Angiographic assessment of dimensions of 6F and 7F Mallinckrodt Softouch coronary contrast catheters from digital and cine arteriograms.

G Koning1, P M van der Zwet, C D von Land, J H Reiber.   

Abstract

Coronary contrast catheters are almost exclusively used for calibration purposes in quantitative coronary arteriography. In this study we have assessed the suitability of new 6F and 7F Mallinckrodt nylon catheter with Softouch tip with improved imaging specifications for such calibration purposes both from digital and cinefilm images using new analytical QCA-software packages (Philips DCI/ACA and Medis CMS). The average signed differences between the angiographically measured dimensions at 100% contrast fillings and acquired at 3 different kV-levels (60, 75 and 90 kV) were -3.3% and 0.6% for the 6F and 7F catheter tips, respectively as measured with the ACA-package on digital images, and -0.4% and 2.1%, respectively, as measured with the CMS-system on cinefilm images. The pooled standard deviations were 0.102 mm and 0.107 mm for the 6F and 7F catheter tips, respectively, as measured with the ACA-package, and 0.080 mm and 0.083 mm, as measured with the CMS-system. The deviations for the nylon shafts were much larger. It became also clear that neither the filling of the catheters, nor the kV-level used, had any appreciable effect on the measurement accuracy for the Softouch tips, which facilitates the frame selection in QCA-studies. From these data it can be concluded that the nontapering parts of the 6F and 7F Mallinckrodt Softouch tips are very well suitable for QCA calibration purposes, but that the nylon shafts are not.

Entities:  

Mesh:

Year:  1992        PMID: 1629641     DOI: 10.1007/bf01137536

Source DB:  PubMed          Journal:  Int J Card Imaging        ISSN: 0167-9899


  4 in total

1.  A new approach for the automated definition of path lines in digitized coronary angiograms.

Authors:  P M van der Zwet; I M Pinto; P W Serruys; J H Reiber
Journal:  Int J Card Imaging       Date:  1990

2.  Coronary artery quantitation and data management system for paired cineangiograms.

Authors:  W H Leung; W Sanders; E L Alderman
Journal:  Cathet Cardiovasc Diagn       Date:  1991-10

3.  Assessment of short-, medium-, and long-term variations in arterial dimensions from computer-assisted quantitation of coronary cineangiograms.

Authors:  J H Reiber; P W Serruys; C J Kooijman; W Wijns; C J Slager; J J Gerbrands; J C Schuurbiers; A den Boer; P G Hugenholtz
Journal:  Circulation       Date:  1985-02       Impact factor: 29.690

4.  Assessment of dimensions and image quality of coronary contrast catheters from cineangiograms.

Authors:  J H Reiber; C J Kooijman; A den Boer; P W Serruys
Journal:  Cathet Cardiovasc Diagn       Date:  1985
  4 in total
  4 in total

1.  Validation of an accurate method for three-dimensional reconstruction and quantitative assessment of volumes, lengths and diameters of coronary vascular branches and segments from biplane angiographic projections.

Authors:  E Wellnhofer; A Wahle; I Mugaragu; J Gross; H Oswald; E Fleck
Journal:  Int J Card Imaging       Date:  1999-10

2.  Impact of catheter size on reliability of quantitative coronary angiographic measurements (comparison of 4Fr and 6Fr catheters).

Authors:  Shigenori Ito; Kanako Kinoshita; Akiko Endo; Masato Nakamura; Toshiya Muramatsu
Journal:  Heart Vessels       Date:  2016-02-05       Impact factor: 2.037

3.  The impact of vessel and catheter position on the measurement accuracy in catheter-based quantitative coronary angiography.

Authors:  W Wunderlich; B Roehrig; F Fischer; H R Arntz; R Agrawal; A Morguet; H P Schultheiss; D Horstkotte
Journal:  Int J Card Imaging       Date:  1998-08

4.  Angiographically borderline left main coronary artery lesions: correlation of transthoracic doppler echocardiography and intravascular ultrasound: a pilot study.

Authors:  Zoltán Ruzsa; Attila Pálinkás; Tamás Forster; Imre Ungi; Albert Varga
Journal:  Cardiovasc Ultrasound       Date:  2011-06-14       Impact factor: 2.062

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.