Bench model to simulate upper airway obstruction for analyzing automatic continuous positive airway pressure devices.

Jordi Rigau1, Josep M Montserrat, Holger Wöhrle, Diana Plattner, Matthias Schwaibold, Daniel Navajas, Ramon Farré.   

Abstract

BACKGROUND: Automatic positive airway pressure (APAP) devices are increasingly being used in patients with obstructive sleep apnea. Some APAP devices present an unstable behavior when subjected to some events or artifacts. The aims were to develop a bench model capable of reproducing real flow, snoring, and obstructive patterns and to compare the response of APAP devices based on flow and snoring with other devices using, in addition, the forced oscillation technique (FOT).
METHODS: The bench model subjected APAP devices to apneas with and without obstruction, obstructive hypopneas with and without snoring, periods of flow limitation, and artifacts such as leaks and mouth expiration.
RESULTS: Almost all the devices increased the pressure when subjected to apneas with obstruction, but at different rates. The time required by each device to reach 10 cm H(2)O ranged from 2.5 to 13 min. In the presence of apneas without obstruction, all the devices based on flow and snoring increased the pressure at the same rate as during apneas with obstruction. However, the devices using FOT did not modify the pressure. Four devices did not modify the pressure in the presence of obstructive hypopneas, and all but one device increased the pressure in the presence of snoring. Mask leaks had little effect on the response of the devices, but four devices increased the pressure during mouth expiration artifacts.
CONCLUSIONS: When, in addition to the flow and snoring signals, the measurement of the upper airway resistance is included, the accuracy of the event detection algorithms is improved.

Entities:  

Mesh:

Year:  2006        PMID: 16899832     DOI: 10.1378/chest.130.2.350

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  19 in total

1.  Obstructive pressure peak: a new method for differentiation of obstructive and central apneas under auto-CPAP therapy.

Authors:  K H Ruhle; U Domanski; G Nilius
Journal:  Sleep Breath       Date:  2013-03       Impact factor: 2.816

2.  Bench test evaluation of adaptive servoventilation devices for sleep apnea treatment.

Authors:  Kaixian Zhu; Haissam Kharboutly; Jianting Ma; Mourad Bouzit; Pierre Escourrou
Journal:  J Clin Sleep Med       Date:  2013-09-15       Impact factor: 4.062

3.  Is there an optimal nasal pressure for treating obstructive sleep apnea-and if so, what is it?

Authors:  Ramon Farré; Daniel Navajas; Josep M Montserrat
Journal:  Sleep       Date:  2013-04-01       Impact factor: 5.849

4.  Advances in Positive Airway Pressure Treatment Modalities for Hypoventilation Syndromes.

Authors:  Dan Combs; Safal Shetty; Sairam Parthasarathy
Journal:  Sleep Med Clin       Date:  2014-09

5.  APAP and Alternative Titration Methods.

Authors:  Omer Ahmed; Sairam Parthasarathy
Journal:  Sleep Med Clin       Date:  2010-09-01

Review 6.  CPAP and bi-level PAP therapy: new and established roles.

Authors:  Andreea Antonescu-Turcu; Sairam Parthasarathy
Journal:  Respir Care       Date:  2010-09       Impact factor: 2.258

7.  Accuracy of autotitrating CPAP to estimate the residual Apnea-Hypopnea Index in patients with obstructive sleep apnea on treatment with autotitrating CPAP.

Authors:  Himanshu Desai; Anil Patel; Pinal Patel; Brydon J B Grant; M Jeffery Mador
Journal:  Sleep Breath       Date:  2009-05-01       Impact factor: 2.816

8.  Pressure-Relief Features of Fixed and Autotitrating Continuous Positive Airway Pressure May Impair Their Efficacy: Evaluation with a Respiratory Bench Model.

Authors:  Kaixian Zhu; Sami Aouf; Gabriel Roisman; Pierre Escourrou
Journal:  J Clin Sleep Med       Date:  2016-03       Impact factor: 4.062

9.  Autotitration positive airway pressure therapy in patients with obstructive sleep apnea who are intolerant of fixed continuous positive airway pressure.

Authors:  James M Parish; Bernie W Miller; Joseph G Hentz
Journal:  Sleep Breath       Date:  2007-11-28       Impact factor: 2.816

10.  Control of OSA during automatic positive airway pressure titration in a clinical case series: predictors and accuracy of device download data.

Authors:  Hsin-Chia Carol Huang; David R Hillman; Nigel McArdle
Journal:  Sleep       Date:  2012-09-01       Impact factor: 5.849

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.