Literature DB >> 16798767

Twenty years of paediatric cardiac pacing: 515 pacemakers and 480 leads implanted in 292 patients.

Massimo Stefano Silvetti1, Fabrizio Drago, Giorgia Grutter, Antonella De Santis, Vincenzo Di Ciommo, Lucilla Ravà.   

Abstract

AIMS: The aim of this study was to evaluate long-term outcome of pacemakers (PMs) in paediatric patients. METHODS AND
RESULTS: Patients' data were retrospectively reviewed. We recorded the techniques and systems used, any complication, and outcome. Endocardial leads were inserted by transcutaneous puncture of subclavian vein and fixed with a non-absorbable ligature, and epicardial leads by standard surgical technique. Lead survival was calculated and plotted with the product limit method of Kaplan-Meier. Between 1982 and 2002, 292 patients, aged 8+/-7 years (range 1 day-18 years), underwent PM implantation: the first PM had endocardial leads in 165 patients and epicardial in 127 patients. Structural heart disease (HD) was present in 239 patients. Follow-up was 5+/-4 (range 0.1-18) years. There were no pacing-related deaths. In total, 211 endocardial implantation procedures with 90 atrial and 165 ventricular leads and 145 epicardial procedures with 103 atrial and 123 ventricular leads were performed. Early (<3 months) complications: haemothorax occurred in 3.5% of endocardial leads and dislodgement was not significantly different for atrial and ventricular endocardial leads. Late complications: 63 leads failed (48 epicardial), with the worst outcome for conventional epicardial leads (31 vs. 9% endocardial, P<0.05; steroid eluting 8% epicardial vs. 5% endocardial, P=NS). Endocardial atrial leads failed (7%) in operated HD and ventricular leads failed (6%) after body growth, without difference in estimated mean survival time (11 years). Early and late PM infection/erosion was approximately 2% in all patients.
CONCLUSION: Pacing in children shows good results, but complications are frequent and related to leads. Endocardial pacing showed better long-term outcome.

Entities:  

Mesh:

Year:  2006        PMID: 16798767     DOI: 10.1093/europace/eul062

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  15 in total

1.  Pacing and Defibrillators in Complex Congenital Heart Disease.

Authors:  Henry Chubb; Mark O'Neill; Eric Rosenthal
Journal:  Arrhythm Electrophysiol Rev       Date:  2016-05

2.  Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict.

Authors:  M C Post; W Budts; A Van de Bruaene; R Willems; B Meyns; F Rega; M Gewillig
Journal:  Neth Heart J       Date:  2011-08       Impact factor: 2.380

3.  Longitudinal Outcomes of Epicardial and Endocardial Pacemaker Leads in the Adult Fontan Patient.

Authors:  Geoffrey D Huntley; Abhishek J Deshmukh; Carole A Warnes; Suraj Kapa; Alexander C Egbe
Journal:  Pediatr Cardiol       Date:  2018-06-14       Impact factor: 1.655

4.  Clinical outcomes and programming strategies of implantable cardioverter-defibrillator devices in paediatric hypertrophic cardiomyopathy: a UK National Cohort Study.

Authors:  Gabrielle Norrish; Henry Chubb; Ella Field; Karen McLeod; Maria Ilina; Georgia Spentzou; Jan Till; Piers E F Daubeney; Alan Graham Stuart; Jane Matthews; Dominic Hares; Elspeth Brown; Katie Linter; Vinay Bhole; Krishnakumar Pillai; Michael Bowes; Caroline B Jones; Orhan Uzun; Amos Wong; Arthur Yue; Shankar Sadagopan; Tara Bharucha; Norah Yap; Eric Rosenthal; Sujeev Mathur; Satish Adwani; Zdenka Reinhardt; Jasveer Mangat; Juan Pablo Kaski
Journal:  Europace       Date:  2021-03-08       Impact factor: 5.214

5.  Surgical approaches to epicardial pacemaker placement: does pocket location affect lead survival?

Authors:  Brian J Lichtenstein; David P Bichell; Dana M Connolly; John J Lamberti; Suzanne M Shepard; Stephen P Seslar
Journal:  Pediatr Cardiol       Date:  2010-08-06       Impact factor: 1.655

Review 6.  Pacing device therapy in infants and children: a review.

Authors:  Daiji Takeuchi; Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2012-10-27       Impact factor: 1.731

7.  Late Outcome and Predictors of Adverse Events Related to the Implantation of a Permanent Pacemaker in Patients with Isolated Congenital Atrioventricular Block.

Authors:  Grasiele Bess de Oliveira; Fábio Silveira de Oliveira; Ana Paula Susin Osório; Alexandre Kreling Medeiros; João Ricardo Michielin Sant'anna; Tiago Luiz Luz Leiria
Journal:  Pediatr Cardiol       Date:  2016-06-23       Impact factor: 1.655

Review 8.  Cardiac resynchronization therapy in congenital heart disease.

Authors:  Jan Janoušek; Peter Kubuš
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-05-25

9.  Permanent atrial pacing lead implant route after Fontan operation.

Authors:  Kazuhiro Takahashi; Frank Cecchin; Elizabeth Fortescue; Charles I Berul; Mark E Alexander; Edward P Walsh; Francis Fynn-Thompson; John K Triedman
Journal:  Pacing Clin Electrophysiol       Date:  2009-06       Impact factor: 1.976

10.  Surveillance Cultures and Infection in 230 Pacemaker and Defibrillator Generator Changes in Pediatric and Adult Congenital Patients.

Authors:  Gregory Webster; Lauren C Balmert; Ami B Patel; Larry K Kociolek; Melanie Gevitz; Rachael Olson; Ahmed S Chaouki; Osama El-Tayeb; Michael C Monge; Carl Backer
Journal:  World J Pediatr Congenit Heart Surg       Date:  2021-05
View more