Disinfection of needleless catheter connectors and access ports with alcohol may not prevent microbial entry: the promise of a novel antiseptic-barrier cap.

Steve Z Menyhay1, Dennis G Maki.   

Abstract

BACKGROUND: Needleless valve connectors for vascular catheters are widely used throughout the United States because they reduce the risk of biohazardous injuries from needlesticks and exposure to bloodborne pathogens, such as human immunodeficiency virus and hepatitis C virus. Patients with long-term central venous catheters are at significant risk of acquiring catheter-related bloodstream infections caused by microbes that gain access through the connection between the administration set and the catheter or an injection port. Most healthcare practitioners wipe the membranous septum of the needleless connector or the injection port with 70% alcohol before accessing it. We report a simulation study of the efficacy of conventional alcohol disinfection before access, compared with that of a novel antiseptic-barrier cap that, when threaded onto a needleless luer-activated valved connector, allows a chlorhexidine-impregnated sponge to come into continuous contact with the membranous surface; after removal of the cap, there is no need to disinfect the surface with alcohol before accessing it.
METHODS: One hundred five commercial, needleless luer-activated valved connectors, each accessible by a blunt male-connector luer-lock attachment, were purchased from 3 manufacturers and were tested. The membranous septum of each test device was first heavily contaminated with approximately 10(5) colony-forming units of Enterococcus faecalis and then was allowed to dry for 24 hours. Fifteen of the contaminated devices were not disinfected (positive controls), 30 were conventionally disinfected with a commercial 70% alcohol pledget, and 60 had the antiseptic cap threaded onto the connector and then removed after 10 minutes. The test connectors were then accessed with a sterile syringe containing nutrient broth media, which was injected, captured on the downstream side of the intraluminal fluid pathway, and cultured quantitatively.
RESULTS: All 15 control connectors (100%) showed massive transmission of microorganisms across the membranous septum (4,500-10,000 colony-forming units). Of the 30 connectors accessed after conventional disinfection with 70% alcohol, 20 (67%) showed transmission of microorganisms (442-25,000 colony-forming units). In contrast, of the 60 connectors cultured after application of the novel antiseptic cap, only 1 (1.6%) showed any transmission of microorganisms (P < .001).
CONCLUSIONS: The findings of this study show that, if the membranous septum of a needleless luer-activated connector is heavily contaminated, conventional disinfection with 70% alcohol does not reliably prevent entry of microorganisms. In contrast, the antiseptic-barrier cap provided a high level of protection, even in the presence of very heavy contamination. This novel technology deserves to be studied in a clinical trial.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16418982     DOI: 10.1086/500280

Source DB:  PubMed          Journal:  Infect Control Hosp Epidemiol        ISSN: 0899-823X            Impact factor:   3.254


  12 in total

1.  Leaving more than your fingerprint on the intravenous line: a prospective study on propofol anesthesia and implications of stopcock contamination.

Authors:  Devon C Cole; Tezcan Ozrazgat Baslanti; Nikolaus L Gravenstein; Nikolaus Gravenstein
Journal:  Anesth Analg       Date:  2015-04       Impact factor: 5.108

2.  Guidelines for the prevention of intravascular catheter-related infections.

Authors:  Naomi P O'Grady; Mary Alexander; Lillian A Burns; E Patchen Dellinger; Jeffrey Garland; Stephen O Heard; Pamela A Lipsett; Henry Masur; Leonard A Mermel; Michele L Pearson; Issam I Raad; Adrienne G Randolph; Mark E Rupp; Sanjay Saint
Journal:  Clin Infect Dis       Date:  2011-04-01       Impact factor: 9.079

3.  Needleless closed system does not reduce central venous catheter-related bloodstream infection: a retrospective study.

Authors:  Mitsuru Ishizuka; Hitoshi Nagata; Kazutoshi Takagi; Keiichi Kubota
Journal:  Int Surg       Date:  2013 Jan-Mar

4.  Central venous port systems as an integral part of chemotherapy.

Authors:  Ulf K Teichgräber; Robert Pfitzmann; Herbert A F Hofmann
Journal:  Dtsch Arztebl Int       Date:  2011-03-04       Impact factor: 5.594

5.  Needleless connectors: the vascular access catheter's microbial gatekeeper.

Authors:  Evonne Curran
Journal:  J Infect Prev       Date:  2016-07-25

Review 6.  Disinfection of Needleless Connector Hubs: Clinical Evidence Systematic Review.

Authors:  Nancy L Moureau; Julie Flynn
Journal:  Nurs Res Pract       Date:  2015-05-14

7.  Closed catheter access system implementation in reducing the bloodstream infection rate in low birth weight preterm infants.

Authors:  Lily Rundjan; Rinawati Rohsiswatmo; Tiara Nien Paramita; Chrissela Anindita Oeswadi
Journal:  Front Pediatr       Date:  2015-03-16       Impact factor: 3.418

8.  A simple and easy in vitro model to test the efficacy of IV lines' needleless connectors against contamination.

Authors:  María Guembe; María Jesús Pérez-Granda; Luis Alcalá; Pablo Martín-Rabadán; Emilio Bouza
Journal:  Intensive Care Med Exp       Date:  2014-11-07

Review 9.  Reducing the risk of infection associated with vascular access devices through nanotechnology: a perspective.

Authors:  Li Zhang; Samantha Keogh; Claire M Rickard
Journal:  Int J Nanomedicine       Date:  2013-11-21

Review 10.  The Role of Antifungals against Candida Biofilm in Catheter-Related Candidemia.

Authors:  Emilio Bouza; Jesús Guinea; María Guembe
Journal:  Antibiotics (Basel)       Date:  2014-12-25
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.