       Impact factor: 2.359

Review 4.  Management of proximal humerus fractures in adults.

Authors:  Leonidas Vachtsevanos; Lydia Hayden; Aravind S Desai; Asterios Dramis
Journal:  World J Orthop       Date:  2014-11-18

5.  Treatment of proximal and middle one-third humeral fractures with lateral distal tibial helical plate.

Authors:  Liang Zhang; Lin-Wei Chen; Wen-Jie Zhang; Chun-Ming Zhao; Bo Huang; Qing Yu; Bin Ni
Journal:  Eur J Orthop Surg Traumatol       Date:  2011-12-09

6.  Treatment for proximal humeral fractures with percutaneous plating: our first results.

Authors:  D Imarisio; A Trecci; L Sabatini; R Scagnelli
Journal:  Musculoskelet Surg       Date:  2013-04-16

7.  Additive fiber-cerclages in proximal humeral fractures stabilized by locking plates: no effect on fracture stabilization and rotator cuff function in human shoulder specimens.

Authors:  Christine Voigt; Christof Hurschler; Louise Rechi; Rolf Vosshenrich; Helmut Lill
Journal:  Acta Orthop       Date:  2009-08       Impact factor: 3.717

8.  Clinical and radiological outcomes of unstable proximal humeral fractures treated with a locking plate and fibular strut allograft.

Authors:  Karnav Panchal; Jae-Jung Jeong; Sang-Eun Park; Weon-Yoo Kim; Hyung-Ki Min; Ju-Yeong Kim; Jong-Hun Ji
Journal:  Int Orthop       Date:  2015-08-11       Impact factor: 3.075

9.  [Effect of additive fiber cerclage of the rotator cuff in angle-stable plate-stabilized proximal humeral fractures: a biomechanical study of human shoulder specimens].

Authors:  C Voigt; C Hurschler; J Althainz; R Vosshenrich; H Lill
Journal:  Unfallchirurg       Date:  2008-07       Impact factor: 1.000

10.  The amount of impaction and loss of reduction in osteoporotic proximal humeral fractures after surgical fixation.

Authors:  S Carbone; M Papalia
Journal:  Osteoporos Int       Date:  2015-09-01       Impact factor: 4.507

View more