Literature DB >> 16352274

Extra-ascending aortic versus intra-descending aortic balloon counterpulsation-effect on coronary artery blood flow.

Andrew N Davies1, William S Peters, Tonglin Su, Colin E Sullivan, Theo Perkidides, F Paget Milsom, Geoffrey White.   

Abstract

BACKGROUND: Diastolic counterpulsation has been used to provide circulatory augmentation for chronic heart failure or for short-term cardiac support. Recently an extra-aortic balloon (EAB) counterpulsation device has been proposed. AIM: To compare the circulatory effects of counterpulsation using the EAB or an intra-aortic balloon (IAB) in the acute pig model.
METHODS: In six anaesthetized great white pigs (paced at 100 bpm), ECG, arterial and central venous pressures, flow in the coronary circulation and descending thoracic aorta were measured. Baseline data was collected, then with the EAB or an IAB fitted using 1:1 and 1:2 counterpulsation modes. Baseline data was compared to EAB and IAB data in 1:1 mode. Assisted beat data compared to unassisted beat data was also analysed in 1:2 mode.
RESULTS: Both devices augmented peak diastolic arterial pressure, and decreased afterload. EAB counterpulsation increased diastolic coronary flow in both the 1:1 mode by 69% (p < 0.05) and in the 1:2 mode by 63% (assisted versus unassisted beat, p < 0.05). The IAB significantly increased diastolic coronary flow only in the 1:2 mode by 28% (p < 0.01). Both devices augmented total coronary flow and some augmentation of aortic flow was observed.
CONCLUSION: The circulatory effect of the EAB and IAB counterpulsation were comparable. This suggests the EAB could be used as a non-blood contacting heart assist device in patients suffering moderate-severe heart failure.

Entities:  

Mesh:

Year:  2005        PMID: 16352274     DOI: 10.1016/j.hlc.2005.03.018

Source DB:  PubMed          Journal:  Heart Lung Circ        ISSN: 1443-9506            Impact factor:   2.975


  5 in total

1.  The future of adult cardiac assist devices: novel systems and mechanical circulatory support strategies.

Authors:  Carlo R Bartoli; Robert D Dowling
Journal:  Cardiol Clin       Date:  2011-11       Impact factor: 2.213

2.  Concept and first experimental results of a new ferromagnetic assist device for extra-aortic counterpulsation.

Authors:  Christoph T Starck; Jakob Becker; Roland Fuhrer; Simon Sündermann; Jan Wendelin Stark; Volkmar Falk
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-09-22

Review 3.  Understanding the C-pulse device and its potential to treat heart failure.

Authors:  Virna L Sales; Patrick M McCarthy
Journal:  Curr Heart Fail Rep       Date:  2010-03

Review 4.  Aortic counterpulsation: C-pulse and other devices for cardiac support.

Authors:  Pallavi Solanki
Journal:  J Cardiovasc Transl Res       Date:  2014-02-20       Impact factor: 4.132

5.  Preliminary Results From the C-Pulse OPTIONS HF European Multicenter Post-Market Study.

Authors:  Antonia Schulz; Thomas Krabatsch; Jan D Schmitto; Roland Hetzer; Mirko Seidel; Pascal M Dohmen; Holger Hotz
Journal:  Med Sci Monit Basic Res       Date:  2016-02-18