Literature DB >> 16783475

Spray pattern analysis for metered dose inhalers: effect of actuator design.

Hugh Smyth1, Geoff Brace, Tony Barbour, Jim Gallion, Joe Grove, Anthony J Hickey.   

Abstract

PURPOSE: This study was conducted to identify the device factors influencing spray pattern and particle size to gain a more complete understanding of spray plume measurements.
METHODS: A statistically designed experiment was used to investigate the influence of three actuator features (orifice diameter, expansion chamber depth, and orifice length) on spray pattern and particle size profiles. Custom-built actuators were manufactured and analyzed with laser light sheet illumination methods for spray patterns and laser diffraction for particle size analysis.
RESULTS: In addition to orifice size, spray patterns were significantly influenced by the actuator orifice length and sump depth. Particle size analysis of the plumes generated from actuators used in these studies showed that all actuator features (orifice size, length, and sump depth) were significant factors influencing particle size.
CONCLUSIONS: The performance of propellant-based metered dose inhaler aerosols seems to be significantly related to sump depth and orifice length, in addition to orifice size. Rational design of propellant-based metered dose inhalers should therefore consider these variables in addition to formulation strategies and simply modifying orifice diameter.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16783475     DOI: 10.1007/s11095-006-0280-z

Source DB:  PubMed          Journal:  Pharm Res        ISSN: 0724-8741            Impact factor:   4.200


  11 in total

1.  An experimental investigation of the spray issued from a pMDI using laser diagnostic techniques.

Authors:  C A Dunbar; A P Watkins; J F Miller
Journal:  J Aerosol Med       Date:  1997

2.  Multimodal particle size distributions emitted from HFA-134a solution pressurized metered-dose inhalers.

Authors:  Hugh D C Smyth; Anthony J Hickey
Journal:  AAPS PharmSciTech       Date:  2003       Impact factor: 3.246

3.  Influence of the metering chamber volume and actuator design on the aerodynamic particle size of a metered dose inhaler.

Authors:  Julianne Berry; Susan Heimbecher; John L Hart; Joel Sequeira
Journal:  Drug Dev Ind Pharm       Date:  2003-09       Impact factor: 3.225

4.  The formulation and evaluation of a CFC-free budesonide pressurised metered dose inhaler.

Authors:  D Ganderton; D Lewis; R Davies; B Meakin; T Church
Journal:  Respir Med       Date:  2003-11       Impact factor: 3.415

Review 5.  The influence of formulation variables on the performance of alternative propellant-driven metered dose inhalers.

Authors:  Hugh D C Smyth
Journal:  Adv Drug Deliv Rev       Date:  2003-07-18       Impact factor: 15.470

6.  Modulite: a means of designing the aerosols generated by pressurized metered dose inhalers.

Authors:  D Ganderton; D Lewis; R Davies; B Meakin; G Brambilla; T Church
Journal:  Respir Med       Date:  2002-08       Impact factor: 3.415

7.  Delivery of propellant soluble drug from a metered dose inhaler.

Authors:  H L Ashworth; C G Wilson; E E Sims; P K Wotton; J G Hardy
Journal:  Thorax       Date:  1991-04       Impact factor: 9.139

8.  Spray pattern analysis for metered dose inhalers I: Orifice size, particle size, and droplet motion correlations.

Authors:  H Smyth; A J Hickey; G Brace; T Barbour; J Gallion; J Grove
Journal:  Drug Dev Ind Pharm       Date:  2006-10       Impact factor: 3.225

9.  Influence of formulation on aerosol particle size.

Authors:  G P Polli; W M Grim; F A Bacher; M H Yunker
Journal:  J Pharm Sci       Date:  1969-04       Impact factor: 3.534

10.  Modulation of aerosol clouds produced by pressurised inhalation aerosols.

Authors:  G Brambilla; D Ganderton; R Garzia; D Lewis; B Meakin; P Ventura
Journal:  Int J Pharm       Date:  1999-09-10       Impact factor: 5.875

View more
  3 in total

1.  Temporally and Spatially Resolved x-ray Fluorescence Measurements of in-situ Drug Concentration in Metered-Dose Inhaler Sprays.

Authors:  Daniel J Duke; Alan L Kastengren; Nicholas Mason-Smith; Yang Chen; Paul M Young; Daniela Traini; David Lewis; Daniel Edgington-Mitchell; Damon Honnery
Journal:  Pharm Res       Date:  2015-11-12       Impact factor: 4.200

2.  The influence of actuator materials and nozzle designs on electrostatic charge of pressurised metered dose inhaler (pMDI) formulations.

Authors:  Yang Chen; Paul M Young; David F Fletcher; Hak Kim Chan; Edward Long; David Lewis; Tanya Church; Daniela Traini
Journal:  Pharm Res       Date:  2013-12-03       Impact factor: 4.200

3.  Tuning aerosol particle size distribution of metered dose inhalers using cosolvents and surfactants.

Authors:  Imran Y Saleem; Hugh D C Smyth
Journal:  Biomed Res Int       Date:  2013-07-24       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.