Performance of mechanical ventilators at the patient's home: a multicentre quality control study.

R Farré1, D Navajas, E Prats, S Marti, R Guell, J M Montserrat, C Tebe, J Escarrabill.   

Abstract

BACKGROUND: Quality control procedures vary considerably among the providers of equipment for home mechanical ventilation (HMV).
METHODS: A multicentre quality control survey of HMV was performed at the home of 300 patients included in the HMV programmes of four hospitals in Barcelona. It consisted of three steps: (1) the prescribed ventilation settings, the actual settings in the ventilator control panel, and the actual performance of the ventilator measured at home were compared; (2) the different ventilator alarms were tested; and (3) the effect of differences between the prescribed settings and the actual performance of the ventilator on non-programmed readmissions of the patient was determined.
RESULTS: Considerable differences were found between actual, set, and prescribed values of ventilator variables; these differences were similar in volume and pressure preset ventilators. The percentage of patients with a discrepancy between the prescribed and actual measured main ventilator variable (minute ventilation or inspiratory pressure) of more than 20% and 30% was 13% and 4%, respectively. The number of ventilators with built in alarms for power off, disconnection, or obstruction was 225, 280 and 157, respectively. These alarms did not work in two (0.9%), 52 (18.6%) and eight (5.1%) ventilators, respectively. The number of non-programmed hospital readmissions in the year before the study did not correlate with the index of ventilator error.
CONCLUSIONS: This study illustrates the current limitations of the quality control of HMV and suggests that improvements should be made to ensure adequate ventilator settings and correct ventilator performance and ventilator alarm operation.

Entities:  

Mesh:

Year:  2006        PMID: 16467068      PMCID: PMC2111198          DOI: 10.1136/thx.2005.052647

Source DB:  PubMed          Journal:  Thorax        ISSN: 0040-6376            Impact factor:   9.139


  12 in total

1.  Inaccuracy of tidal volume delivered by home mechanical ventilators.

Authors:  F Lofaso; R Fodil; H Lorino; K Leroux; A Quintel; A Leroy; A Harf
Journal:  Eur Respir J       Date:  2000-02       Impact factor: 16.671

Review 2.  Rationale for the use of non-invasive ventilation in chronic ventilatory failure.

Authors:  P M Turkington; M W Elliott
Journal:  Thorax       Date:  2000-05       Impact factor: 9.139

3.  Performance characteristics of 10 home mechanical ventilators in pressure-support mode: a comparative bench study.

Authors:  Anne Battisti; Didier Tassaux; Jean-Paul Janssens; Jean-Bernard Michotte; Samir Jaber; Philippe Jolliet
Journal:  Chest       Date:  2005-05       Impact factor: 9.410

4.  Quality control of equipment in home mechanical ventilation: a European survey.

Authors:  R Farre; S J Lloyd-Owen; N Ambrosino; G Donaldson; J Escarrabill; B Fauroux; D Robert; B Schoenhofer; A Simonds; J A Wedzicha
Journal:  Eur Respir J       Date:  2005-07       Impact factor: 16.671

5.  Quality control of mechanical ventilation at the patient's home.

Authors:  Ramon Farré; Esther Giró; Vinyet Casolivé; Daniel Navajas; Joan Escarrabill
Journal:  Intensive Care Med       Date:  2002-12-14       Impact factor: 17.440

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Patterns of home mechanical ventilation use in Europe: results from the Eurovent survey.

Authors:  S J Lloyd-Owen; G C Donaldson; N Ambrosino; J Escarabill; R Farre; B Fauroux; D Robert; B Schoenhofer; A K Simonds; J A Wedzicha
Journal:  Eur Respir J       Date:  2005-06       Impact factor: 16.671

8.  The appropriate setting of noninvasive pressure support ventilation in stable COPD patients.

Authors:  M Vitacca; S Nava; M Confalonieri; L Bianchi; R Porta; E Clini; N Ambrosino
Journal:  Chest       Date:  2000-11       Impact factor: 9.410

9.  Frequency, causes, and outcome of home ventilator failure.

Authors:  S Srinivasan; S M Doty; T R White; V H Segura; M T Jansen; S L Davidson Ward; T G Keens
Journal:  Chest       Date:  1998-11       Impact factor: 9.410

10.  Assessment of physiologic variables and subjective comfort under different levels of pressure support ventilation.

Authors:  Michele Vitacca; Luca Bianchi; Ercole Zanotti; Andrea Vianello; Luca Barbano; Roberto Porta; Enrico Clini
Journal:  Chest       Date:  2004-09       Impact factor: 9.410

View more
  8 in total

1.  Risk management of the home ventilator dependent patient.

Authors:  A K Simonds
Journal:  Thorax       Date:  2006-05       Impact factor: 9.139

2.  Pediatric Chronic Home Invasive Ventilation.

Authors:  Paul E Moore; Debra Boyer; Michael G O'Connor; Christopher D Baker; Jordan S Rettig; Laura Sterni; Ann Halbower; Kevin C Wilson; Carey C Thomson
Journal:  Ann Am Thorac Soc       Date:  2016-07

3.  An Official American Thoracic Society Clinical Practice Guideline: Pediatric Chronic Home Invasive Ventilation.

Authors:  Laura M Sterni; Joseph M Collaco; Christopher D Baker; John L Carroll; Girish D Sharma; Jan L Brozek; Jonathan D Finder; Veda L Ackerman; Raanan Arens; Deborah S Boroughs; Jodi Carter; Karen L Daigle; Joan Dougherty; David Gozal; Katharine Kevill; Richard M Kravitz; Tony Kriseman; Ian MacLusky; Katherine Rivera-Spoljaric; Alvaro J Tori; Thomas Ferkol; Ann C Halbower
Journal:  Am J Respir Crit Care Med       Date:  2016-04-15       Impact factor: 21.405

4.  Actual performance of mechanical ventilators in ICU: a multicentric quality control study.

Authors:  Leonardo Govoni; Raffaele L Dellaca'; Oscar Peñuelas; Giacomo Bellani; Antonio Artigas; Miquel Ferrer; Daniel Navajas; Antonio Pedotti; Ramon Farré
Journal:  Med Devices (Auckl)       Date:  2012-12-20

5.  Home mechanical ventilation in childhood-onset hereditary neuromuscular diseases: 13 years' experience at a single center in Korea.

Authors:  Young Joo Han; June Dong Park; Bongjin Lee; Yu Hyeon Choi; Dong In Suh; Byung Chan Lim; Jong-Hee Chae
Journal:  PLoS One       Date:  2015-03-30       Impact factor: 3.240

Review 6.  Technology for noninvasive mechanical ventilation: looking into the black box.

Authors:  Ramon Farré; Daniel Navajas; Josep M Montserrat
Journal:  ERJ Open Res       Date:  2016-03-11

Review 7.  Trends in mechanical ventilation: are we ventilating our patients in the best possible way?

Authors:  Raffaele L Dellaca'; Chiara Veneroni; Ramon Farre'
Journal:  Breathe (Sheff)       Date:  2017-06

8.  Monitoring Patient/Ventilator Interactions: Manufacturer's Perspective.

Authors:  Gerard Evers; Carl Van Loey
Journal:  Open Respir Med J       Date:  2009-03-12
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.