Evaluation of transmission methodology and attenuation correction for the microPET Focus 220 animal scanner.

Wencke Lehnert1, Steven R Meikle, Stefan Siegel, Danny Newport, Richard B Banati, Anatoly B Rosenfeld.   

Abstract

An accurate, low noise estimate of photon attenuation in the subject is required for quantitative microPET studies of molecular tracer distributions in vivo. In this work, several transmission-based measurement techniques were compared, including coincidence mode with and without rod windowing, singles mode with two different energy sources ((68)Ge and (57)Co), and postinjection transmission scanning. In addition, the effectiveness of transmission segmentation and the propagation of transmission bias and noise into the emission images were examined. The (57)Co singles measurements provided the most accurate attenuation coefficients and superior signal-to-noise ratio, while (68)Ge singles measurements were degraded due to scattering from the object. Scatter correction of (68)Ge transmission data improved the accuracy for a 10 cm phantom but over-corrected for a mouse phantom. (57)Co scanning also resulted in low bias and noise in postinjection transmission scans for emission activities up to 20 MBq. Segmentation worked most reliably for transmission data acquired with (57)Co but the minor improvement in accuracy of attenuation coefficients and signal-to-noise may not justify its use, particularly for small subjects. We conclude that (57)Co singles transmission scanning is the most suitable method for measured attenuation correction on the microPET Focus 220 animal scanner.

Entities:  

Mesh:

Year:  2006        PMID: 16885620     DOI: 10.1088/0031-9155/51/16/008

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  5 in total

1.  Feasibility of template-guided attenuation correction in cat brain PET imaging.

Authors:  Jin Su Kim; Jae Sung Lee; Min-Hyun Park; Kyeong Min Kim; Seung-Ha Oh; Gi Jeong Cheon; In Chan Song; Dae Hyuk Moon; June-Key Chung; Dong Soo Lee
Journal:  Mol Imaging Biol       Date:  2009-12-01       Impact factor: 3.488

2.  Anatomical and functional assessment of brown adipose tissue by magnetic resonance imaging.

Authors:  Y Iris Chen; Aaron M Cypess; Christina A Sass; Anna-Liisa Brownell; Kimmo T Jokivarsi; C Ronald Kahn; Kenneth K Kwong
Journal:  Obesity (Silver Spring)       Date:  2012-02-20       Impact factor: 5.002

3.  137Cs transmission imaging and segmented attenuation corrections in a small animal PET scanner.

Authors:  Ying-Hwey Nai; Takayuki Ose; Miho Shidahara; Hiroshi Watabe
Journal:  Radiol Phys Technol       Date:  2017-07-08

4.  Optimization of PET-MR registrations for nonhuman primates using mutual information measures: a Multi-Transform Method (MTM).

Authors:  Christine M Sandiego; David Weinzimmer; Richard E Carson
Journal:  Neuroimage       Date:  2012-08-25       Impact factor: 6.556

5.  [18F]FDG Uptake in the Aortic Wall Smooth Muscle of Atherosclerotic Plaques in the Simian Atherosclerosis Model.

Authors:  Takayuki Iwaki; Hiroshi Mizuma; Kazuya Hokamura; Hirotaka Onoe; Kazuo Umemura
Journal:  Biomed Res Int       Date:  2016-12-22       Impact factor: 3.411

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.