María Guembe1, María Jesús Pérez-Granda, Luis Alcalá, Pablo Martín-Rabadán, Emilio Bouza.
Abstract
BACKGROUND: Hub colonization after manipulation is responsible for 29% to 60% of catheter-related bloodstream infections (C-RBSI). Prevention can be achieved by the use of hub connectors, but its efficacy is generally based on instillation of high concentrations of microorganisms, which do not reflect the real contamination in daily practice. Our purpose was to create an in vitro model lasting long enough to be used for the comparison of the efficacy between various connectors against contamination simulating the real daily handling.Entities:
Year: 2014 PMID: 26266924 PMCID: PMC4513010 DOI: 10.1186/s40635-014-0027-9
Source DB: PubMed Journal: Intensive Care Med Exp ISSN: 2197-425X
Figure 1Experimental model of the blood culture with the cannula and the CLAVE™.
Characteristics of the experiment regarding contamination model and type of instillation fluid
|
|
|
|
|
|
|---|---|---|---|---|
| Handsa | Saline |
| 76.25 (65.92 to 125.01) |
|
| Propofol |
| 74.91 (51.45 to 99.00) |
| |
| 0.5 MF SAb | Saline | MSSA (ATCC 29213) | 104.92 (45.97 to 239.94) | NA |
| Propofol | MSSA (ATCC 29213) | 147.89 (66.04 to 236.58) | NA | |
| 0.05 MF SAc | Saline | MSSA (ATCC 29213) | 240.20 (154.82 to 360.00) | NA |
| Propofol | MSSA (ATCC 29213) | 66.11 (58.01 to 69.11) | NA | |
| Controld | Saline | NA | NA | NA |
| Propofol | NA | NA | NA |
aManipulation of the connector without gloves. bManipulation of the connector with gloves impregnated with a 0.5 McFarland solution of Staphylococcus aureus ATCC 29213. cManipulation of the connector with gloves impregnated with a 0.05 McFarland solution of Staphylococcus aureus ATCC 29213. dStandard sterile manipulation with clean gloves and with disinfection of connectors using alcohol. MTP, median time to positivity; IQR, interquartile range; SA, Staphylococcus aureus; MF, McFarland; CoNS, coagulase-negative staphylococci.
Median time to positivity of the bottles comparing different contamination models and different fluids
|
|
|
| ||
|---|---|---|---|---|
|
|
|
| ||
| Saline | 76.25 (65.95 to 125.00) | 104.92 (45.96 to 239.94) | 240.20 (154.82 to 360.00) |
|
| Propofol | 74.91 (51.44 to 98.99) | 147.89 (66.03 to 236.58) | 66.11 (58.01 to 69.11) | 0.24 |
|
| 0.69 | 1.00 |
| |
aWe found statistically significant differences between the 0.05 McFarland S. aureus model and the hand model. MTP, median time to positivity; SA, Staphylococcus aureus; MF, McFarland.
Figure 2Main results. MF, McFarland; SA, Staphylococcus aureus; MTP, median time to positivity.