[Signal loss in magnetic resonance imaging caused by intraoral anchored dental magnetic materials].

F H Blankenstein1, B Truong, A Thomas, R J Schröder, M Naumann.   

Abstract

PURPOSE: To measure the maximum extent of the signal loss areas in the center of the susceptibility artifacts generated by ferromagnetic dental magnet attachments using three different sequences in the 1.5 and 3.0 Tesla MRI.
MATERIALS AND METHODS: Five different pieces of standard dental magnet attachments with volumes of 6.5 to 31.4 mm(3) were used: a NdFeB magnet with an open magnetic field, a NdFeB magnet with a closed magnetic field, a SmCo magnet with an open magnetic field, a stainless steel keeper (AUM-20) and a PdCo piece. The attachments were placed between two cylindrical phantoms and examined in 1.5 and 3.0 Tesla MRI using gradient echo and T1- and T2-weighted spin echoes. We measured the maximum extent of the generated signal loss areas parallel and perpendicular to the direction of B (O).
RESULTS: In gradient echoes the artifacts were substantially larger and symmetrically adjusted around the object. The areas with total signal loss were mushroom-like with a maximum extent of 7.4 to 9.7 cm parallel to the direction of B (O) and 6.7 to 7.4 cm perpendicular to B (O). In spin echoes the signal loss areas were obviously smaller, but not centered. The maximum values ranged between 4.9 and 7.2 cm (parallel B (O)) and 3.6 and 7.0 cm (perpendicular B (O)). The different ferromagnetic attachments had no clinically relevant influence on the signal loss neither in 1.5 T nor 3.0 T MRI.
CONCLUSIONS: Ferromagnetic materials used in dentistry are not intraorally standardized. To ensure, that the area of interest is not affected by the described artifacts, the maximum extent of the signal loss area should be assumed: a radius of up to 7 cm in 1.5 and 3.0 T MRI by T1 and T2 sequences, and a radius of up to 10 cm in T2* sequences. To decide whether magnet attachments have to be removed before MR imaging, physicians should consider both the intact retention of the keepers and the safety distance between the ferromagnetic objects and the area of interest.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16862505     DOI: 10.1055/s-2006-926817

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  7 in total

1.  Predictability of magnetic susceptibility artifacts from metallic orthodontic appliances in magnetic resonance imaging.

Authors:  F Blankenstein; B T Truong; A Thomas; N Thieme; C Zachriat
Journal:  J Orofac Orthop       Date:  2014-11-26       Impact factor: 1.938

2.  MRI with intraoral orthodontic appliance-a comparative in vitro and in vivo study of image artefacts at 1.5 T.

Authors:  C Zachriat; P Asbach; K I Blankenstein; I Peroz; F H Blankenstein
Journal:  Dentomaxillofac Radiol       Date:  2015-03-03       Impact factor: 2.419

3.  [Alteration of flux density of highly coercive SmCo magnets used in prosthodontics and epithetics in 1.5 T and 3 T MRI].

Authors:  Buu-Tai Truong; Felix H Blankenstein
Journal:  Radiologe       Date:  2019-10       Impact factor: 0.635

4.  [Metal objects of the head and neck region in magnetic resonance imaging : Survey among radiologists].

Authors:  Ulrike Kielburg; Felix H Blankenstein
Journal:  Radiologe       Date:  2019-10       Impact factor: 0.635

5.  [Comparison of the two magnetic resonance deartifact techniques in imaging of different porcelain-fused-to-metal crowns].

Authors:  Wen-Jin Li; Jin-Liang Niu; Li Zhu; Yu Wang; Ying An; Shuang-Yuan Zhang
Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi       Date:  2019-02-01

6.  Magnetic permeability as a predictor of the artefact size caused by orthodontic appliances at 1.5 T magnetic resonance imaging.

Authors:  Felix H Blankenstein; Patrick Asbach; Florian Beuer; Johannes Glienke; Stefan Mayer; Christine Zachriat
Journal:  Clin Oral Investig       Date:  2016-03-17       Impact factor: 3.573

Review 7.  Unwanted effects due to interactions between dental materials and magnetic resonance imaging: a review of the literature.

Authors:  Sherin Jose Chockattu; Deepak Byathnal Suryakant; Sophia Thakur
Journal:  Restor Dent Endod       Date:  2018-08-30
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.