| Literature DB >> 16518312 |
Sotirios Stathakis1, Constantin Kappas, Kiki Theodorou, Nikos Papanikolaou, Jean-Claude Rosenwald.
Abstract
A number of treatment-planning systems still use conventional correction methods for body inhomogeneities. Most of these methods (power law method, tissue-air ratio (TAR), etc.) consider only on-axis points, rectangular fields, and inhomogeneous slabs covering the whole irradiating field. A new method is proposed that overcomes the above limitations. The new method uses the principle of the Clarkson method on sector integration to take into account the position and lateral extent of the inhomogeneity with respect to the point of calculation, as well as the shape of the irradiating field. The field is divided into angular sectors, and each sector is then treated separately for the presence of inhomogeneities using a conventional correction method. Applying this method, we can predict the correction factors for Co-60 and 6-MV photon beams for irregular fields that include inhomogeneities of lower or higher densities relative to water. Validation of the predicted corrections factors was made against Monte Carlo calculations for the same geometries. The agreement between the predicted correction factors and the Monte Carlo calculations was within 1.5%. In addition, the new method was able to predict the behavior of the correction factor when the point of calculation was approaching or moving away from the interface between two materials.Entities:
Mesh:
Year: 2006 PMID: 16518312 PMCID: PMC5722477 DOI: 10.1120/jacmp.v7i1.2042
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Case of a rectangular inhomogeneity in a rectangular field of dimensions 2x and 2y. and : distances from point P to the inhomogeneity limits and .
Figure 2The BSM applied for each sector for a rectangular field with blocks shaping the field
Figure 3Case 2 beam's‐eye view of the BSM applied for a sector containing two inhomogeneities with the last one adjacent to the edge of the field
Figure 4Setup geometry for the experiment. The dimensions of the blocks when projected to the level of the point of interest are .
Figure 5Comparison of the correction factors for lung inhomogeneity at a depth of 10 cm along the x‐ and y‐axes for the Co‐60 field
Figure 8Comparison of the correction factors for bone inhomogeneity at a depth of 10 cm along the x‐ and y‐axes for the 6‐MV field
Figure 6Comparison of the correction factors for bone inhomogeneity at a depth of 10 cm along the x‐ and y‐axes for the Co‐60 field