Literature DB >> 16500737

Experimental evaluation of the holding power/stiffness of the self-tapping bone screws in normal and osteoporotic bone material.

Suneel Battula1, Andrew Schoenfeld, Gregory Vrabec, Glen O Njus.   

Abstract

OBJECTIVE: The goal of this study is to compare the holding power of the self-tapping bone screws in normal and osteoporotic bone materials.
BACKGROUND: Self-tapping screws are increasingly being used in orthopaedic surgery due to their advantages over the other bone screws.
METHODS: Screws were divided into five groups (six screws per group) based on the depth of insertion in the bone coupons that represented normal and osteoporotic bones. Screws were randomly inserted into the bone coupons with tips of the screws being -1 mm, 0 mm, 1 mm, 2 mm and 3 mm relative to the far cortex. Biomechanical testing was performed using an Instron 8,511 in accordance with the American Society for Testing and Materials standards for bone screws. Two-factor analysis of variance (ANOVA) was used to determine if the holding power of the screws were different with respect to insertion depths and bone materials.
FINDINGS: The bone materials had a significant difference (P < 0.05) in the holding power and depths of insertion past the far cortex were significantly different from one another in holding power. The affect of the screw material on the holding power of the self-tapping screws in different bone materials was also examined. The performance of stainless steel screws was superior to that of titanium screws in the osteoporotic material.
INTERPRETATION: Based on the results it can be concluded that the depth of insertion of the tip of the screw for adequate fracture fixation in normal bone is 1mm or more past the far cortex and in osteoporotic bone it is at least 2mm past the far cortex.

Entities:  

Mesh:

Year:  2006        PMID: 16500737     DOI: 10.1016/j.clinbiomech.2005.12.020

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  9 in total

1.  Pullout Strength After Multiple Reinsertions in Radial Bone Fixation.

Authors:  Alfonso Mejia; Giovanni Solitro; Elena Gonzalez; Amit Parekh; Mark Gonzalez; Farid Amirouche
Journal:  Hand (N Y)       Date:  2018-09-06

2.  A surrogate long-bone model with osteoporotic material properties for biomechanical testing of fracture implants.

Authors:  Mark B Sommers; Daniel C Fitzpatrick; Steven M Madey; Corey Vande Zanderschulp; Michael Bottlang
Journal:  J Biomech       Date:  2007-06-18       Impact factor: 2.712

3.  A study on the use of the Osstell apparatus to evaluate pedicle screw stability: An in-vitro study using micro-CT.

Authors:  Daisuke Nakashima; Ken Ishii; Morio Matsumoto; Masaya Nakamura; Takeo Nagura
Journal:  PLoS One       Date:  2018-06-28       Impact factor: 3.240

4.  Factors Associated With the Accuracy of Depth Gauge Measurements.

Authors:  Pengcheng Liu; Joanna Xi Xiao; Chen Zhao; Xiaodong Li; Guantong Sun; Fei Yang; Xiaoqing Wang
Journal:  Front Surg       Date:  2022-01-13

5.  Parametric analysis of orthopedic screws in relation to bone density.

Authors:  Elisabetta M Zanetti; Massimiliano Salaorno; Giovanni Grasso; Alberto L Audenino
Journal:  Open Med Inform J       Date:  2009-04-21

6.  Validity of synthetic bone as a substitute for osteoporotic cadaveric femoral heads in mechanical testing: A biomechanical study.

Authors:  F O'Neill; F Condon; T McGloughlin; B Lenehan; C Coffey; M Walsh
Journal:  Bone Joint Res       Date:  2012-04-01       Impact factor: 5.853

7.  Compressive properties of commercially available polyurethane foams as mechanical models for osteoporotic human cancellous bone.

Authors:  Purvi S D Patel; Duncan E T Shepherd; David W L Hukins
Journal:  BMC Musculoskelet Disord       Date:  2008-10-09       Impact factor: 2.362

8.  Biomechanical evaluation of bone screw fixation with a novel bone cement.

Authors:  Tiina Juvonen; Juha-Pekka Nuutinen; Arto P Koistinen; Heikki Kröger; Reijo Lappalainen
Journal:  Biomed Eng Online       Date:  2015-07-30       Impact factor: 2.819

9.  Analysis of Locking Self-Taping Bone Screws for Angularly Stable Plates.

Authors:  Radek Čada; Karel Frydrýšek; František Sejda; Jiří Demel; Leopold Pleva
Journal:  J Med Biol Eng       Date:  2017-06-21       Impact factor: 1.553

  9 in total