The influence of photon depth of interaction and non-collinear spread of annihilation photons on PET image spatial resolution.

Alejandro Sánchez-Crespo1, Stig A Larsson.   

Abstract

PURPOSE: The quality of PET imaging is impaired by parallax errors. These errors produce misalignment between the projected location of the true origin of the annihilation event and the line of response determined by the coincidence detection system. Parallax errors are due to the varying depths of photon interaction (DOI) within the scintillator and the non-collinear (NC) emission of the annihilation photons. The aim of this work was to address the problems associated with the DOI and the NC spread of annihilation photons and to develop a quantitative model to assess their impact on image spatial resolution losses for various commonly used scintillators and PET geometries.
METHODS: A theoretical model based on Monte Carlo simulations was developed to assess the relative influence of DOI and the NC spread of annihilation photons on PET spatial resolution for various scintillator materials (BGO, LSO, LuAP, GSO, NaI) and PET geometries.
RESULTS: The results demonstrate good agreement between simulated, experimental and published overall spatial resolution for some commercial systems, with maximum differences around 1 mm in both 2D and 3D mode. The DOI introduces an impairment of non-stationary spatial resolution along the radial direction, which can be very severe at peripheral positions. As an example, the radial spatial resolution loss due to DOI increased from 1.3 mm at the centre to 6.7 mm at 20 cm from the centre of a BGO camera with a 412-mm radius in 2D mode. Including the NC, the corresponding losses were 3.0 mm at the centre and 7.3 mm 20 cm from the centre.
CONCLUSION: Without a DOI detection technique, it seems difficult to improve PET spatial resolution and increase sensitivity by reducing the detector ring radius or by extending the detector in the axial direction. Much effort is expended on the design and configuration of smaller detector elements but more effort should be devoted to the DOI complexity.

Entities:  

Mesh:

Year:  2006        PMID: 16568208     DOI: 10.1007/s00259-005-0024-z

Source DB:  PubMed          Journal:  Eur J Nucl Med Mol Imaging        ISSN: 1619-7070            Impact factor:   9.236


  8 in total

1.  Influence of the angle of incidence on the sensitivity of gamma camera based PET.

Authors:  Stefaan Vandenberghe; Yves D'Asseler; Jeff Kolthammer; Rik Van de Walle; Ignace Lemahieu; Rudi A Dierckx
Journal:  Phys Med Biol       Date:  2002-01-21       Impact factor: 3.609

2.  Positron flight in human tissues and its influence on PET image spatial resolution.

Authors:  Alejandro Sánchez-Crespo; Pedro Andreo; Stig A Larsson
Journal:  Eur J Nucl Med Mol Imaging       Date:  2003-10-10       Impact factor: 9.236

3.  Performance evaluation of the positron scanner ECAT EXACT.

Authors:  K Wienhard; L Eriksson; S Grootoonk; M Casey; U Pietrzyk; W D Heiss
Journal:  J Comput Assist Tomogr       Date:  1992 Sep-Oct       Impact factor: 1.826

4.  PETSIM: Monte Carlo simulation of all sensitivity and resolution parameters of cylindrical positron imaging systems.

Authors:  C J Thompson; J Moreno-Cantu; Y Picard
Journal:  Phys Med Biol       Date:  1992-03       Impact factor: 3.609

5.  Performance measurements of a depth-encoding PET detector module based on position-sensitive avalanche photodiode read-out.

Authors:  P A Dokhale; R W Silverman; K S Shah; R Grazioso; R Farrell; J Glodo; M A McClish; G Entine; V H Tran; S R Cherry
Journal:  Phys Med Biol       Date:  2004-09-21       Impact factor: 3.609

Review 6.  PET instrumentation: what are the limits?

Authors:  T F Budinger
Journal:  Semin Nucl Med       Date:  1998-07       Impact factor: 4.446

7.  Scintillation crystals for PET.

Authors:  C L Melcher
Journal:  J Nucl Med       Date:  2000-06       Impact factor: 10.057

8.  Fully three-dimensional positron emission tomography.

Authors:  J G Colsher
Journal:  Phys Med Biol       Date:  1980-01       Impact factor: 3.609

  8 in total
  1 in total

1.  Hallmarks in prostate cancer imaging with Ga68-PSMA-11-PET/CT with reference to detection limits and quantitative properties.

Authors:  Alejandro Sanchez-Crespo; Emma Jussing; Ann-Charlotte Björklund; Katja Pokrovskaja Tamm
Journal:  EJNMMI Res       Date:  2018-04-04       Impact factor: 3.138

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.