Literature DB >> 16875333

Radiation dose reduction to breast and thyroid during MDCT: effectiveness of an in-plane bismuth shield.

C Hohl1, J E Wildberger, C Süss, C Thomas, G Mühlenbruch, T Schmidt, D Honnef, R W Günther, A H Mahnken.   

Abstract

PURPOSE: To evaluate dose reduction and image deterioration using in-plane bismuth breast-shielding and thyroid-shielding for MDCT.
MATERIAL AND METHODS: Skin and organ doses of thyroid and breast were measured with thermoluminescent dosimeters in a female Alderson-Rando Phantom with and without a 4-ply in-plane bismuth shield. Routine neck (120 kVp, 150 mAs(eff); 16 x 1.5 mm) and chest (120 kVp, 100 mAs(eff); 16 x 1.5 mm) scan protocols were simulated on a 16-row MDCT scanner in three different settings: without shielding, with the shield directly on the surface, and with a 1-cm-thick cotton spacer between surface and shield. Image noise was quantified and compared using the t test.
RESULTS: On average, shielding resulted in a 47% organ-dose reduction for the thyroid and 32% for the breast. Placement of the spacer between shield and surface had no significant impact on the measured doses, but significantly decreased the image noise (P < 0.05).
CONCLUSION: In-plane bismuth breast and thyroid shielding significantly decreases radiation dose in MDCT without deteriorating image quality.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16875333     DOI: 10.1080/02841850600702150

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  17 in total

1.  Bismuth shielding in CT: support for use in children.

Authors:  Sangroh Kim; Donald P Frush; Terry T Yoshizumi
Journal:  Pediatr Radiol       Date:  2010-08-24

2.  Effect of bismuth breast shielding on radiation dose and image quality in coronary CT angiography.

Authors:  Andrew J Einstein; Carl D Elliston; Daniel W Groves; Bin Cheng; Steven D Wolff; Gregory D N Pearson; M Robert Peters; Lynne L Johnson; Sabahat Bokhari; Gary W Johnson; Ketan Bhatia; Theodore Pozniakoff; David J Brenner
Journal:  J Nucl Cardiol       Date:  2011-11-09       Impact factor: 5.952

3.  Prevalence of Protective Shielding Utilization for Radiation Dose Reduction in Adult Patients Undergoing Body Scanning Using Computed Tomography.

Authors:  Shoaib Safiullah; Roshan Patel; Brittany Uribe; Kyle Spradling; Chandana Lall; Lishi Zhang; Zhamshid Okhunov; Ralph V Clayman; Jaime Landman
Journal:  J Endourol       Date:  2017-08-14       Impact factor: 2.942

4.  Inadvertent thyroid irradiation in protocol-driven trauma CT: a survey of hospital ERs.

Authors:  Stephen R Baker; Yih-Hann H Hsieh; Pierre D Maldjian; Michael T Scanlan
Journal:  Emerg Radiol       Date:  2009-01-09

5.  The effects of bismuth breast shields in conjunction with automatic tube current modulation in CT imaging.

Authors:  Sabah Servaes; Xiaowei Zhu
Journal:  Pediatr Radiol       Date:  2013-05-23

Review 6.  State of the art: technologies for computed tomography dose reduction.

Authors:  Martin L D Gunn; Jennifer R Kohr
Journal:  Emerg Radiol       Date:  2009-11-20

7.  New organ-based tube current modulation method to reduce the radiation dose during computed tomography of the head: evaluation of image quality and radiation dose to the eyes in the phantom study.

Authors:  Jung-Su Kim; Soon-Mu Kwon; Jung-Min Kim; Sang-Wook Yoon
Journal:  Radiol Med       Date:  2017-03-24       Impact factor: 3.469

8.  Eye-lens bismuth shielding in paediatric head CT: artefact evaluation and reduction.

Authors:  Maria Raissaki; Kostas Perisinakis; John Damilakis; Nicholas Gourtsoyiannis
Journal:  Pediatr Radiol       Date:  2010-06-16

9.  Dose reduction in CT using bismuth shielding: measurements and Monte Carlo simulations.

Authors:  Kyung-Hwan Chang; Wonho Lee; Dong-Myung Choo; Choon-Sik Lee; Youhyun Kim
Journal:  Radiat Prot Dosimetry       Date:  2009-12-03       Impact factor: 0.972

10.  Shielding effect of radiation dose reduction fiber during the use of C-arm fluoroscopy: a phantom study.

Authors:  Hyemi Cha; Kisung Lee; Moon Seok Park; Kyoung Min Lee; Kyeyoung Cho; Ki Hyuk Sung
Journal:  J Radiat Res       Date:  2020-09-08       Impact factor: 2.724

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.