Literature DB >> 16799007

Computer navigation of the acetabular component: a cadaver reliability study.

J M F Spencer1, R E Day, K E Sloan, R J Beaver.   

Abstract

Our aim was to assess the intra- and inter-observer reliability in the establishment of the anterior pelvic plane used in imageless computer-assisted navigation. From this we determined the subsequent effects on version and inclination of the acetabular component. A cadaver model was developed with a specifically-designed rod which held the component tracker at a fixed orientation to the pelvis, leaving the anterior pelvic plane as the only variable. Eight surgeons determined the anterior pelvic plane by palpating and registering the bony landmarks as reference points. The exact anterior pelvic plane was then established by using anatomically-placed bone screws as reference points. The difference between the surgeons was found to be highly significant (p < 0.001). The variation was significantly larger for anteversion (sd 9.6 degrees ) than for inclination (sd 6.3 degrees ). The present method for registering pelvic landmarks shows significant inaccuracy, which highlights the need for improved methods of registration before this technique is considered to be safe.

Mesh:

Year:  2006        PMID: 16799007     DOI: 10.1302/0301-620X.88B7.17468

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  12 in total

1.  Measuring acetabular component version after THA: CT or plain radiograph?

Authors:  Benjamin McArthur; Michael Cross; Christina Geatrakas; David Mayman; Bernard Ghelman
Journal:  Clin Orthop Relat Res       Date:  2012-03-01       Impact factor: 4.176

2.  Navigated non-image-based positioning of the acetabulum during total hip replacement.

Authors:  Jean-Yves Jenny; Cyril Boeri; Jean-Claude Dosch; Marius Uscatu; Eugen Ciobanu
Journal:  Int Orthop       Date:  2007-11-13       Impact factor: 3.075

3.  Six sigma analysis of minimally invasive acetabular arthroplasty: a preliminary investigation.

Authors:  David A Heck; James B Stiehl
Journal:  Clin Orthop Relat Res       Date:  2009-05-22       Impact factor: 4.176

4.  Intra- and intersurgeon variability in image-free navigation system for THA.

Authors:  Hirotsugu Ohashi; Masanori Matsuura; Yusaku Okamoto; Yoshiaki Okajima
Journal:  Clin Orthop Relat Res       Date:  2009-04-21       Impact factor: 4.176

5.  Accuracy of computer navigation for acetabular component placement in THA.

Authors:  James A Ryan; Amir A Jamali; William L Bargar
Journal:  Clin Orthop Relat Res       Date:  2009-07-24       Impact factor: 4.176

6.  Computer Navigation for Pediatric Femoral ACL Tunnel Placement.

Authors:  Charles A Popkin; Charles M Chan; Jared A Nowell; Stephen G Crowley; Margaret Wright; Christopher S Ahmad
Journal:  Iowa Orthop J       Date:  2019

7.  Evaluation of the three-dimensional bony coverage before and after rotational acetabular osteotomy.

Authors:  Takeyuki Tanaka; Toru Moro; Yoshio Takatori; Hirofumi Oshima; Hideya Ito; Naohiko Sugita; Mamoru Mitsuishi; Sakae Tanaka
Journal:  Int Orthop       Date:  2018-02-26       Impact factor: 3.075

8.  Accuracy of acetabular cup positioning using imageless navigation.

Authors:  Erik Hohmann; Adam Bryant; Kevin Tetsworth
Journal:  J Orthop Surg Res       Date:  2011-08-10       Impact factor: 2.359

9.  Native femoral anteversion should not be used as reference in cementless total hip arthroplasty with a straight, tapered stem: a retrospective clinical study.

Authors:  Michael Worlicek; Markus Weber; Benjamin Craiovan; Michael Wörner; Florian Völlner; Hans R Springorum; Joachim Grifka; Tobias Renkawitz
Journal:  BMC Musculoskelet Disord       Date:  2016-09-20       Impact factor: 2.362

Review 10.  Navigated cup implantation in hip arthroplasty.

Authors:  Johannes Beckmann; Dirk Stengel; Markus Tingart; Jürgen Götz; Joachim Grifka; Christian Lüring
Journal:  Acta Orthop       Date:  2009-10       Impact factor: 3.717

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.