Evaluation of a noninvasive cardiac output monitor in mechanically ventilated children.

Astrid Botte1, Francis Leclerc, Yvon Riou, Ahmed Sadik, Véronique Neve, Tameur Rakza, Adelaide Richard.   

Abstract

OBJECTIVE: To compare measurements of cardiac output (CO) and cardiac index (CI) obtained by a recently developed noninvasive continuous cardiac output system, NICO (CONICO), and transthoracic Doppler echocardiography (COTTE) in mechanically ventilated children. DESIGN AND
SETTING: Prospective study in a university-affiliated tertiary pediatric intensive care unit. PATIENTS: A total of 21 mechanically ventilated children, weighing >15 kg, in stable respiratory and hemodynamic condition. MEASUREMENTS: Sets of three successive measurements of CO with the NICO system and transthoracic Doppler echocardiography were obtained. Bland-Altman analysis was used to compare the agreement between the two methods.
RESULTS: The mean +/- sd CO values were 4.06 +/- 1.43 L/min for CONICO and 4.67 +/- 1.78 L/min for COTTE. Bias +/- sd between the two methods was -0.61 +/- 0.94 L/min. The variability of the difference between the two methods increased as the magnitude of the CO measurement increased. Similar results were obtained for cardiac index: 4.01 +/- 1.40 L.min.m for CINICO and 4.59 +/- 1.48 L.min.m for CITTE. Bland-Altman analysis revealed a nonuniform relationship between CI difference and the magnitude (y = -0.299 - 0.0655 x mean). The variability of the differences did not increase as the magnitude of the CO measurement increased (sd of estimate was 0.827 L.min.m). With both CONICO and CINICO, each measurement was highly repeatable, with coefficient of variation of only 2.88% +/- 2.31%. Repeatability with Doppler echocardiography was 7.02% +/- 4.33%.
CONCLUSIONS: The NICO system is a new device that measures CO easily and automatically in mechanically ventilated children weighing >15 kg. CO values obtained with this technique were in agreement with those obtained with Doppler echocardiography in children in respiratory and hemodynamic stable condition. The NICO system needs further investigation in children in unstable respiratory and hemodynamic condition.

Entities:  

Mesh:

Year:  2006        PMID: 16575346     DOI: 10.1097/01.PCC.0000216691.69976.E4

Source DB:  PubMed          Journal:  Pediatr Crit Care Med        ISSN: 1529-7535            Impact factor:   3.624


  7 in total

Review 1.  Accuracy and precision of minimally-invasive cardiac output monitoring in children: a systematic review and meta-analysis.

Authors:  Koichi Suehiro; Alexandre Joosten; Linda Suk-Ling Murphy; Olivier Desebbe; Brenton Alexander; Sang-Hyun Kim; Maxime Cannesson
Journal:  J Clin Monit Comput       Date:  2015-08-29       Impact factor: 2.502

2.  Correlation between Regional Cerebral Saturation and Invasive Cardiac Index Monitoring after Heart Transplantation Surgery.

Authors:  Awni M Al-Subu; Christoph P Hornik; Ira M Cheifetz; Andrew J Lodge; George Ofori-Amanfo
Journal:  J Pediatr Intensive Care       Date:  2018-06-11

3.  Capability of a new paediatric oesophageal Doppler monitor to detect changes in cardiac output during testing of external pacemakers after cardiac surgery.

Authors:  Thilo Fleck; Stephan Schubert; Brigitte Stiller; Matthias Redlin; Peter Ewert; Nicole Nagdyman; Felix Berger
Journal:  J Clin Monit Comput       Date:  2011-11-12       Impact factor: 2.502

4.  Volumetric Capnography in Pediatric Extracorporeal Membrane Oxygenation: A Case Series.

Authors:  Kim R Derespina; Shivanand S Medar; Scott I Aydin; Shubhi Kaushik; Awni Al-Subu; George Ofori-Amanfo
Journal:  J Pediatr Intensive Care       Date:  2020-10-05

5.  Continuous, non-invasive techniques to determine cardiac output in children after cardiac surgery: evaluation of transesophageal Doppler and electric velocimetry.

Authors:  Stephan Schubert; Thomas Schmitz; Markus Weiss; Nicole Nagdyman; Michael Huebler; Vladimir Alexi-Meskishvili; Felix Berger; Brigitte Stiller
Journal:  J Clin Monit Comput       Date:  2008-07-30       Impact factor: 2.502

6.  Volumetric Capnography Monitoring and Effects of Epinephrine on Volume of Carbon Dioxide Elimination during Resuscitation after Cardiac Arrest in a Swine Pediatric Ventricular Fibrillatory Arrest.

Authors:  Awni M Al-Subu; Timothy A Hacker; Jens C Eickhoff; George Ofori-Amanfo; Marlowe W Eldridge
Journal:  J Pediatr Intensive Care       Date:  2020-06-01

7.  Comparison of three non-invasive hemodynamic monitoring methods in critically ill children.

Authors:  Chanapai Chaiyakulsil; Marut Chantra; Poomiporn Katanyuwong; Anant Khositseth; Nattachai Anantasit
Journal:  PLoS One       Date:  2018-06-18       Impact factor: 3.240

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.