Literature DB >> 16567025

Biomechanical performance of the new BeadEx implant in the treatment of osteoporotic vertebral body compression fractures: restoration and maintenance of height and stability.

A Kettler1, W Schmoelz, Y Shezifi, N Ohana, A Ben-Arye, L Claes, H-J Wilke.   

Abstract

BACKGROUND: Vertebral compression fractures are counted among the most common complications of osteoporosis. For treatment, a new, alternative implant has been developed (BeadEx, Expandis, Hof HaCarmel, Israel). The aim of the present in vitro study was to evaluate whether this implant is able to restore the initial height and three-dimensional stability after fracture and whether it is able to maintain this height and stability during complex cyclic loading.
METHODS: The BeadEx implant consists of small titanium rolls, which are pressed into the vertebral body through specially designed, hollow pedicle screws. The height and the three-dimensional flexibility of 18 bisegmental spine specimens (nine T12-L2, nine L3-L5) was measured, first, before and after creating a wedge compression fracture at the middle vertebral body (L1 resp. L4), second, after treatment of the fracture, and, third, during and after complex cyclic loading. The fractures were treated either with BeadEx plus internal fixator, BeadEx plus bone cement or vertebroplasty for comparison.
FINDINGS: The height before fracture could almost be restored by BeadEx plus bone cement but not by BeadEx plus fixator and vertebroplasty. The total height loss after cyclic loading was smallest with BeadEx plus bone cement (in median -4.7mm with respect to the intact specimens) but -6.2mm with BeadEx plus fixator and -7.8mm with vertebroplasty. The three-dimensional stability of the specimens was clearly higher if treated with BeadEx plus fixator than with BeadEx plus bone cement or vertebroplasty.
INTERPRETATION: From a biomechanical point of view, BeadEx plus bone cement can be recommended as an alternative to vertebroplasty in the treatment of osteoporotic vertebral body fractures. BeadEx plus fixator can be recommended if additional stability is needed.

Entities:  

Mesh:

Year:  2006        PMID: 16567025     DOI: 10.1016/j.clinbiomech.2006.02.005

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  11 in total

1.  Vertebroplasty with self-locking hexagonal metal implants shows comparable primary and secondary stiffness to PMMA cement augmentation techniques in a biomechanical vertebral compression fracture model.

Authors:  W Schmoelz; A C Disch; J F Huber
Journal:  Eur Spine J       Date:  2010-03-07       Impact factor: 3.134

Review 2.  [Biomechanical aspects of vertebral augmentation].

Authors:  H-J Wilke
Journal:  Unfallchirurg       Date:  2015-10       Impact factor: 1.000

3.  Surgical treatment of osteoporotic thoraco-lumbar compressive fractures: the use of pedicle screw with augmentation PMMA.

Authors:  Massimo Girardo; P Cinnella; G Gargiulo; P Viglierchio; A Rava; S Aleotti
Journal:  Eur Spine J       Date:  2017-03-21       Impact factor: 3.134

4.  Biomechanical in vitro comparison of radiofrequency kyphoplasty and balloon kyphoplasty.

Authors:  Gerhard Achatz; Hans-Joachim Riesner; Benedikt Friemert; Raimund Lechner; Nicolas Graf; Hans-Joachim Wilke
Journal:  Eur Spine J       Date:  2017-04-27       Impact factor: 3.134

5.  A demineralized calf vertebra model as an alternative to classic osteoporotic vertebra models for pedicle screw pullout studies.

Authors:  Atilla Akbay; Gokhan Bozkurt; Ozgur Ilgaz; Selcuk Palaoglu; Nejat Akalan; Edward C Benzel
Journal:  Eur Spine J       Date:  2007-11-17       Impact factor: 3.134

6.  Effectiveness and reliability of cannulated fenestrated screws augmented with polymethylmethacrylate cement in the surgical treatment of osteoporotic vertebral fractures.

Authors:  Angela Coniglio; Alessandro Rava; Federico Fusini; Gabriele Colò; Alessandro Massè; Massimo Girardo
Journal:  J Craniovertebr Junction Spine       Date:  2021-03-04

7.  Straight-Forward versus Bicortical Fixation Penetrating Endplate in Lumbosacral Fixation-A Biomechanical Study.

Authors:  Ahmet Karakasli; Nihat Acar; Bora Uzun
Journal:  J Korean Neurosurg Soc       Date:  2018-02-28

8.  Biomechanical assessment of new surgical method instead of kyphoplasty to improve the mechanical behavior of the vertebra: Micro finite element study.

Authors:  Seyed Aref Hosseini Faradonbeh; Nima Jamshidi
Journal:  World J Orthop       Date:  2017-11-18

Review 9.  Surgical Trend Analysis for Use of Cement Augmented Pedicle Screws in Osteoporosis of Spine: A Systematic Review (2000-2017).

Authors:  Vishwajeet Singh; Rajat Mahajan; Kalidutta Das; Harvinder Singh Chhabra; Tarush Rustagi
Journal:  Global Spine J       Date:  2018-09-27

10.  Stent-Screw Assisted Internal Fixation of Osteoporotic Vertebrae: A Comparative Finite Element Analysis on SAIF Technique.

Authors:  Luigi La Barbera; Alessandro Cianfoni; Andrea Ferrari; Daniela Distefano; Giuseppe Bonaldi; Tomaso Villa
Journal:  Front Bioeng Biotechnol       Date:  2019-10-25
View more