Literature DB >> 16772625

Manual chest compression vs use of an automated chest compression device during resuscitation following out-of-hospital cardiac arrest: a randomized trial.

Al Hallstrom1, Thomas D Rea, Michael R Sayre, James Christenson, Andy R Anton, Vince N Mosesso, Lois Van Ottingham, Michele Olsufka, Sarah Pennington, Lynn J White, Stephen Yahn, James Husar, Mary F Morris, Leonard A Cobb.   

Abstract

CONTEXT: High-quality cardiopulmonary resuscitation (CPR) may improve both cardiac and brain resuscitation following cardiac arrest. Compared with manual chest compression, an automated load-distributing band (LDB) chest compression device produces greater blood flow to vital organs and may improve resuscitation outcomes.
OBJECTIVE: To compare resuscitation outcomes following out-of-hospital cardiac arrest when an automated LDB-CPR device was added to standard emergency medical services (EMS) care with manual CPR. DESIGN, SETTING, AND PATIENTS: Multicenter, randomized trial of patients experiencing out-of-hospital cardiac arrest in the United States and Canada. The a priori primary population was patients with cardiac arrest that was presumed to be of cardiac origin and that had occurred prior to the arrival of EMS personnel. Initial study enrollment varied by site, ranging from late July to mid November 2004; all sites halted study enrollment on March 31, 2005. INTERVENTION: Standard EMS care for cardiac arrest with an LDB-CPR device (n = 554) or manual CPR (n = 517). MAIN OUTCOME MEASURES: The primary end point was survival to 4 hours after the 911 call. Secondary end points were survival to hospital discharge and neurological status among survivors.
RESULTS: Following the first planned interim monitoring conducted by an independent data and safety monitoring board, study enrollment was terminated. No difference existed in the primary end point of survival to 4 hours between the manual CPR group and the LDB-CPR group overall (N = 1071; 29.5% vs 28.5%; P = .74) or among the primary study population (n = 767; 24.7% vs 26.4%, respectively; P = .62). However, among the primary population, survival to hospital discharge was 9.9% in the manual CPR group and 5.8% in the LDB-CPR group (P = .06, adjusted for covariates and clustering). A cerebral performance category of 1 or 2 at hospital discharge was recorded in 7.5% of patients in the manual CPR group and in 3.1% of the LDB-CPR group (P = .006).
CONCLUSIONS: Use of an automated LDB-CPR device as implemented in this study was associated with worse neurological outcomes and a trend toward worse survival than manual CPR. Device design or implementation strategies require further evaluation. TRIAL REGISTRATION: clinicaltrials.gov Identifier: NCT00120965.

Entities:  

Mesh:

Year:  2006        PMID: 16772625     DOI: 10.1001/jama.295.22.2620

Source DB:  PubMed          Journal:  JAMA        ISSN: 0098-7484            Impact factor:   56.272


  55 in total

Review 1.  Cardiopulmonary resuscitation using electrically driven devices: a review.

Authors:  Anatol Prinzing; Stefan Eichhorn; Marcus-André Deutsch; Ruediger Lange; Markus Krane
Journal:  J Thorac Dis       Date:  2015-10       Impact factor: 2.895

2.  New strategies for cardiopulmonary resuscitation.

Authors:  Jonas A Cooper; Joshua M Cooper
Journal:  Curr Treat Options Cardiovasc Med       Date:  2008-02

3.  [Emergency physician and AutoPulse--a good duo in preclinical emergency services?: case example and report on experience].

Authors:  J-C Schewe; U Heister; A Hoeft; H Krep
Journal:  Anaesthesist       Date:  2008-06       Impact factor: 1.041

Review 4.  [Mechanical resuscitation assist devices].

Authors:  M Fischer; M Breil; M Ihli; M Messelken; S Rauch; J-C Schewe
Journal:  Anaesthesist       Date:  2014-03       Impact factor: 1.041

Review 5.  Mechanical cardiopulmonary resuscitation for patients with cardiac arrest.

Authors:  Lei Jiang; Jin-Song Zhang
Journal:  World J Emerg Med       Date:  2011

6.  Effect of the AutoPulse automated band chest compression device on hemodynamics in out-of-hospital cardiac arrest resuscitation.

Authors:  François-Xavier Duchateau; Papa Gueye; Sonja Curac; Florence Tubach; Claire Broche; Patrick Plaisance; Didier Payen; Jean Mantz; Agnès Ricard-Hibon
Journal:  Intensive Care Med       Date:  2010-03-06       Impact factor: 17.440

7.  [Comments on the 2010 guidelines on cardiopulmonary resuscitation of the European Resuscitation Council].

Authors:  V Wenzel; S G Russo; H R Arntz; J Bahr; M A Baubin; B W Böttiger; B Dirks; U Kreimeier; M Fries; C Eich
Journal:  Anaesthesist       Date:  2010-12       Impact factor: 1.041

8.  Mechanical versus manual chest compressions for cardiac arrest.

Authors:  Peter L Wang; Steven C Brooks
Journal:  Cochrane Database Syst Rev       Date:  2018-08-20

9.  Liver injury diagnosed on computed tomography after use of an automated cardiopulmonary resuscitation device.

Authors:  Jeremy R Camden; Laura R Carucci
Journal:  Emerg Radiol       Date:  2011-04-06

10.  [Outcomes research: definitions, methods and challenges in trauma and orthopaedic surgery].

Authors:  D Stengel; E A Neugebauer; N M Meenen
Journal:  Unfallchirurg       Date:  2007-09       Impact factor: 1.000