Follow-up and final results of the Oslo I Study comparing screen-film mammography and full-field digital mammography with soft-copy reading.

P Skaane1, A Skjennald, K Young, E Egge, I Jebsen, E M Sager, B Scheel, E Søvik, A K Ertzaas, S Hofvind, M Abdelnoor.   

Abstract

PURPOSE: To compare cancer detection rates of screen-film (SFM) and full-field digital mammography (FFDM) with soft-copy reading in a screening program including the initial positive scores for interval cancers and cancers in the subsequent screening round, and to analyze the false-negative FFDM interpretations.
MATERIAL AND METHODS: Using a paired study design, 3683 women underwent SFM and FFDM in a population-based screening program. Two standard views of each breast were acquired. The images were interpreted without previous films for comparison. Independent double reading using a 5-point rating scale for probability of cancer was used for each modality. An examination was defined as positive if at least one of the two independent readers scored 2 or higher on the 5-point rating scale. SFM-positive cases were discussed in a SFM consensus meeting and FFDM-positive cases in a separate FFDM consensus meeting before recall. The study population was followed for more than 2 years so that interval cancers and screen-detected cancers in the subsequent screening round could be included. Cancer detection rates were compared using the McNemar test for paired proportions. The kappa statistic and Wilcoxon signed-rank test for matched pairs were used for comparing rating scores. The reading time was recorded for all FFDM interpretations.
RESULTS: A total of 31 cancers (detection rate 0.84%) were diagnosed initially, of which SFM detected 28 and FFDM 23 (McNemar test P=0.23, discordant pair 8 and 3). Two cancers with a positive score at initial SFM reading and three with a positive score at initial FFDM reading were dismissed at SFM and FFDM consensus meetings, respectively. The difference in cancer detection after recall (discordant pair 11 and 5) was not significant (McNemar test, P=0.21). Of the 10 interval cancers and 16 screen-detected cancers in the subsequent round, 3 had true-positive SFM scores while 4 had true-positive FFDM scores in the initial reading session. A total of 38 cancers therefore had a positive result at double reading at one or both modalities, 31 at SFM and 27 at FFDM (McNemar test, P=0.48). Comparison of SFM and FFDM interpretations using the mean score for each case revealed no statistically significant difference between the two modalities (Wilcoxon signed-rank test for matched pairs; P-value=0.228). Two initial round cancers (one tumor found incidentally at work-up for a mass proved to be a simple cyst with a positive score at FFDM but a negative score at SFM, and one tumor with positive score at SFM but negative score at FFDM due to positioning failure) were excluded from the further analysis. Excluding these two cancers from comparison, there were 31% (22 of 72) false-negative SFM and 47% (34 of 72) false-negative FFDM individual interpretations. The overall mean interpretation time for normal FFDM examinations was 45 s. For most false-negative FFDM results, the reading time was shorter or longer than for normal examinations. The recorded FFDM interpretation time was noticeably short for several overlooked cancers manifesting as microcalcifications (ductal carcinoma in situ).
CONCLUSION: There is no statistically significant difference in cancer detection rate between SFM and FFDM with soft-copy reading in a mammography screening program. Analysis of cancers missed at FFDM with soft-copy reading indicates that close attention has to be paid to systematic use of image display protocols.

Entities:  

Mesh:

Year:  2005        PMID: 16372686     DOI: 10.1080/02841850500223547

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  13 in total

1.  Impact of the Introduction of Digital Mammography in an Organized Screening Program on the Recall and Detection Rate.

Authors:  Cinzia Campari; Paolo Giorgi Rossi; Carlo Alberto Mori; Sara Ravaioli; Andrea Nitrosi; Rita Vacondio; Pamela Mancuso; Antonella Cattani; Pierpaolo Pattacini
Journal:  J Digit Imaging       Date:  2016-04       Impact factor: 4.056

2.  Mammographic performance in a population-based screening program: before, during, and after the transition from screen-film to full-field digital mammography.

Authors:  Solveig Hofvind; Per Skaane; Joann G Elmore; Sofie Sebuødegård; Solveig Roth Hoff; Christoph I Lee
Journal:  Radiology       Date:  2014-04-01       Impact factor: 11.105

3.  Computer-aided detection system for clustered microcalcifications: comparison of performance on full-field digital mammograms and digitized screen-film mammograms.

Authors:  Jun Ge; Lubomir M Hadjiiski; Berkman Sahiner; Jun Wei; Mark A Helvie; Chuan Zhou; Heang-Ping Chan
Journal:  Phys Med Biol       Date:  2007-01-23       Impact factor: 3.609

Review 4.  Digital mammography: what do we and what don't we know?

Authors:  Ulrich Bick; Felix Diekmann
Journal:  Eur Radiol       Date:  2007-02-14       Impact factor: 5.315

5.  Observer variability in screen-film mammography versus full-field digital mammography with soft-copy reading.

Authors:  Per Skaane; Felix Diekmann; Corinne Balleyguier; Susanne Diekmann; Jean-Charles Piguet; Kari Young; Michael Abdelnoor; Loren Niklason
Journal:  Eur Radiol       Date:  2008-02-27       Impact factor: 5.315

Review 6.  Breast cancer screening: an evidence-based update.

Authors:  Mackenzie S Fuller; Christoph I Lee; Joann G Elmore
Journal:  Med Clin North Am       Date:  2015-03-05       Impact factor: 5.456

7.  Conspicuity of breast cancer according to histopathological type and breast density when imaged by full-field digital mammography compared with screen-film mammography.

Authors:  Katja Pinker; Nicholas Perry; S Vinnicombe; S Shiel; M Weber
Journal:  Eur Radiol       Date:  2010-08-04       Impact factor: 5.315

8.  A novel image toggle tool for comparison of serial mammograms: automatic density normalization and alignment-development of the tool and initial experience.

Authors:  Satoshi Honda; Hiroko Tsunoda; Wataru Fukuda; Yukihisa Saida
Journal:  Jpn J Radiol       Date:  2014-09-20       Impact factor: 2.374

9.  The impact of digital mammography on screening a young cohort of women for breast cancer in an urban specialist breast unit.

Authors:  Nicholas M Perry; N Patani; S E Milner; K Pinker; K Mokbel; P C Allgood; S W Duffy
Journal:  Eur Radiol       Date:  2010-10-01       Impact factor: 5.315

10.  Full-field digital mammography compared to screen film mammography in the prevalent round of a population-based screening programme: the Vestfold County Study.

Authors:  Einar Vigeland; Herman Klaasen; Tor Audun Klingen; Solveig Hofvind; Per Skaane
Journal:  Eur Radiol       Date:  2007-08-07       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.