Literature DB >> 11514010

Impact of simple tissue inhomogeneity correction algorithms on conformal radiotherapy of lung tumours.

M Engelsman1, E M Damen, P W Koken, A A van 't Veld, K M van Ingen, B J Mijnheer.   

Abstract

BACKGROUND AND
PURPOSE: Conformal radiotherapy requires accurate dose calculation at the dose specification point, at other points in the planning target volume (PTV) and in organs at risk. To assess the limitations of treatment planning of lung tumours, errors in dose values, calculated by some simple tissue inhomogeneity correction algorithms available in a number of currently applied treatment planning systems, have been quantified.
MATERIALS AND METHODS: Single multileaf collimator-shaped photon beams of 6, 8, 15 and 18 MV nominal energy were used to irradiate a 50 mm diameter spherical solid tumour, simulated by polystyrene, which was located centrally inside lung tissue, simulated by cork. The planned dose distribution was made conformal to the PTV, which was a 15 mm three-dimensional expansion of the tumour. Values of both the absolute dose at the International Commission on Radiation Units and Measurement (ICRU) reference point and relative dose distributions inside the PTV and in the lung were calculated using three inhomogeneity correction algorithms. The algorithms investigated in this study are the pencil beam algorithm with one-dimensional corrections, the modified Batho algorithm and the equivalent path length algorithm. The calculated data were compared with measurements for a simple beam set-up using radiographic film and ionization chambers.
RESULTS: For this specific configuration, deviations of up to 3.5% between calculated and measured values of the dose at the ICRU reference point were found. Discrepancies between measured and calculated beam fringe values (distance between the 50 and 90% isodose lines) of up to 14 mm have been observed. The differences in beam fringe and penumbra width (20-80%) increase with increasing beam energy. Our results demonstrate that an underdosage of the PTV up to 20% may occur if calculated dose values are used for treatment planning. The three algorithms predict a considerably higher dose in the lung, both along the central beam axis and in the lateral direction, compared with the actual delivered dose values.
CONCLUSIONS: The dose at the ICRU reference point of such a tumour in lung geometry is calculated with acceptable accuracy. Differences between calculated and measured dose distributions are primarily due to changes in electron transport in the lung, which are not adequately taken into account by the simple tissue inhomogeneity correction algorithms investigated in this study. Particularly for high photon beam energies, clinically unacceptable errors will be introduced in the choice of field sizes employed for conformal treatments, leading to underdosage of the PTV. In addition, the dose to the lung will be wrongly predicted which may influence the choice of the prescribed dose level in dose-escalation studies.

Entities:  

Mesh:

Year:  2001        PMID: 11514010     DOI: 10.1016/s0167-8140(01)00387-5

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  11 in total

1.  Dosimetric verification of the anisotropic analytical algorithm in lung equivalent heterogeneities with and without bone equivalent heterogeneities.

Authors:  Kaoru Ono; Satoru Endo; Kenichi Tanaka; Masaharu Hoshi; Yutaka Hirokawa
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

2.  The importance of surrounding tissues and window settings for contouring of moving targets.

Authors:  Kai Joachim Borm; Markus Oechsner; Johannes Berndt; Stephanie Elisabeth Combs; Michael Molls; Marciana Nona Duma
Journal:  Strahlenther Onkol       Date:  2015-06-19       Impact factor: 3.621

3.  An institutional experience of quality assurance of a treatment planning system on photon beam.

Authors:  Yıldıray Ozgüven; Kadir Yaray; Fadime Alkaya; Birsen Yücel; Serdar Soyuer
Journal:  Rep Pract Oncol Radiother       Date:  2013-12-08

4.  Clinical implications in the use of the PBC algorithm versus the AAA by comparison of different NTCP models/parameters.

Authors:  Antonella Bufacchi; Barbara Nardiello; Roberto Capparella; Luisa Begnozzi
Journal:  Radiat Oncol       Date:  2013-07-04       Impact factor: 3.481

5.  Lung dose analysis in loco-regional hypofractionated radiotherapy of breast cancer.

Authors:  Mohammad A Attar; Yasir A Bahadur; Camelia T Constantinescu; Maha M Eltaher
Journal:  Saudi Med J       Date:  2016-06       Impact factor: 1.484

6.  Dosimetric comparison of different inhomogeneity correction algorithms for external photon beam dose calculations.

Authors:  Khadijeh Asnaashari; Mohammad Rasa Golrokh Nodehi; Seyyed Rabie Mahdavi; Somaye Gholami; Hamid Reza Khosravi
Journal:  J Med Phys       Date:  2013-04

7.  An inhomogeneity correction algorithm for irregular fields of high-energy photon beams based on Clarkson integration and the 3D beam subtraction method.

Authors:  Sotirios Stathakis; Constantin Kappas; Kiki Theodorou; Nikos Papanikolaou; Jean-Claude Rosenwald
Journal:  J Appl Clin Med Phys       Date:  2006-02-15       Impact factor: 2.102

8.  Initial evaluation of a four-dimensional computed tomography system using a programmable motor.

Authors:  Luc Simon; Philippe Giraud; Vincent Servois; Jean-Claude Rosenwald
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102

9.  Inhomogeneity correction and the analytic anisotropic algorithm.

Authors:  Don Robinson
Journal:  J Appl Clin Med Phys       Date:  2008-05-01       Impact factor: 2.102

10.  Dosimetric comparison of pencil beam and Monte Carlo algorithms in conformal lung radiotherapy.

Authors:  Yelda Elcim; Bahar Dirican; Omer Yavas
Journal:  J Appl Clin Med Phys       Date:  2018-08-05       Impact factor: 2.102