Literature DB >> 16466152

Evaluation of a contemporary CAD/CAM system.

T McGarry1, B McHugh.   

Abstract

The objective of this study is to investigate the accuracy of the TracerCAD system in measuring a model of known dimensions and volumes. A cylindrical nylon 6.6 model was prepared. Landmarks were added to indicate proximal, distal, anterior, medial, and lateral. Four additional landmarks were added at regular intervals along the length of the cylinder with a view to calculating diameters and volumes relative to these landmarks. The model was measured using a comparator with a guaranteed accuracy of 0.01 mm and was traced using the TracerCAD system (Test 1). The test was repeated with the model rotated by 90 degrees (Test 2), to determine whether there were any effects related to the orientation of the model in relation to the transmitter. The difference in average volume between measured results and TracerCAD scans was between 0.20% and -1.96%. Individual trace volumes varied between -0.0085% and -4.50%. In all volumes measured in Tests 1 and 2, all maximum volume percentage differences measured greater than 3%.

Entities:  

Mesh:

Year:  2005        PMID: 16466152     DOI: 10.1080/03093640500199497

Source DB:  PubMed          Journal:  Prosthet Orthot Int        ISSN: 0309-3646            Impact factor:   1.895


  4 in total

Review 1.  Residual limb volume change: systematic review of measurement and management.

Authors:  Joan E Sanders; Stefania Fatone
Journal:  J Rehabil Res Dev       Date:  2011

2.  Reliability and validity of 3D limb scanning for ankle-foot orthosis fitting.

Authors:  Olivia A Powers; Jeff R Palmer; Jason M Wilken
Journal:  Prosthet Orthot Int       Date:  2022-02-01       Impact factor: 1.672

3.  Residual limb volume fluctuations in transfemoral amputees.

Authors:  Linda Paternò; Michele Ibrahimi; Elisa Rosini; Giuseppe Menfi; Vito Monaco; Emanuele Gruppioni; Leonardo Ricotti; Arianna Menciassi
Journal:  Sci Rep       Date:  2021-06-10       Impact factor: 4.379

4.  Validity and reliability of a novel 3D scanner for assessment of the shape and volume of amputees' residual limb models.

Authors:  Elena Seminati; David Canepa Talamas; Matthew Young; Martin Twiste; Vimal Dhokia; James L J Bilzon
Journal:  PLoS One       Date:  2017-09-08       Impact factor: 3.240