Inaccuracy of tidal volume delivered by home mechanical ventilators.

F Lofaso1, R Fodil, H Lorino, K Leroux, A Quintel, A Leroy, A Harf.   

Abstract

Ideally, the inspired (tidal) volume (V(T)) provided by a volume-controlled ventilation device should not change when the pressure imposed on the ventilator varies. A bench study evaluation of V(T) versus pressure was performed on 10 commercially available devices. The difference between the desired V(T) and the observed V(T) reached 100 mL for some devices when inspiratory resistance was at its lowest, rising to 150 mL when inspiratory resistance was increased to obtain peak airway pressure of 60 cmH2O. The present data indicate that some home ventilators are inaccurate in delivering the preset tidal volume when the pressure imposed on the ventilator is increased to simulate high airway resistance.

Mesh:

Year:  2000        PMID: 10706502     DOI: 10.1034/j.1399-3003.2000.15b20.x

Source DB:  PubMed          Journal:  Eur Respir J        ISSN: 0903-1936            Impact factor:   16.671


  5 in total

1.  Bench studies evaluating devices for non-invasive ventilation: critical analysis and future perspectives.

Authors:  Carlo Olivieri; Roberta Costa; Giorgio Conti; Paolo Navalesi
Journal:  Intensive Care Med       Date:  2011-11-29       Impact factor: 17.440

2.  Accuracy of tidal volume delivered by home mechanical ventilation during mouthpiece ventilation: A bench evaluation.

Authors:  Adam Ogna; Helene Prigent; Line Falaize; Karl Leroux; Dante Santos; Isabelle Vaugier; David Orlikowski; Frederic Lofaso
Journal:  Chron Respir Dis       Date:  2016-07-08       Impact factor: 2.444

3.  Quality control of mechanical ventilation at the patient's home.

Authors:  Ramon Farré; Esther Giró; Vinyet Casolivé; Daniel Navajas; Joan Escarrabill
Journal:  Intensive Care Med       Date:  2002-12-14       Impact factor: 17.440

4.  Performance of mechanical ventilators at the patient's home: a multicentre quality control study.

Authors:  R Farré; D Navajas; E Prats; S Marti; R Guell; J M Montserrat; C Tebe; J Escarrabill
Journal:  Thorax       Date:  2006-02-07       Impact factor: 9.139

5.  Speech effects of a speaking valve versus external PEEP in tracheostomized ventilator-dependent neuromuscular patients.

Authors:  Hélène Prigent; Marine Garguilo; Sophie Pascal; Samuel Pouplin; Justine Bouteille; Michèle Lejaille; David Orlikowski; Frédéric Lofaso
Journal:  Intensive Care Med       Date:  2010-06-10       Impact factor: 17.440

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.