| Literature DB >> 16878838 |
Jonathan Bravman1, Charbel Ishak, Jonathan Gelber, Suk Namkoong, Laith M Jazrawi, Frederick J Kummer.
Abstract
In a recent experiment examining the effect of interference screw sizing on the fixation stability of multi-strand anterior cruciate ligament grafts, we noticed a large variation in fixation strengths and attributed it to the suturing of the grafts and its interaction with the screw. We performed an experiment using interference screws for fixation of hamstring grafts within a bone tunnel model to compare how the presence of sutures affected graft fixation. We found that having sutures along the region of the graft that contacts the screw within the bone tunnel can increase graft fixation strength 100%.Mesh:
Year: 2006 PMID: 16878838
Source DB: PubMed Journal: Bull Hosp Jt Dis ISSN: 0018-5647