Literature DB >> 16425310

Forces and torques during fracture reduction: Intraoperative measurements in the femur.

Thomas Gösling1, Rolf Westphal, Jens Faülstich, Kirsten Sommer, Friedrich Wahl, Christian Krettek, Tobias Hufner.   

Abstract

Reduction is a crucial step in fracture treatment. We determined intraoperative peak forces and torques during fracture reduction in seven patients with eight fractures of the femoral shaft. All fractures were temporarily stabilized by external fixation. Force and torque measurements were performed during the subsequent intramedullary nailing procedure. A three-dimensional load cell was attached to the distal femur fragment using two Schanz screws. All forces and torques were registered on-line during the reduction process. The maximum resulting force was 411 N, the maximum resulting torque 74 N x m. The highest force was observed along the shaft axis with 396 N for distraction. The maximum torque value was measured around the frontal axis, being 74 N x m for antecurvature. These results may assist the development of new reduction techniques and devices. Copyright 2006 Orthopaedic Research Society.

Entities:  

Mesh:

Year:  2006        PMID: 16425310     DOI: 10.1002/jor.20045

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  5 in total

1.  Introduction of a computer-based method for automated planning of reduction paths under consideration of simulated muscular forces.

Authors:  Jan Buschbaum; Rainer Fremd; Tim Pohlemann; Alexander Kristen
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-03-20       Impact factor: 2.924

2.  Computer-assisted fracture reduction: a new approach for repositioning femoral fractures and planning reduction paths.

Authors:  Jan Buschbaum; Rainer Fremd; Tim Pohlemann; Alexander Kristen
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-05-10       Impact factor: 2.924

Review 3.  Medical Robotics in Bone Fracture Reduction Surgery: A Review.

Authors:  Long Bai; Jianxing Yang; Xiaohong Chen; Yuanxi Sun; Xingyu Li
Journal:  Sensors (Basel)       Date:  2019-08-18       Impact factor: 3.576

4.  Design and Evaluation of a Percutaneous Fragment Manipulation Device for Minimally Invasive Fracture Surgery.

Authors:  Ioannis Georgilas; Giulio Dagnino; Beatriz Alves Martins; Payam Tarassoli; Samir Morad; Konstantinos Georgilas; Paul Koehler; Roger Atkins; Sanja Dogramadzi
Journal:  Front Robot AI       Date:  2019-10-30

5.  Design and evaluation of an intelligent reduction robot system for the minimally invasive reduction in pelvic fractures.

Authors:  Chunpeng Zhao; Yu Wang; Xinbao Wu; Gang Zhu; Shuchang Shi
Journal:  J Orthop Surg Res       Date:  2022-04-04       Impact factor: 2.359

  5 in total