A Garcia-Diaz1, G S Clewley, C L Booth, W Labett, N McAllister, A M Geretti.
Abstract
The performance of the new Abbott real-time human immunodeficiency virus type 1 (HIV-1) assay for HIV-1 RNA load determination in plasma was compared to that of the Abbott LCx HIV-1 RNA quantitative assay following automated RNA isolation by the Abbott m1000 extractor. The measured viral loads of 89 clinical specimens differed by mean 0.19 log10 copies/ml (95% confidence interval, 0.12 to 0.26 log10 copies/ml). Although the difference in viral load determinations was positively skewed in favor of the LCx assay, it did not reach statistical significance (P = 0.42). Results were linearly associated (R2 = 0.94) and strongly correlated (R = 0.96). Good performance was observed with HIV-1 subtypes other than B and circulating recombinant forms, although results obtained with two subtype G specimens and one H specimen showed a more substantial difference.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16672408 PMCID: PMC1479164 DOI: 10.1128/JCM.44.5.1788-1791.2006