PURPOSE: To investigate the effects of high dielectric material padding on RF field distribution in the human head at 7.0 T, and demonstrate the feasibility and effectiveness of RF passive shimming and focusing with such an approach.
MATERIALS AND METHODS: The intensity distribution changes of gradient-recalled-echo (GRE) and spin-echo (SE) images of a human head acquired with water pads (dielectric constant = 78) placed in specified configurations around the head at 7.0 T were evaluated and compared with computer simulation results using the finite difference time domain (FDTD) method. The contributions to the B(1) field distribution change from the displacement current and conductive current of a given configuration of dielectric padding were determined with computer simulations.
RESULTS: MR image intensity distribution in the human head with an RF coil at 7.0 T can be changed drastically by placing water pads around the head. Computer simulations reveal that the high permittivity of water pads results in a strong displacement current that enhances image intensity in the nearby region and alters the intensity distribution of the entire brain.
CONCLUSION: The image intensity distribution in the human head at ultra-high field strengths can be effectively manipulated with high permittivity padding. Utilizing this effect, the B(1) field inside the human head of a given RF coil can be adjusted to reduce the B(1) field inhomogeneity artifact associated with the wave behavior (RF passive shimming) or to locally enhance the signal-to-noise ratio (SNR) in targeted regions of interest (ROIs; RF field focusing). (c) 2006 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2006        PMID: 16755543     DOI: 10.1002/jmri.20603

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  41 in total

1.  Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject-dependent transmit phase measurements.

Authors:  Gregory J Metzger; Carl Snyder; Can Akgun; Tommy Vaughan; Kamil Ugurbil; Pierre-Francois Van de Moortele
Journal:  Magn Reson Med       Date:  2008-02       Impact factor: 4.668

2.  Image homogenization using pre-emphasis method for high field MRI.

Authors:  Ye Li; Chunsheng Wang; Baiying Yu; Daniel Vigneron; Wei Chen; Xiaoliang Zhang
Journal:  Quant Imaging Med Surg       Date:  2013-08

3.  Image quality improvement of composed whole-spine MR images by applying a modified homomorphic filter--first results in cases of multiple myeloma.

Authors:  M P Lichy; C Mueller-Horvat; V Jellus; W Horger; M Horger; C Pfannenberg; B Kiefer; C D Claussen; H P Schlemmer
Journal:  Eur Radiol       Date:  2008-05-29       Impact factor: 5.315

4.  Large improvement of RF transmission efficiency and reception sensitivity for human in vivo31P MRS imaging using ultrahigh dielectric constant materials at 7T.

Authors:  Byeong-Yeul Lee; Xiao-Hong Zhu; Sebastian Rupprecht; Michael T Lanagan; Qing X Yang; Wei Chen
Journal:  Magn Reson Imaging       Date:  2017-07-21       Impact factor: 2.546

5.  A novel method to decrease electric field and SAR using an external high dielectric sleeve at 3 T head MRI: numerical and experimental results.

Authors:  Bu S Park; Sunder S Rajan; Joshua W Guag; Leonardo M Angelone
Journal:  IEEE Trans Biomed Eng       Date:  2014-10-22       Impact factor: 4.538

6.  Variable slice thickness (VAST) EPI for the reduction of susceptibility artifacts in whole-brain GE-EPI at 7 Tesla.

Authors:  Sascha Brunheim; Sören Johst; Viktor Pfaffenrot; Stefan Maderwald; Harald H Quick; Benedikt A Poser
Journal:  MAGMA       Date:  2017-07-10       Impact factor: 2.310

7.  Mapping electrical properties heterogeneity of tumor using boundary informed electrical properties tomography (BIEPT) at 7T.

Authors:  Yicun Wang; Qi Shao; Pierre-Francois Van de Moortele; Emilian Racila; Jiaen Liu; John Bischof; Bin He
Journal:  Magn Reson Med       Date:  2018-09-19       Impact factor: 4.668

8.  [Problems and chances of high field magnetic resonance imaging].

Authors:  M E Ladd; M Bock
Journal:  Radiologe       Date:  2013-05       Impact factor: 0.635

9.  Improvements of transmit efficiency and receive sensitivity with ultrahigh dielectric constant (uHDC) ceramics at 1.5 T and 3 T.

Authors:  Sebastian Rupprecht; Christopher T Sica; Wei Chen; Michael T Lanagan; Qing X Yang
Journal:  Magn Reson Med       Date:  2017-09-25       Impact factor: 4.668

10.  Manipulating transmit and receive sensitivities of radiofrequency surface coils using shielded and unshielded high-permittivity materials.

Authors:  Manushka V Vaidya; Cem M Deniz; Christopher M Collins; Daniel K Sodickson; Riccardo Lattanzi
Journal:  MAGMA       Date:  2017-11-06       Impact factor: 2.310

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.