Outcome of percutaneous hybrid coronary revascularization: bare metal stents jeopardize the benefit of sirolimus-eluting stents in the real world.

Massimo Fineschi1, Tommaso Gori, Carlo Pierli, Stefano Casini, Giuseppe Sinicropi, Alberto Buti, Alessia Del Pasqua, Achille Bravi.   

Abstract

BACKGROUND: In an effort to contain procedural costs while limiting the risk of in-stent restenosis, hybrid percutaneous revascularization (ie, stenting with at least one sirolimus-eluting stent [SES] and at least one bare metal stent [BMS] in the same patient) is felt to be a cost-effective alternative to exclusive SES use.
OBJECTIVE: To describe the outcome of hybrid percutaneous revascularization for the treatment of patients with multiple coronary artery lesions. METHODS AND
RESULTS: Fifty-six patients (42 men; mean age [+/- SEM] 64+/-2) underwent hybrid stenting (average of 1.2 SES/patient and 1.3 BMS/patient). SES were used to treat lesions at higher restenotic potential, including longer lesions, smaller target vessels and bifurcation lesions (mean stent length [+/- SEM] was 21.1+/-1.2 mm for SES and 16.0+/-0.6 mm for BMS; stent diameter mean [+/- SEM] was 2.9+/-0.0 mm for SES and 3.1+/-0.1 mm for BMS; bifurcation lesions were 43% for SES and 7% for BMS; all P<0.01). At nine months of clinical follow-up, no death or myocardial infarction was reported. Twenty-one patients underwent clinically driven repeat coronary angiography at a mean (+/- SEM) of 8+/-1 of months (range two to 12 months) follow-up. Target lesion revascularization procedures were recorded in six patients (11%) for nine lesions (6%). Of these lesions, seven were categorized after blinded analysis as due to in-BMS restenosis and two to in-SES restenosis (P=0.01); three patients (5.4%) underwent reangioplasty for de novo lesions. There was one case of acute in-SES thrombosis. SES showed significantly less neointimal hyperplasia (late lumen loss was 0.4+/-0.1 mm for SES and 1.3+/-0.1 mm for BMS; loss index was 0.15+/-0.05 for SES and 0.48+/-0.05 for BMS; all P<0.001).
CONCLUSIONS: The use of SES resulted in less neointimal hyperplasia even when used to treat lesions at higher risk for restenosis based on angiographic characteristics. BMS implantation significantly limits this beneficial effect, compromising the outcome of hybrid percutaneous coronary revascularization.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16341297

Source DB:  PubMed          Journal:  Can J Cardiol        ISSN: 0828-282X            Impact factor:   5.223


  2 in total

1.  Concurrent drug eluting/bare metal stent implantation during percutaneous coronary intervention in target vessel: outcomes and 1-year follow-up.

Authors:  Alessandro Cuneo; Peter Bramlage; Matthias Hochadel; Jochen Senges; Christoph Nienaber; Karl-Heinz Kuck; Ulrich Tebbe
Journal:  Clin Res Cardiol       Date:  2011-12-09       Impact factor: 5.460

2.  Symptomatic failure after sirolimus-eluting stent implantation: a rare but challenging condition.

Authors:  Massimo Fineschi; Tommaso Gori; Carlo Pierli; Stefano Casini; Giuseppe Sinicropi; Alberto Buti; Alessandro Iadanza; Achille Bravi
Journal:  Can J Cardiol       Date:  2007-02       Impact factor: 5.223