New "plasma referenced" POCT glucose monitoring systems--are they suitable for glucose monitoring and diagnosis of diabetes?

Reto Savoca1, Beata Jaworek, Andreas R Huber.   

Abstract

BACKGROUND: Results from portable glucose meters should be reported as plasma values as recommended by IFCC. Three new "plasma calibrated" blood glucose meters (Abbott Precision Xcceed, Bayer Ascensia Contour and Roche Accu-Chek Aviva) were compared against laboratory venous plasma glucose measurements to determine their suitability for patient monitoring and diabetes diagnosis.
METHODS: 115 capillary blood samples were analyzed on each of the three meters and compared to venous plasma measurements on the Dade Behring RXL analyzer (Hexokinase method).
RESULTS: We found a significant positive bias of all meters compared to the laboratory reference method. Deviations of more than 10% were seen in more than a third of all glucose values in all three devices. 13%, 8.7% and 10.4%, respectively, of the values from the Abbott, Bayer and Roche devices fell in zone B or C of the Parkes' error grid.
CONCLUSIONS: Neither glucose meter met the ADA or the NACB goals and should therefore not be used as screening tests to diagnose diabetes but are well suitable for patient monitoring.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16697360     DOI: 10.1016/j.cca.2006.03.012

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  8 in total

Review 1.  Glucose meters: a review of technical challenges to obtaining accurate results.

Authors:  Ksenia Tonyushkina; James H Nichols
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

2.  Evaluation of the analytical performance of the coulometry-based Optium Omega blood glucose meter.

Authors:  Bogdan Solnica; Beata Kusnierz-Cabala; Krystyna Slowinska-Solnica; Przemyslaw Witek; Agnieszka Cempa; Maciej T Malecki
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

3.  Markers of cardiovascular risk are not changed by increased whole-grain intake: the WHOLEheart study, a randomised, controlled dietary intervention.

Authors:  Iain A Brownlee; Carmel Moore; Mark Chatfield; David P Richardson; Peter Ashby; Sharron A Kuznesof; Susan A Jebb; Chris J Seal
Journal:  Br J Nutr       Date:  2010-03-23       Impact factor: 3.718

Review 4.  Analytical and clinical performance of blood glucose monitors.

Authors:  Suzanne Austin Boren; William L Clarke
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

5.  Correlation Between Capillary and Venous Blood Glucose in the Lactose Tolerance Test.

Authors:  José Luis Domínguez Jiménez; Antonio Fernández Suárez
Journal:  Dig Dis Sci       Date:  2015-08-30       Impact factor: 3.199

6.  Validity and Reliability of a Glucometer Against Industry Reference Standards.

Authors:  Amanda J Salacinski; Micah Alford; Kathryn Drevets; Sarah Hart; Brian E Hunt
Journal:  J Diabetes Sci Technol       Date:  2014-01-01

7.  International recommendations for glucose control in adult non diabetic critically ill patients.

Authors:  Carole Ichai; Jean-Charles Preiser
Journal:  Crit Care       Date:  2010-09-14       Impact factor: 9.097

8.  Validation of Point-of-Care Glucose Testing for Diagnosis of Type 2 Diabetes.

Authors:  Marijana Vučić Lovrenčić; Vanja Radišić Biljak; Sandra Božičević; Edita Pape-Medvidović; Spomenka Ljubić
Journal:  Int J Endocrinol       Date:  2013-12-08       Impact factor: 3.257

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.